2
dc2349af
DEMO MANUAL DC2349A
QUICK START PROCEDURE
NOTE 1. Evaluation of demonstration circuit 2349A requires
the Linear Technology USB serial controller, DC590B, and
the Linear Technology QuikEval™ software. Please refer
to DC590B demo manual for their setup and installation.
NOTE 2. An internet connection is required for the automatic
download and installation of the LTC5586 software GUI.
If an internet connection is not available, contact Linear
Technology for installation files
NOTE 3. Refer to Figure 1 for proper equipment setup.
NOTE 4. Make all connections without RF and DC applied.
NOTE 5. Care should be taken to never exceed absolute
maximum input ratings.
1. If not already done, install the QuikEval software
and connect the DC590B USB serial controller to an
internet-connected computer’s USB port.
2. Set the DC590B’s isolated supply voltage to 3.3V by
moving its VCCIO jumper (JP6) to 3.3V position.
3. Verify that the DC2349A’s jumpers are set to:
JP1 = INT (on-board 3.3V digital supply)
JP2 = RFA (RFA input selected)
4. While the RF sources and the DC power supplies are
switched off, connect all test equipment as shown in
Figure1.
5. When the DC2349A demo board is connected to the
DC590B USB serial controller, the QuikEval software
will automatically download and install the proper
software GUI.
6. While monitoring the voltage on VSENSE, increase
the VSUPPLY DC power supply voltage until VSENSE
reaches 5V (VSUPPLY voltage should be approxi-
mately5.5V). Verify the current consumption is close
to the figure listed in the data sheet.
7. Increase the IF amplifier common-mode output volt-
age, VCM, to 0.9V.
8. In the LTC5586 QuikEval computer GUI, verify that
“Connected” is displayed, then press the “Soft Reset”
button. See Figure 2.
9. Set the desired LO frequency and LO power, and turn
on the LO signal source.
10. Use the LTC5586 QuikEval GUI to configure the
LTC5586 to the same LO frequency.
11. Set the desired RF frequency and RF power, and turn
on the RF source.
12. Perform measurements.
13. The LTC5586 QuikEval GUI can be used to control and
configure various functions of the LTC5586. Press the
button “Show Regs” will make the LTC5586’s internal
register visible and accessible on screen. See Figure3.