date 06/20/2017│ page 2 of 5CUI Inc │ SERIES: ETMA 250W U │ DESCRIPTION: MEDICAL AC-DC POWER SUPPLY
INPUT
parameter conditions/description min typ max units
voltage 90 264 Vac
frequency 47 63 Hz
current at 90 Vac, 60 Hz 3.5 A
inrush current at 230 Vac, cold start 150 A
no load power consumption 0.5 W
OUTPUT
parameter conditions/description min typ max units
line regulation ±1 %
load regulation ±5 %
hold-up time at 115 Vac 16 ms
temperature coefcient ±0.05 %/°C
PROTECTIONS
parameter conditions/description
over voltage protection
12 V output
16 V output
19 V output
20 V output
24, 36 V outputs
48 V output
110
110
111
118
110
110
133
130
137
130
125
120
%
%
%
%
%
%
overload protection shutdown and auto restart 110 160 %
short circuit protection continuous, auto restart
SAFETY & COMPLIANCE
parameter conditions/description min typ max units
isolation voltage input to output for 2 seconds
input to frame ground for 2 seconds
4,242
2,121
Vdc
Vdc
insulation resistance input to output, at 500 Vdc
input to frame ground, at 500 Vdc
20
20
MΩ
MΩ
safety approvals medical UL/cUL UL 60601-1, TUV EN 60601-1, CB IEC 60601-1 4th edition
EMI/EMC FCC Part 18 Class B, EN 60601-1, EN 55011
MTBF at 25°C, max. load 140,000 hours
RoHS compliant yes
ENVIRONMENTAL
parameter conditions/description min typ max units
operating temperature 0 50 °C
storage temperature -10 70 °C
humidity non-condensing 20 90 %