B-45Z500
Z500
Profile Measuring System
Z500
High-Precision Sensor that Measures and
Displays an Object's Profile.
Features
OMRON's original line beam method provides a
complete solution to profile measurement problems.
Conventional non-contact measurement of the profile of an object commonly
uses a displacement sensor to measure the height of the object, by moving the
object or the sensor.
However, this system has several disadvantages, such as lower measurement
accuracy resulting from object or sensor movement as well as high system con-
struction cost.
By utilizing a unique wide beam method and 2-dimensional CCD, OMRON's
Z500 eliminates these problems.
Through its capability of measuring a diversely-shaped objects in a stable man-
ner, the Z500 can meet a variety of application needs.
Application
Principle of line beam method
A wide beam is applied to the object to be measured. A 2-dimensional CCD receives
the reflected light to measure the 2-dimensional profile of the object.
Measurement of connector pin
configuration
Height measurement for cream-
soldered joint on PCB
Rivet height measurement Door gap measurement
Detected image
Projector
lens Receiver lens
SW-CCD
Laser
F502-EN2-04.book Seite 45 Dienstag, 26. Juli 2005 5:48 17
B-46 Displacement sensors / Width-measuring Sensors
Features
Measures the shape of object instantly.
A complete solution to the disadvantages of conventional
measurement systems.
Measurement by displacement sensor
Measurement by Z500 model
Accurate and stable measurement.
OMRON's original 2-dimensional SW-CCD and multiple light
intensity control system enable stable measurement of ob-
jects with round shape and other surface conditions.
A variety of measurement items.
Various measurement items, such as level difference, width,
and edge position, can be selected depending on the specific
application.
Permitting simultaneous measurement of up to 8 items, the
Z500 is applicable to various measurement purposes.
Four types of monitor screens
Measurement data can be displayed on 4 types of monitor screens.
These screens enable analysis and evaluation of measurement data from various viewpoints.
When you
use Z500, ---
Sensor Head
Judgment/
measurement
Output
Workpiece
Moving
Personal computer/PLC
Displacement
sensor
Problems to be resolved
Object or sensor needs to be moved.
Personal computer or other device is
required for measurement data processing.
Increased system construction cost
Measurement accuracy deterioratio
n
caused by movement of object or
sensor and data processing.
Measurement takes a long time.
There is nothing for a conventional
measurement system but to make the best of it,
or we have to give up measurement.
Sensor Head
Monitor
Workpiece
Enjoy the following advantages from Z500 !
Moving the object or sensor is no longer necessary.
Various data processing and calculation functions are
performed automatically.
Lower system construction cost
Higher measurement accuracy
Shorter measurement time
Sophisticated and
high accuracy inline
inspection Judgment/
measurement
Output
2-point level difference (Step: 2 pts)
Width
Edge position
Profile monitor
Time-series change of profile (data
on cross section height) can be
checked on a 3D gray scale image.
Image monitor
Both measurement data and
profile image can be checked
at the same time.
Digital monitor
Two or more measurement
data can be checked at the
same time.
Trend monitor
Time-series change of mea-
surement data can be
checked.
F502-EN2-04.book Seite 46 Dienstag, 26. Juli 2005 5:48 17
B-47Z500
Z500
System configuration
Ordering Information
Note: Specify the required cable length when ordering.
Name Item Model Remarks
Sensor
Z500-SW2T Cable length: 2 m
Z500-SW6 Cable length: 0.5 m
Z500-SW17 Cable length: 2 m
Controller Z500-MC10E NPN input/output
Z500-MC15E PNP input/output
Console Z300-KP ——
Liquid Crystal Monitor F150-M05L ——
Sensor Extension Cable Z309-SC1R (See note) Cable length: 1.5 m, 3 m, 6 m, 8 m or 13 m
Monitor Cable F150-VM Cable length: 2 m
Power Supply Synchronization Sensor
Linear Sensor Controller
K3AS
Controller
Z500-MC10E
Z500-MC15E
Monitor Cable
F150-VM (cable length: 2 m)
Liquid Crystal Monitor
F150-M05L (pin input)
BNC Jack
(Provided with F150-VM)
Console
Z300-KP
(cable length: 2 m)
Monitor
Programmable Controller
Sensor
Z500-SW2T (cable length: 2 m)
Z500-SW6 (cable length: 0.5 m)
Z500-SW17(cable length: 2 m)
(Up to 2 Sensors can be connected.)
Sensor Extension Cable
Z309-SC1R (cable length:
1.5 m, 3 m, 6 m, 8 m or 13 m)
F502-EN2-04.book Seite 47 Dienstag, 26. Juli 2005 5:48 17
B-48 Displacement sensors / Width-measuring Sensors
Rating/Performance
Sensor Z500-SW2T/-SW6/-SW17
Note: 1 .Defined at 1/e2 (13.5%) of the density at the light center. Light may, however, be present outside this range and if the reflection factor of the light around the
workpiece is high compared to the workpiece, measurement may be affected.
2 .Error with respect to the theoretical line representing the displacement output for measurement of OMRON standard quartz glass. The linearity varies with
the type of workpiece.
3 .Error with respect to the theoretical line representing the displacement output for measurement of OMRON standard SUS blocks. The linearity varies with the
type of workpiece.
4 .Error with respect to the theoretical line representing the displacement output for measurement of OMRON standard white alumina ceramics. The linearity
varies with the type of workpiece.
5 .Displacement conversion value for peak-to-peak of displacement output. These figures are for measurement of OMRON standard quartz glass (mirror reflec-
Model Z500-SW2T Z500-SW6 Z500-SW17
Diffuse reflection Mirror reflection Diffuse reflection Mirror reflection Diffuse reflection Mirror reflection
Item
Measurement mode
Distance to
measurement center 5.2 mm
20 mm
(with beam cover
attached: 16 mm)
50 mm 44 mm 100 mm 94 mm
Measurement range ±0.8 mm ±5 mm ±4 mm ±20 mm ±16 mm
Light source
Visible-light semiconductor laser
(See note 10)
(Wavelength 650 nm,
1 mW max., Class 2)
Visible-light semiconductor laser
(Wavelength 658 nm, 15 mW max., Class 3B)
Beam dimensions (See note 1)
Reference distance:
20 µm × 4 mm TYP.
(Measurement region: 2 mm)
Reference distance:
30 µm × 24 mm TYP.
(Measurement region: 6 mm)
Reference distance:
60 µm × 45 mm TYP.
(Measurement region: 17 mm)
Linearity ±0.1 %F.S.
(See note 3)
±0.1 %F.S.
(See note 2) ±0.1 %F.S. (See note 4)
Resolution 0.25 µm (See notes 5 and 6) 0.3 µm (See notes 7 and 8) 1 µm (See notes 7 and 8)
Sampling cycle 9.94 ms
LED indicators
(LASER indicator) Lit while laser is ON.
Temperature characteristic
(See note 9) 0.01 %F.S./°C
Environ-
ment
resistance
Degree of protection IEC IP64 IEC IP66
Ambient operating
illumination Illumination at light-receiving surface: 3,000 lx max., incandescent light
Ambient temperature Operating: 0 to +50 °C, Storage: -15 to +60 °C (with no icing or condensation)
Ambient humidity Operating and storage: 35 to 85 % RH (with no condensation)
Vibration resistance 10 to 150 Hz (single amplitude: 0.35 mm) for 80 min. each in X, Y, and Z directions
Materials
Unit: Die-cast aluminum
Cable sheathing: Heat-resistant chlorinated vinyl
Connector: zinc alloy and brass
Cable length 2 m 0.5 m 2 m
Minimum bending radius 68 mm
Weight (including packaging) Approx. 600 g
(Unit: Approx. 350 g)
Approx. 700 g
(Unit: Approx. 600 g)
Approx. 800 g
(Unit: Approx. 600 g)
Accessories 3 ferrite cores, laser warning labels (English)
±0.1 % max.
80 mm 120 mm Distance
120 mm
80 mm
Digital output
Z500-SW17
F502-EN2-04.book Seite 48 Dienstag, 26. Juli 2005 5:48 17
B-49Z500
Z500
tion mode) or OMRON standard SUS blocks (diffuse reflection mode) at the measurement center. In strong magnetic fields, it may not be possible to maintain
resolution performance characteristics.
6 .These figures are for when the Sensor is connected to the Z500-MC10E/MC15E, the average number of measurements is 16. Measurement data are sent
to PC via RS-232C cable for calculation of their average values.
7 .Displacement conversion value for peak-to-peak of displacement output (for measurement of OMRON standard white alumina ceramic at the measurement
center). In strong magnetic fields, it may not be possible to maintain resolution performance characteristics.
8 .With the Z500-MC10E/MC15E, at an average number of measurements of 64. Measurement data are sent to PC via RS-232C cable for calculation of their
average values.
9 .Value for measurement with the space between the Sensor and the workpiece (white alumina ceramic) secured with an aluminum jig.
10.Higher power laser type (Class 3B) is also available. For further information, please contact us.
F502-EN2-04.book Seite 49 Dienstag, 26. Juli 2005 5:48 17
B-50 Displacement sensors / Width-measuring Sensors
Controller Z500-MC10E/MC15E
Note: For measurement at an average number of times of 64 with an OMRON K3AS Linear Sensor Controller connected.
Model Z500-MC10E Z500-MC15E
Item Input/Output Type NPN PNP
Perfor-
mance
specifica-
tions
Number of Sensors that can be mounted
2
Number of scenes 16
Light intensity tracking
function
Automatic (The light intensity tracking range can be specified)/Fixed (Select from 31 stages)/
Multiple (The light intensity range can be specified)
Measurement item Select from the following 8 types:
Height, Step: 2 pts, Step: 3 pts, Edge position, With, Edge center, Peak/Bottom, Define
Region specification Region specification of line beam and displacement direction is possible.
Number of data to be stored
2048 points max.
Trigger function Free/External 1/External 2/Auto
Results output
Terminal block 11 input points: TRIGGER, LD-OFF, RESET, D10 to D17
21 output points: DO0 to DO19, GATE
Monitor interface 1CH (for pin jack or overscan monitor)
Analog output resolution The full scale for output can be divided into a maximum of 40,000 gradations.
Resolution (See note): 0.25 mV (±5 V), 0.4 µA (4 to 20 mA)
General
specifica-
tions
Power supply voltage 21.6 to 26.4 VDC
Current consumption 1 A max. (with 2 Sensors connected)
Insulation resistance 20 M min. between all DC external terminals and GR terminal (100 VDC megger)
(with internal surge absorber removed)
Dielectric strength 1000 VAC, 50/60 Hz between all DC external terminals and GR terminal
(with internal surge absorber removed)
Leakage current 10 mA max.
Noise resistance 1500 Vp-p; pulse width: 0.1 µs/1 µs, Rising edge: 1-ns pulse
Vibration resistance 10 to 150 Hz (double amplitude: 0.1 mm) for 8 min. each in X, Y, and Z directions
Shock resistance 200 m/s2, 3 times each in 6 directions
Ambient temperature Operating: 0 to +50 °C, Storage: -15 to +60 °C (with no icing or condensation)
Ambient humidity Operating and storage: 35 to 85 %RH (with no condensation)
Ambient environment No corrosive gases
Ground Ground the Z500's ground terminal to less than 100 .
Degree of protection IEC IP20 (in-panel)
Material Unit: ABS
Weight (including packaging)
Approx. 1300 g (Unit: Approx. 700 g)
Accessories 2 manuals, 1 resistor (250 , 1/2 W)
J
u
d
gmen
t
ou
t
pu
t
RS-232C output
Terminal block output
Measurement value output (measurement value)
RS-232C output
Analog output
F502-EN2-04.book Seite 50 Dienstag, 26. Juli 2005 5:48 17
B-51Z500
Z500
Monitor
Laser Safety
The Z500-SW2T Sensor Head is a Class 2 Laser Product ac-
cording to EN60825-1 (IEC60825-1) and Class II Laser Prod-
uct according to FDA (21 CFR1040.10) (see note). The
Z500-SW6 and Z500-SW17 Sensor Heads are Class 3B and
Class IIIB Laser Products, respectively. The Z500 Series is
meant to be built into final system equipment. Pay special at-
tention to the following precautions for the safe use of the
product:
Note: Europe: Class 2 and Class 3B of EN60825-1: 1994 = IEC60825-1: 1993
U.S.A.: Class II and Class IIIB of FDA (21 CFR1040.10)
(1)Use this product as specified in the operation manual. Oth-
erwise, you may be exposed to hazardous laser radiation.
(2)The Z500 series radiates laser beams in the visible light
range. Do not expose your eyes directly to the laser radi-
ation. Ensure that the laser beam path is terminated dur-
ing use. If a mirror or shiny surface is positioned in the
laser beam path, ensure that the reflected beam path is
also terminated. If the Unit must be used without terminat-
ing the laser beam path, position the laser beam path so
that it is not at eye level.
(3)To avoid exposure to hazardous laser radiation, do not dis-
place nor remove the protective housing during operation,
maintenance, and any other servicing.
(4)The user should return the product to OMRON for all repair
and servicing.
(5)As for countries other than those of Europe and the U.S.A.,
observe the regulations and standards specified by each
country.
Dimensions (Unit: mm)
Monitor Liquid Crystal Monitor
Item Model F150-M05L
Panel size 5.5 inches
Panel type TFT color liquid crystal
Resolution 320 × 240 dots
Input signal NTSC composite video (1.0 V/75 )
Power supply voltage 20.4 to 26.4 VDC
Current consumption Approx. 700 mA
Ambient temperature Operating: 0 to +50 °C, Storage: -25 to +65 °C (with no icing or condensation)
Ambient humidity Operating and storage: 35 to 85 %RH (with no condensation)
Weight (including packaging) Approx. 870 g (Unit: Approx. 610 g)
Accessories Operation manual, 4 mounting brackets
Z500-SW2T
Z500-SW6/Z500-SW17
Wavelength 650 nm 658 nm
Maximum pulse duration
10 ms 17.5 ms
Cycle 0.5 to 10 ms 0.5 to 25 ms
Peak power 1 mW max. 15 mW max.
Class 2 3B
Sensor
Z500-SW2T
17 dia.
4
Operation indicators
Optical axis
Reference plane
Receiver
26.4
12
3
Emitter
Monitor
Light
Two, 4.5-dia. holes
Emission axis
Reference plane
Reception axis
Connector
18.1 20
20 65
6.5 20
4
12
25
˚
56.5
3.5
32
20
16
4
(4)
65
57
Vinyl-insulated round cable 6.8 dia.
Standard length: 2 m
Measurement center
2
F502-EN2-04.book Seite 51 Dienstag, 26. Juli 2005 5:48 17
B-52 Displacement sensors / Width-measuring Sensors
Z500-SW6
Sensor
Z500-SW17
Liquid Crystal Monitor
F150-M05L
Controller
Z500-MC10E/MC15E
Console
Z300-KP
36
92.1
46
22.7
82
Operation indicators
(5)
Connector
18
Reference plane
Optical axis
Emitter
Receiver
80
50
72
70
69 25
45.1
16.9
100
Vinyl-insulated round cable 6.8 dia.
Standard length: 0.5 m
4
16.4
25
9.5
Emission axis
Reception axis
Reference plane
30˚
Measurement center
Three, 4.5-dia. holes
821 dia.
(15)
9.5
13 dia.
36
92.1
46
22.7
82
Operation indicators
(5)
Connector
18
Reference plane
Optical axis
Emitter
Receiver
80
100
72
70
69 25
45.1
16.9
100
Vinyl-insulated round cable 6.8 dia.
Standard length: 2 m
4
16.4
25
9.5
Emission axis
Reception axis
Reference plane
25˚
Measurement center
Three, 4.5-dia. holes
821 dia.
(15)
9.5
13 dia.
POWER
SYNC
143
(145)
(155)
132
50 min.
(46 max.)
Mounting plate thickness:
1.6 to 4.8
(100)
185
(5.5)
174
Mounting bracket
Monitor Cable (F150-VM)
Tolerance: ±1 mm
The dimensions in parentheses are
provided for reference.
42.2
133.5
175.5
Panel opening dimensions
+0.5 mm
0
+0.5 mm
0
195
Indicators
10
10
185
Four, 4.5-dia. holes
Connector
4.1
(66.5)
8
110
100
90
12.5 dia.
142.6
50
500
MULTI DIMENSION CONTROLER
ESC TRIG
SHIFT
CONSOLE
ENT
50
23
2000 510
131.5
87
12.5
12 dia.
Z300-KP
In the interest of product improvement, specifications are subject to change without notice.
ALL DIMENSIONS SHOWN ARE IN MILLIMETERS.
To convert millimeters into inches, multiply by 0.03937. To convert grams into ounces, multiply by 0.03527.
Cat. No. Q02E-EN-02
F502-EN2-04.book Seite 52 Dienstag, 26. Juli 2005 5:48 17