SPECIFICATION SHEET FOR CO SENSOR TYPE CO/MF-500
PERFORMANCE CHARACTERISTICS
REV.: 2/2008 Page 1 of 2
CROSS-SENSITIVITY DATA
Interfering Gas Concentration Reading
H2 100 ppm < 40 ppm
H2S 10 ppm ~ 0 ppm
Performance data conditions:
20 °C, 50% RH and 1013 mbar
APPLICATIONS
Safety and Environmental Control
For Portable Gas Detectors
PHYSICAL CHARACTERISTICS
BOTTOM VIEW
Miniature-Size Outline Dimensions
Reference
Sensing
Counter
13.5 mm
PCD
18 mm
90°
SIDE VIEW
20 mm
5 mm
Pin Projection 1.5 mm Pins
on 13.5 mm PCD
Nominal Range 0 – 500 ppm
Maximum Overload 2000 ppm
Expected Operation Life 2 years in air
Output Signal 70 ± 20 nA/ppm
Inboard Filter To remove H2S and
SO2
Resolution 1 ppm
Temperature Range - 20 °C to + 50 °C
Pressure Range Atmospheric ± 10 %
Pressure Coefficient No data
t90 Response Time < 25 sec
Relative Humidity Range 15 % to 90 % R.H.
non-condensing
Typical Baseline Range (pure
air, 20°C)
- 2 ppm to +3 ppm
Maximum Zero Shift (+20°C
to +40 °C)
9 ppm equivalent
Expected Long Term Output
Drift
< 2 % signal
loss/month
Recommended Load Resistor 10 Ohm
Bias Voltage Not required
Repeatability < 2 % of signal
Output Linearity Linear
Weight ~ 5.4 g
Position Sensitivity None
Storage Life Six months in
container
Recommended Storage
Temperature
5 °C – 20 °C
Warranty Period 12 months from date
of dispatch
Phone: +41 43 311 72 00 MEMBRAPOR AG
Fax : +41 43 311 72 01 Rautistrasse 164
Email: info@membrapor.ch CH-8048 Zürich
www.membrapor.ch Switzerland
MEMBRAPOR
SPECIFICATION SHEET FOR CO SENSOR TYPE CO/MF-500
TEMPERATURE DEPENDENCE
The output of an electrochemical sensor varies with temperature. The graphs below show the variation in output with
temperature for this type of sensor. The results are shown in the graphs as a mean for a batch of sensors, along with
confidence intervals corresponding to ±3 times the standard deviation. The sensitivity dependence is expressed as a
percentage of the signal at 20 °C.
Sensitivity Temperature Dependence
50%
60%
70%
80%
90%
100%
110%
120%
-20-100 10 20304050
TemperatureC)
Output (referenced to 20 °C)
Mean
+3SD
-3SD
The data contained in this document is for guidance only. Membrapor AG accepts no liability for any consequential losses, injury or damage resulting
from the use of this document or the information contained within it. The data is given for guidance only. Customers should test under their own
conditions, to ensure that the sensors are suitable for their own requirements.
REV.: 2/2008 Page 2 of 2
Phone: +41 43 311 72 00 MEMBRAPOR AG
Fax : +41 43 311 72 01 Rautistrasse 164
Email: info@membrapor.ch CH-8048 Zürich
www.membrapor.ch Switzerland