Features & Benefits
60 MHz, 100 MHz, and
200 MHz Bandwidths
Sample Rates up to 2 GS/s
2 or 4 Channels
Color or Monochrome
LCD Display
CompactFlash®Mass
Storage Option with
TDS2MEM Module
Autoset Menu with
Waveform Selection
Probe Check Wizard to
Ensure Correct Probe Usage
Context-sensitive Help
Dual Time Base
Advanced Triggers,
Including Pulse Width
Trigger and Line-Selectable
Video Trigger
11 Automatic Measurements
Multi-language User Interface
Waveform and
Setup Memories
FFT Standard on All Models
OpenChoiceSolutions
Speed Documentation and
Analysis of Measurement
Results
TDS2CMAX Communication
Module
TDS2MEM Storage Memory
and Communication Module
TDSPCS1 OpenChoice
Software
–WaveStarSoftware
Integration with
Third-party Software
Applications
Design and Debug
Education and Training
Manufacturing Test
and Quality Control
Service and Repair
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
TDS1000 and TDS2000 Series Oscilloscopes.
Colorful Performance at a Black and White Price.
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000
1
The TDS1000 Series and TDS2000 Series
digital storage oscilloscopes deliver an
unbeatable combination of superior
performance, unmatched ease-of-use,
and affordability in an ultra lightweight,
portable package.
Affordable Digital Precision
With up to 200 MHz bandwidth and 2 GS/s
maximum sample rate, no other color digital
storage oscilloscope offers as much band-
width and sample rate for the price. The
TDS1000 and TDS2000 Series oscilloscopes
provide accurate real-time acquisition up to
their full bandwidth, advanced triggers to
isolate signals of interest, and 11 standard
automatic measurements on all models.
Their Fast Fourier Transform (FFT) math
function allows the user to analyze, charac-
terize and troubleshoot circuits by viewing
frequency and signal strength (standard
on all models).
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000
2
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
Ultra-fast Setup and Use
The simple user interface with classic,
analog-style controls makes these instruments
easy to use, reducing learning time and
increasing efficiency. Innovative features,
such as the autoset menu, probe check
wizard, context-sensitive help menu and
color LCD display (TDS2000 Series) optimize
instrument setup and operation.
Simple, Speedy
Documentation
and Analysis
OpenChoicesolutions deliver simple,
seamless integration between the oscillo-
scope and the personal computer, providing
you with multiple choices to easily document
and analyze your measurement results.
Choose from optional communication modules,
CompactFlash®mass storage capability,
OpenChoice software and integration
with third-party software.
Automatically detects sine waves, square
waves and video signals, with readouts of
relevant measurements and additional
user-selectable views of the signal.
Pulse-width and field- and line-selectable
video triggers make the oscilloscope ideal
for a wide range of applications.
Unparalleled measurement accuracy with
up to 200 MHz bandwidth and 2 GS/s
maximum sample rate.
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
Characteristics
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000 3
Easily document and analyze measurement
results using OpenChoicesoftware. Quickly transfer waveforms and measure-
ment data to an external PC or between
oscilloscopes.
Precisely analyze, characterize and
troubleshoot circuits – at the push of
a button with the autoset menu.
TDS1000 and TDS2000 Series Electrical Characteristics
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
Display (1/4 VGA LCD) Mono Mono Color Color Color Color Color
Bandwidth*160 MHz 100 MHz 60 MHz 100 MHz 100 MHz 200 MHz 200 MHz
Channels 2 2 2 2 4 2 4
External Trigger Input Yes on all models
Sample Rate 1.0 GS/s 1.0 GS/s 1.0 GS/s 1.0 GS/s 1.0 GS/s 2.0 GS/s 2.0 GS/s
on each channel
Record Length 2.5 K points on all models
Vertical Resolution 8-bits
Vertical Sensitivity 2 mV to 5 V/div on all models with calibrated fine adjustment
DC Vertical Accuracy ±3% on all models
Vertical Zoom Vertically expand or compress a live or stopped waveform
Max Input Voltage 300 VRMS CAT II; derated a t 20 dB/decade above 100 kHz to 13 Vp-p AC at 3 MHz and above
Position Range 2 mV to 200 mV/div ±2 V; >200 mV to 5 V/div ±50 V
BW Limit 20 MHz for all models
Input Coupling AC, DC,GND on all models
Input Impedance 1 Min parallel with 20 pF
Time Base Range 5 ns to 5 ns to 5 ns to 5 ns to 5 ns to 2.5 ns to 2.5 ns to
50 sec/div 50 sec/div 50 sec/div 50 sec/div 50 sec/div 50 sec/div 50 sec/div
Time Base Accuracy 50 ppm
Horizontal Zoom Horizontally expand or compress a live or stopped waveform
*1Bandwidth is 20 MHz at 2 mV/div, all models.
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000
4
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
Acquisition Modes
Peak Detect High frequency and random glitch
capture. Captures glitches as narrow as 12 ns
typical using acquisition hardware at all time/div
settings from 5 µs/div to 50 s/div.
Sample Sample data only.
Average Waveform averaged, selectable: 4, 16,
64, 128.
Single Sequence Use the Single Sequence
button to capture a single triggered acquisition
sequence at a time.
Trigger System (Main Only)
Trigger Modes Auto,Normal, Single Sequence.
Trigger Types
Edge (rising or falling) Conventional level-driven
trigger. Positive or negative slope on any channel.
Coupling selections:AC,DC,Noise Reject, HF Reject,
LF Reject.
Video Trigger on all lines or individual line,
odd/even or all fields from composite video, or
broadcast standards (NTSC, PAL,SECAM).
Pulse Width (or Glitch) Trigger on a pulse width
less than, greater than, equal to, or not equal to a
selectable time limit ranging from 33 ns to 10 sec.
Trigger Source
2-channel models CH1,CH2, Ext, Ext/5,AC Line.
4-channel models CH1, CH2, CH3, CH4, Ext,
Ext/5,AC Line.
Trigger View
Displays trigger signal while trigger view button
is depressed.
Trigger Signal
Frequency Readout
Provides a frequency readout of the trigger source.
Cursors
Types Voltage,Time.
Measurements []T, 1/[]T (frequency), []V.
Measurement System
Automatic Waveform Measurements Period,
Frequency ,+Width, –Width,Rise Time,Fall Time,
Max, Min, Peak-to-Peak, Mean,Cycle RMS.
Waveform Processing
Operators Add, Subtract, FFT.
FFT Windows: Hanning,Flat Top, Rectangular;
2048 sample points.
Sources 2-channel models: CH1 - CH2,
CH2 - CH1, CH1 + CH2.
4-channel models: CH1 - CH2, CH2 - CH1,
CH3 - CH4, CH4 - CH3, CH1 + CH2, CH3 + CH4.
Autoset Menu Single-button, automatic setup
of all channels for vertical, horizontal and trigger
systems, with undo autoset.
Display Characteristics
Display
Monochrome models: 1/4 VGA,backlit passive LCD
with adjustable multi-level contrast and inverse video
selectable from front panel.
Color models: 1/4 VGA,passive color LCD with
color on black background with adjustable
multi-level contrast.
Interpolation Sin(x)/x.
Display Types Dots, vectors.
Persistence Off, 1 sec,2 sec, 5 sec,infinite.
Format YT and XY.
Autoset Menu for
Multiple Signal Types
Signal Autoset Menu
Type Choices
Square Wave Single-cyc le,Multi-cyc le,
Rising or Falling Edge
Sine Wave Single-cyc le,Multi-cyc le,
FFT Spectrum
Video (NTSC, Field:All, Odd or Even
PAL, SECAM) Line:All or Selectable
Line Number
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000 5
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
I/O Interface
I/O interface TDS2CMAX Communication Module TDS2MEM Storage Memory
and Communication Module
Compatibility TDS200,TDS1000,TDS2000 Series Oscilloscopes TDS1000, TDS2000 Series Oscilloscopes
Printer Port Centronics-type Parallel
Printer Capability Graphics File Formats – TIFF,PCX (PC Paint Brush), BMP (Microsoft Windows), EPS (Encapsula ted Postscript) and RLE
(Requires Module) Printer Formats – Bubble Jet, DPU-411, DPU-412, DPU-3445,Thinkjet, Deskjet, Laser Jet,
Epson Dot (9- or 24-pin)*2,Epson C60*2,Epson C80
Layout – Landscape and Portrait
RS-232 Programmability Full talk/listen modes. Control of all modes, settings and measurements.
Baud rate up to 19,200. 9-pin DTE RS-232 Cable (012-1651-00) included
GPIB (IEEE std. 488-1987) Full talk/listen modes. Control of all modes, Not Available
Programmability settings and measurements.
Mass Storage Not Available Type 1 CompactFlash memor y card supplied with Module*3
CompactFlash®Memory Accepts any Type 1 CompactFlash card, up to and including 1 GB
Built-in Clock/Calendar
Includes CompactFlash to USB Memory Card Reader
for use with Personal Computer
OpenChoicePC TDSPCS1 OpenChoice PC Communications Software included with each Module
Communications Seamless connection from oscilloscope to PC through GPIB and RS-232
Software Transfer and save settings, waveforms, measurements, and screen images
Includes a Windo ws desktop data transfer a pplication in addition to convenient Microsoft Word and Excel Toolbar Add-ins
*2Requires instrument firmware version 2.12/4.12 or higher to support Epson C60/C80 printer formats.
*3Supplied Type 1 CompactFlash memory card stora ge capacity 8 MB or grea ter.
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000
6
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
Environmental and Safety
Temperature
Operating: 0 ºC to +50 ºC.
Nonoperating: –40 ºC to +71 ºC.
Humidity
Operating and Nonoperating: up to 90% RH
at or below +30 ºC.
Operating: up to 60% RH up to 50 ºC.
Nonoperating: up to 60% RH up to 55 ºC.
Altitude Operating and Nonoperating: up to
3,000 m.
Electromagnetic Compatibility Meets Directive
89/336/EEC, amended by 93/68/EEC, meets
or exceeds EN55011 Class A Radiated and
Conducted Emissions; FCC 47 CFR, Part 15,
Subpart B, Class A; Australian EMC Framework,
demonstrated per Emission Standard AS/NZS
2064; Russian GOST EMC regulations.
Safety UL3111-1, CSA1010.1, IEC61010-1,
EN61010-1.
Physical Characteristics
INSTRUMENT
Dimensions mm in.
Width 323.8 12.75
Height 151.4 5.96
Depth 124.5 4.90
Weight kg lbs.
Instrument only 2.0 4.4
with accessories 2.2 4.9
INSTRUMENT SHIPPING
Package Dimensions mm in.
Width 476.2 18.75
Height 266.7 10.50
Depth 228.6 9.00
RM2000 RACKMOUNT
Dimensions mm in.
Width 482.6 19.00
Height 177.8 7.00
Depth 108.0 4.25
Ordering Information
TDS1002, TDS1012, TDS2002,
TDS2012, TDS2014, TDS2022,
TDS2024
Standard Accessories
Probes – P2220 200 MHz 10X - 1X Switchable
Passive Probes (one per channel).
Documentation – User manual.
Power cord.
NIM/NIST-Traceable Certificate of Calibration.
Please specify the power plug and language
options when ordering.
Recommended Accessories
TDS2CMAX – Communication Module (GPIB,
RS-232, Centronics-type Printer Port) includes:
TDSPCS1 OpenChoicePC Communication
Software, RS-232 cable (012-1651-00),
Programmers manual.
TDS2MEM – Storage Memory and Communication
Module (RS-232, Centronics-type Printer Port)
includes:Type 1 CompactFlash®memory card
(156-9413-00), CompactFlash to USB Memor y
Card Reader (119-6827-00),TDSPCS1
OpenChoice PC Communication Software,
RS-232 cable (012-1651-00), Programmers
manual.
Nonvolatile Storage
Nonvolatile Standard Optional TDS2MEM
Storage Storage Memory and
Communication Module*4
Reference Waveform Display Two 2500 point reference waveforms Same as instrument only
Waveform Stora ge 2 Channel Models 4 Channel Models 96 or more reference waveforms per 8 MB
Two 2500 point reference waveforms Four 2500 point reference waveforms
Setups 10 front panel setups 4000 or more front panel setups per 8 MB
Screen Images N/A 128 or more screen images per 8 MB
(the number of images
depends on file format selected)
Save All N/A 12 or more Save All operations per 8 MB.
A single Save All operation creates
2 to 9 files (setup, ima ge, plus one file
for each displayed waveform)
*4Supplied with Type 1 CompactFlash®memory card, storage capacity 8 MB or greater.
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000 7
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
CompactFlash to USB Memory Card Reader –
(119-6827-00).
TDSPCS1 OpenChoice PC Communication
Software – A collection of programs that enable
fast and easy transfer communication between
MS Windows PCs and TDS3000B,TDS2000,
TDS1000, and TDS200 Series oscilloscopes;
included with TDS2CMAX and TDS2MEM
communication modules.
Minimum System Requirements:
500 MHz Pentium equivalent or greater
128 MB
MS Windows 98 SE, XP Professional,
Me, or 2000
MS Office 2000 or XP (for TDS toolbars
only) – Excel 2000 or 2002; Word 2000
or 2002
Keyboard and mouse
LAN, GPIB or serial connector
WaveStarSoftware for Oscilloscopes
(WSTRO) – Windows 98/2000/ME/NT 4.0
application for waveform capture, analysis,
documentation, and control from your PC. Provides
enhanced oscilloscope data measurement,
analysis, remote-setup, and charting features.
Connects to a wide variety of Tektronix
oscilloscopes over GPIB, RS-232, and
Ethernet connections.
AD007 – LAN/WAN GPIB converter.
AC2100 – Soft case for carrying instrument.
HCTEK321 – Hard plastic case for carrying
instrument (requires AC2100).
RM2000 – Rackmount kit.
Service Manual (TDS1000 and TDS2000 Series
Digital Storage Oscilloscopes) – English Only
(071-1076-00).
Programmer Manual (TDS200,TDS1000
and TDS2000 Series Digital Storage
Oscilloscopes) – English only (071-1075-01).
Operator Training Kit – Extensive instructions
and step-by-step lab exercises provide education
about the operation of the TDS1000 and TDS2000
Series oscilloscopes. Kit includes a self-paced
CD-ROM-based manual and signal source board
(TNGTDS01). Optional hard copy manual is
available for order separately.
Recommended Probes
P2220 – 10X - 1X switchable passive probe
(200 MHz when 10X is selected).
P6015A – 1000X high voltage passive probe
(75 MHz).
P5100 – 100X high voltage passive probe
(250 MHz).
P5200 – High voltage active differential probe
(25 MHz).
P6021 – 60 MHz AC current probe.
P6022 – 120 MHz AC current probe.
A621 – 2000 A AC current probe/BNC.
A622 – 100 A AC/DC current probe/BNC.
AM503S – AC/DC current probe system.
Power Plug Options
Opt. A0 North America power.
Opt. A1 Universal EURO power.
Opt. A2 United Kingdom power.
Opt. A3 Australia power.
Opt. A5 Switzerland power.
Opt. A99 No power cord or AC adapter.
Opt. AC China power.
Accessory Cables
GPIB, 1 m (3.3 ft.) – Order 012-0991-01.
GPIB, 2 m (6.6 ft.) – Order 012-0991-00.
RS-232, 9-Pin female to 25-Pin male,
4.6 m (15 ft.),for Modems – Order
012-1241-00.
RS-232, 9-Pin female to 9-Pin female,null
modem, for computers – Order 012-1651-00.
RS-232, 9-Pin female to 25-Pin female,null
modem, for computers – Order 012-1380-00.
RS-232, 9-Pin female to 25-Pin male,null
modem, for printers – Order 012-1298-00.
Centronics, 25-Pin male to 36-Pin Centronics,
2.4 m (8 ft.),for Parallel Printer Interfaces –
Order 012-1214-00.
International User Manual
(TDS1000 and TDS2000
Series Digital Storage
Oscilloscopes, includes
TDS2CMAX*1user information)
Opt. L0 – English (071-1064-00).
Opt. L1 – French (071-1065-00).
Opt. L2 – Italian (071-1066-00).
Opt. L3 – German (071-1067-00).
Opt. L4 – Spanish (071-1068-00).
Opt. L5 – Japanese (071-1069-00).
Opt. L6 – Portuguese (071-1070-00).
Opt. L7 – Simplified Chinese (071-1071-00).
Opt. L8 – Traditional Chinese (071-1072-00).
Opt. L9 – Korean (071-1073-00).
Opt. LR – Russian (071-1074-00).
Translated front panel overlays included with their
respective user manuals.
Warranty Information
Three year warranty covering all labor and parts,
excluding probes and accessories.
*1The user information for the TDS2CMA communication
module that appears in this manual also applies to the
TDS2CMAX communication module.
Tektronix probes are expressly designed
for your oscilloscope, with identical quality
standards and built-in compatibility for
optimum performance.
Digital Storage Oscilloscopes
TDS1002 TDS1012 TDS2002 TDS2012 TDS2014 TDS2022 TDS2024
Our most up-to-date product information is available at:
www.tektronix.com
Copyright © 2005,Tektronix, Inc.All rights reserved.Tektronix products are covered
by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material.Specification and price change
privileges reserved. TEKTRONIX and TEK are registered trademarks of Tektronix,
Inc. All other trade names referenced are the ser vice marks, trademarks or regis-
tered trademarks of their respective companies.
1/05 DV/WOW 40W-15314-4
Digital Storage Oscilloscopes TDS1000 and TDS2000 Series www.tektronix.com/tds2000
8
Product(s) are manufactured
in ISO registered facilities.
Contact Tektronix:
ASEAN / Australasia / Pakistan (65) 6356 3900
Austria +41 52 675 3777
Balkan, Israel, South Africa and other ISE Countries +41 52 675 3777
Belgium 07 81 60166
Brazil & South America 55 (11) 3741-8360
Canada 1 (800) 661-5625
Central East Europe, Ukraine and Baltics +41 52 675 3777
Central Europe & Greece +41 52 675 3777
Denmark 80 88 1401
Finland +41 52 675 3777
France & North Africa +33 (0) 1 69 81 81
Germany +49 (221) 94 77 400
Hong Kong (852) 2585-6688
India (91) 80-22275577
Italy +39 (02) 25086 1
Japan 81 (3) 6714-3010
Luxembourg +44 (0) 1344 392400
Mexico, Central America & Caribbean 52 (55) 56666-333
Middle East, Asia and North Africa +41 52 675 3777
The Netherlands 090 02 021797
Norway 800 16098
People’s Republic of China 86 (10) 6235 1230
Poland +41 52 675 3777
Portugal 80 08 12370
Republic of Korea 82 (2) 528-5299
Russia, CIS & The Baltics 7 095 775 1064
South Africa +27 11 254 8360
Spain (+34) 901 988 054
Sweden 020 08 80371
Switzerland +41 52 675 3777
Taiwan 886 (2) 2722-9622
United Kingdom & Eire +44 (0) 1344 392400
USA 1 (800) 426-2200
USA (Export Sales) 1 (503) 627-1916
For other areas contact Tektronix, Inc. at: 1 (503) 627-7111
Last Updated 3 November 2004
A Critical Component of the
Complete Measurement Solution.
The AFG300 Series arbitrar y function
generator pairs with the TDS3000B,
TDS2000 and TDS1000 Series digital
oscilloscopes t o deliver the two
elements of a complete measure-
ment solution – stimulus and
acquisition.This instrument combines
the capabilities of a function generator with the power of an arbitrary wa veform generator, offering the
performance needed to accurately verify, validate and characterize designs with ease and confidence at a
price you can afford.
Tektronix Support Completes the
Solution. We know you depend on
Tektronix instrument solutions when
you make and meet critical commit-
ments. So we make and meet a sup-
port commitment you can depend on.
Anytime you need support, anywhere
in the world,Tektronix Support gives
you the lowest possible exposure to
inconvenience, delay or disruption
of operations.
– Unsurpassed technical expertise and
experience with 24-hour response to technical questions
– Interactive, online support to request assistance, check ser vice status or arrange for training
– Industry-leading, turn-around service time
– Credible, reliable support with demonstrated on-time deliver y
– 90-day unconditional service warranty
– No fine print, no exclusions, no surprises
– Global support in more than 50 countries
Depend on Tektronix. Visit www.tektronix.com/support