Specification
POWERFACTOR(Typ.)
(Typ.)
(Typ.)
(Typ.)
PF>0.95/230VACPF>0.98/115VACatfullload
RC+/RC-:0~0.8V=poweron;4~10V=poweroffsinkcurrent<4~10mA
REMOTECONTROL
ACINPUTVOLTAGE
ACINPUTCURRENT
INRUSHCURRENT
SETUP,RISETIME
HOLD-UP TIME
LEAKAGECURRENT
DCOUTPUT ADJ.RANGE
OVERLOADPROTECTION
OVERVOLTAGEPROTECTION
OVERTEMPERATUREPROTECTION
POWERGOOD/FAIL SIGNAL
90~264VAC,47~440Hz/127~370VDC.
Maximuminputcurrent3.5A at115VAC,60Hzor
1.6A at230VAC,60Hzwith100%outputload.
Inrushcurrentislessthan25A at115VACor
lessthan40A at230VACundercoldstart
conditions.Limitingprovidedbyinternal
thermistors.
1000ms,20ms/230VAC atfullload
3000ms,20ms/115VACatfullload
16ms/230VACatfullload
16ms/115VACatfullload
Leakagecurrentislessthan180 A at264VACfor
earthleakagecurrent
Leakagecurrentislessthan100 A at264VACfor
patientleakagecurrent
DCoutputvoltage(orCH1ofmultipleoutputmodels)
canbeadjustedbetween-5%~+10%ratedoutput
voltagebypotentialmeter.
Fullyprotectedagainstshortcircuitandoutputoverload.
Thehiccuptypeprotectionwillbeactivatedat120~160%
ratedloadandrecoversautomaticallyafterfaultcondition
isremoved.
Providedonoutputchannel1onlyat115%~135%
ratedoutputvoltage.
Outputwillbeshutdownwhenthisprotectionisactivated.
Whenthetemperatureof TSW1whichdetect onheatsinkof
powertransistorreaches95 Thisprotectionisactivated.
Thenoutputwillbeshutdownandrecoversautomatically
aftertemperaturegoesdown.
*MPS-200-3.3doesnothavethisfunction.
TTL logichighforpowergoodand TTL lowfor
powerfail.Whentheoutputvoltagereaches90%of
ratedvalue,a+5V TTL signalwillbesentoutwitha
10~500msdelay; Atleast1msbeforetheoutput
voltagegoesbelow90%oftheratedvalue,the TTL
signalwillbeturnedoff.
ComplianceLevel
EN55011,ClassB
EN55022,ClassB
EN61000-3-2
EN61000-3-3
ComplianceLevel
EN61000-4-2,
EN61000-4-2,
EN61000-4-3,
EN61000-4-4,
EN61000-4-5,
EN61000-4-6,
EN61000-4-8,
EN61000-4-11,
ENV50204,
Level3,8KV
Level2,4KV
Level2,3V/m
Level3,10V/m
Level2,1KV/5KHz
Level3,2KV/5KHz
Level4,2KV/Line-Line
4KV/Line-Earth
Level2,3Vrms/m
Level3,10Vrms/m
Level2,3A/m
Level3,10A/m
Compliance
Level2,3V/m,900MHz
Level3,10A/m,900MHz
1~4OutputMedical Type 200WMedical series
Universal ACinput/Fullrange
Lowleakagecurrent<180 A
Protections:Shortcircuit/Overload/Overvoltage/Overtemperature
Freeairconvectionfor140Wandforcedairconvectionfor200W
ANSI/AAMIES60601-1medicalsafetyapproved
Withpowergoodandfailsignaloutput
Built-in remoteON-OFFcontrol
Fixedswitchingfrequencyat100KHz
3yearswarranty
Built-inremotesensefunction
Features
TEMP.COEFFICIENT
VIBRATION
SAFETY STANDARDS
:
WITHSTANDVOLTAGE
ISOLATIONRESISTANCE
EMICOMPLIANCE
EMSCOMPLIANCE
MTBF
DIMENSION(L*W*H)
PACKING
WORKINGTEMP.
WORKINGHUMIDITY
STORAGETEMP.,HUMIDITY
Wholeseriescanoperatefrom-20~70 .Pleaserefer
tothederatingcurves.
20~90%RHnon-condensing.
-40~+85 ,10~90%RH
2Gofacceleration,vibratingfrequencyadjustfrom10Hz~500Hzwithina
10-minutecycle.6testingcycles(60minutes)eachalongX, Y,Zaxes.
Medical , TUVEN60601-1
Commercial: AlsodesignrefertoUL60950-1, TUVEN60950-1
4000VACbetweeninputandoutput
1500VACbetweeninput andF.G.
1500VACbetweenoutputandF.G.
EMISpecifications
EMSSpecification
ESDair
ESDcontact
RFfieldsusceptibility
EFT(ElectricalFast Transient)/Burst
Lightning/Surge
ConductedRFsusceptibility
Magneticfieldimmunity
Voltagedip,interruption
Digitalphonecarrierimmunity
262,100hoursmin.atfullloadand25 ofambienttemperature,
calculatedperMIL-HDBK-217F.
177.8x107.2x35.5mmor7"x4.22"x1.4"
0.66Kg;24pcs/16.8Kg/0.99CUFT
, approved
0.04%/ onalloutputsatfullloadbetween0~50 ofambienttemperature.
>100MOhmsforI/P-O/P,I/P-FG,O/P-FGbyusing500VDCtestvoltage.
Conducted&Radiation
Harmonicdistortion
Voltageflicker
ANSI/AAMIES60601- IEC60601-11
FileName:200WMEDICAL-SPEC2012-02-20