PicoScope 2204A PicoScope 2205A PicoScope 2206A PicoScope 2207A PicoScope 2208A
FUNCTION GENERATOR
Standard output signals Sine, square, triangle, DC voltage, ramp, sinc, Gaussian, half-sine
Pseudorandom output signals None White noise, PRBS
Standard signal frequency DC to 100 kHz DC to 1 MHz
Sweep modes Up, down, dual with selectable start/stop frequencies and increments
Triggering - Free-run or up to 1 billion waveform cycles or frequency sweeps.
Triggered from scope trigger or manually.
Output frequency accuracy ±100 ppm ±50 ppm
Output frequency resolution < 0.01 Hz
Output voltage range ±2 V
Output adjustments Any amplitude and offset within ±2 V range
Amplitude flatness (typical) < 1 dB to 100 kHz < 0.5 dB to 1 MHz
DC accuracy ±1% of full scale
SFDR (typical) > 55 dB at 1 kHz full-scale sine wave > 60 dB at 10 kHz full-scale sine wave
Output characteristics Front panel BNC, 600 Ω output impedance
Overvoltage protection ±10 V
ARBITRARY WAVEFORM GENERATOR
Update rate 2 MS/s 20 MS/s
Buffer size 4 kS 8 kS
Resolution 12 bits
Bandwidth > 100 kHz > 1 MHz
Rise time (10% to 90%) < 2 µs < 120 ns
SPECTRUM ANALYZER
Frequency range DC to 10 MHz DC to 25 MHz DC to 50 MHz DC to 100 MHz DC to 200 MHz
Display modes Magnitude, average, peak hold
Windowing functions Rectangular, Gaussian, triangular, Blackman, Blackman-Harris, Hamming, Hann, flat-top
Number of FFT points Selectable from 128 to half available buffer memory in powers of 2
MATH CHANNELS
Functions −x, x+y, x−y, x*y, x/y, x^y, sqrt, exp, ln, log, abs, norm, sign, sin, cos, tan, arcsin, arccos, arctan,
sinh, cosh, tanh, freq, derivative, integral, min, max, average, peak, delay
Operands A, B (input channels), T (time), reference waveforms, constants, Pi
AUTOMATIC MEASUREMENTS
Scope mode AC RMS, true RMS, cycle time, DC average, duty cycle, falling rate, fall time, frequency, high pulse width,
low pulse width, maximum, minimum, peak to peak, rise time, rising rate.
Spectrum mode Frequency at peak, amplitude at peak, average amplitude at peak,
total power, THD %, THD dB, THD plus noise, SFDR, SINAD, SNR, IMD
Statistics Minimum, maximum, average and standard deviation
SERIAL DECODING
Protocols CAN, LIN, I²C, UART/RS-232, SPI, I²S, FlexRay
MASK LIMIT TESTING
Statistics Pass/fail, failure count, total count
DISPLAY
Interpolation Linear or sin(x)/x
Persistence modes Digital color, analog intensity, custom, or none
GENERAL
PC connectivity USB 2.0 (USB 1.1 and 3.0 compatible). USB cable included.
Power requirements Powered from USB port
Dimensions (including connectors) 142 x 92 x 19 mm
Weight < 0.2 kg (7 oz)
Temperature range Operating: 0 °C to 50 °C (20 °C to 30 °C for stated accuracy). Storage: –20 °C to +60 °C.
Humidity range Operating: 5% to 80% RH non-condensing. Storage: 5% to 95% RH non-condensing.
Safety approvals Designed to EN 61010-1:2010
Compliance RoHS, WEEE, and LVD compliant. Tested to meet EN61326-1:2006 and FCC Part 15 Subpart B.
Software included PicoScope 6, Windows and Linux SDK, example programs (C, Visual Basic, Excel VBA, LabVIEW)
PicoScope software PC requirements Microsoft Windows XP (SP3), Windows Vista, Windows 7 or Windows 8 (not Windows RT). 32- or 64-bit
Languages (manual) Chinese (simplified), English, French, German, Italian, Spanish
Languages (software interface) Chinese (simplified & traditional), Czech, Danish, Dutch, English, Finnish, French, German, Greek, Hungarian,
Italian, Japanese, Korean, Norwegian, Polish, Portuguese, Romanian, Russian, Spanish, Swedish, Turkish
Specifications continued...