AC/DC Medical Power Supply * Encased 150 W power supply with screw connection in 2.44" x 4" package * Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP * Low leakage current <100 A rated for BF applications * Risk management process according to ISO 14971 including risk management file * Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3 * * * * Active power factor correction >0.95 TPP 150 Series, 150 Watt CB Protection class l and ll prepared Scheme Operating up to 5000 m altitude prepared 5 year product warranty Open frame version with JST connection see TPP 150A Series www.tracopower.com/overview/tpp150a ES 60601-1 UL 60950-1 The TPP 150 series of 150 Watt AC/DC encased power supplies feature a reinforced double I/O isolation system according to latest medical safety standards (60601-1 3rd edition, 2 x MOPP). The earth leakage current is below 100 A which makes the units suitable for BF (body floating) applications. The excellent efficiency of up to 92% allows a high power density for the standard 2.44" x 4.0" packaging format. The full load operating temperature range is -25C to +70C while it goes up to 80C with 40% load derating. The EMC characteristic is dedicated for applications in industrial and domestic fields. High reliability is provided by the use of industrial quality grade components and an excellent thermal management. It makes the products an ideal solution for medical devices and for demanding safety and space critical applications. Models Models Order code Output voltage Output current max. Efficiency typ. TPP 150-112 12 VDC (10.8 - 13.2 VDC) 12.5 A 91 % TPP 150-115 15 VDC (13.5 - 16.5 VDC) 10 A 92 % TPP 150-124 24 VDC (21.6 - 26.4 VDC) 6.25 A 92 % TPP 150-128 28 VDC (25.2 - 30.8 VDC) 5.36 A 92 % TPP 150-136 36 VDC (32.4 - 39.6 VDC) 4.17 A 92 % TPP 150-148 48 VDC (43.2 - 52.8 VDC) 3.13 A 92 % www.tracopower.com Page 1 of 4 TPP 150 Series, 150 Watt Input Specifications Input voltage range - AC range (universal input) 85 - 264 VAC - DC range 120 - 370 VDC Input frequency 47 - 63 Hz Input current at full load - at 115 VAC / 230 VAC 1.7 A max. / 0.8 A max. Input protection T 3.15 A / 250 VAC (internal fuse) Input inrush current - at 230 VAC 60 A max. Zero load power consumption 0.6 W typ. Power factor 0.95 min. Output Specifications Voltage set accuracy 1% Output voltage adjustment 10% (by trim potentiometer) Regulation - Input variation (Vin min. to Vin max.) - Load variation (0 to 100%) 0.2% max. 0.5% max. Minimum load not required Temperature coefficient 0.02 %/K max. Hold-up time - Vin = 115 VAC 16 ms min. Start-up time 1 s max. Rise time 20 ms typ. Ripple and noise (20 MHz bandwidth) 12 Vout models: 15 Vout models: 24 Vout models: 28 Vout models: 36 Vout models: 48 Vout models: 120 mVp-p typ. w. cap. 1F/25V 1206 X7R MLCC 150 mVp-p typ. w. cap. 1F/25V 1206 X7R MLCC 220 mVp-p typ. w. cap. 1F/50V 1206 X7R MLCC 220 mVp-p typ. w. cap. 1F/50V 1206 X7R MLCC 250 mVp-p typ. w. cap. 1F/50V 1206 X7R MLCC 250 mVp-p typ. w. cap. 0.1F/100V 1206 X7R MLCC Overvoltage protection 115 - 135% of nominal Vout Overload protection 115 - 150% Iout typ. Short circuit protection Hiccup mode, continuous (automatic recovery) Transiente response - Peak deviation (25% load step change) 3% max. - Recovery time 500 s typ. Fan power supply - Temperature-sensitive speed control 12 VDC / 500 mA max. Capacitive load 12 Vout models: 15 Vout models: 24 Vout models: 28 Vout models: 36 Vout models: 48 Vout models: 10`400 F max. 6`600 F max. 2`600 F max. 1`900 F max. 1`150 F max. 650 F max. General Specifications Temperature ranges - Operating temperature -25C to +80C (with derating) - Storage temperature -40C to +80C Derating - Temperature 4.0 %/K above 70C - Low input voltage 1.33 %/V below 100 VAC Humidity (non condensing) 5 - 95 % rel. H. Altitude during operation 5000 m max. Switching frequency (at 230 VAC) Isolation voltage (60 s) (2 x MOPP insulation) 60 kHz typ. (pulse frequency modulation) - Input to Output 4000 VAC - Input/Output to Case 2000 VAC Isolation resistance (at 500 VDC) 100 MOhm min. All specifications valid at nominal input voltage, full load and +25C after warm-up time unless otherwise stated. www.tracopower.com Page 2 of 4 TPP 150 Series, 150 Watt General Specifications (continued) Leakage current (at 264 VAC / 60 Hz) 100 A max. Reliability - calculated MTBF at +25C acc. MIL-HDBK-217F 786`100 h Protection class class I and II prepared EMC emissions - Conducted / radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon Conducted: EN 55032 class B (internal filter) Radiated: EN 55032 class A (internal filter) - Harmonic current emissions IEC/EN 61000-3-2, class A & D - Voltage flicker IEC/EN 61000-3-3 EMC immunity - Electrostatic discharge ESD - RF field immunity - Electrical fast transients/burst immunity - Surge - Conducted RF - Magnetic field - Voltage dip and interruptions IEC/EN 60601-1-2, EN 55024 IEC/EN 61000-4-2, 15kV/8kV perf. criteria A IEC/EN 61000-4-3, 20V/m perf. criteria A IEC/EN 61000-4-4, 2kV perf. criteria A IEC/EN 61000-4-5, 1kV/2kV perf. criteria A IEC/EN 61000-4-6, 20 Vrms perf. criteria A IEC/EN 61000-4-8, 10A/m perf. criteria A IEC/EN 61000-4-11, see below Voltage dip and interruptions according to EN 60601-1-2 Reference: 230 VAC / 50Hz 30%, 500ms perf. criteria A 60%, 100ms perf. criteria A > 95%, 10ms perf. criteria A > 95%, 5000ms perf. criteria B Safety standards and certification - Certification documents IEC/EN/UL 60950-1, IEC/EN 60601-1 3rd edition, ANSI/AAMI ES60601-1:2005(R)2012 www.tracopower.com/overview/tpp150 Environment - Vibration acc. IEC 60068-2-6 - Shock acc. IEC 60068-2-27 - Thermal shock acc. MIL-STD-810F www.tracopower.com/info/reach-declaration.pdf Environmental compliance - Reach - RoHS RoHS directive 2011/65/EU Weight 272 g (9.59 oz) All specifications valid at nominal input voltage, full load and +25C after warm-up time unless otherwise stated. www.tracopower.com Page 3 of 4 TPP 150 Series, 150 Watt Outline Dimensions 116.8 (4.60) Molex 3 CON3 6 Input (CON1) Pin CON2 CON1 1 1 M3x0.5 holes 34.72 (1.367) 62.0 (2.44) Screw Terminal 101.6 (4.00) 109.22 (4.300) Output (CON2) Fan (CON3) Function Pin* Function Pin Function 1 Line 1-3 -Vout 1 -Fan 3 Neutral 4-6 +Vout 2 +Fan Trim potentiometer (under the fan) *Terminal rated for 7 A max. (at higher current connection has to be split) CON 1: Screw Terminal 49.2 (1.94) 39.2 (1.54) 1.2 (0.05) CON 2: Screw Terminal CON 3: Molex series mates with Molex crimp terminals: 2759 and Molex housing: 22-01-1022 103.00 (4.055) 95.20 (3.748) 6.9 (0.27) 10.8 (0.43) The fan`s durability is lower compared to the power supply and has only 2 years warranty. 44.45 (1.750) 25.00 (0.984) 18.5 (0.73) 8.8 (0.35) Dimensions in mm (inch) Tolerances: x.x 0.5 (x.xx 0.02) x.xx 0.25 (x.xxx 0.01) M3 screws (c) Copyright 2018 Traco Electronic AG Specifications can be changed without notice! Rev. May 30, 2018 www.tracopower.com Page 4 of 4 Mouser Electronics Authorized Distributor Click to View Pricing, Inventory, Delivery & Lifecycle Information: TRACO Power: TPP 150-128 TPP 150-124 TPP 150-148 TPP 150-136 TPP 150-112 TPP 150-115