Technical Specifications
82100-0082
THURLBY THANDAR INSTRUMENTS
EL & EX Series
Compact bench DC power supplies
single, dual and triple output models
CYAN MAGENTA YELLOW BLACK
EX series
MAIN OUTPUTS
Voltage Range: 0V to 35V minimum (EX355, EX354D, EX354T)
0V to 75V or 0V to 150V minimum (EX752M)
Current Range: 0A to 2A or 0A to 4A minimum (EX752M)
0A to 4A minimum (EX354D, EX354T)
0A to 5A minimum (EX355)
Operating Mode: Constant voltage or constant current with automatic
cross-over and indication.
Voltage Setting: By coarse and fine controls.
Current Setting: By single logarithmic control.
Load regulation: <0.01% of max. O/P for a 90% load change.
Line regulation: <0.01% of max. O/P for a 10% line voltage change.
Output impedance: Typically <5mΩin constant voltage mode. Typically
>50kΩin constant current mode.
Ripple & Noise
(20MHz bandwidth): Typically <2mV rms, <10mV pk-pk, (CV mode).
Transient Response: <200µs to within 50mV of set level for 90% load change.
Temp. Coefficient: Typically <100ppm/oC.
Output Protection: Outputs will withstand forward voltages of up to 40V
(EX355, EX354D, EX354T) or 85V/170V (EX752M).
Over-voltage clamp operates above these levels.
Reverse protection by diode clamp for currents up to 3A.
Status Indication: Output ON lamps. Constant current mode lamps.
Output Switch: Electronic. Preset voltage and current displayed when
output is off.
Output Terminals: 4mm terminals on 19mm (0.75") pitch (see below).
Special safety terminals on EX752M.
METERS (main outputs)
Meter Types: Separate 3 digit meters for voltage and current with
14mm (0.56") LED displays. Reading rate 4/sec.
(4 digit voltmeter on EX752M).
Meter Resolutions: 100mV, 10mA.
Meter Accuracies: Voltage 0.3% ±1 digit. Current 0.6% ±1 digit.
LOGIC OUTPUT (EX354T only)
Voltage: 5V or 3.3V, selectable by front panel switch.
Voltage accuracy: ±2%
Current limit: 5A minimum
Load regulation: <0.5% for a 90% load change.
Line regulation: <0.1% for a 10% line voltage change.
Ripple & Noise
(20MHz bandwidth): Typically <2mV rms, <10mV pk-pk.
Output Protection: Output will withstand up to 7V forward voltage.
Diode clamp reverse protection for currents to 3A.
Output Terminals: 4mm terminals on 19mm (0.75") pitch (see below).
Status Indication: Current limit lamp.
EL series
MAIN OUTPUTS
Voltage Range: 0V to18V minimum (EL183)
0V to 30V minimum (EL301, EL302, EL302D, EL302T)
0 to 56V minimum (EL561)
Current Range: 0A to 1A minimum (EL301)
0A to 1.1A minimum (EL561)
0A to 2A minimum (EL302, EL302D, EL302T)
0A to 3.3A minimum (EL183)
Operating Mode: Constant voltage or constant current with automatic
cross-over and indication.
Voltage Setting: By coarse and fine controls.
Current Setting: By single logarithmic control.
Load regulation: <0.01% of max. O/P for a 90% load change.
Line regulation: <0.01% of max. O/P for a 10% line voltage change.
Output impedance: Typically <5mΩin constant voltage mode. Typically
>50kΩin constant current mode.
Ripple & Noise
(20MHz bandwidth): Typically <1mV rms, (CV mode).
Transient Response: <20µs to within 50mV of set level for 90% load change.
Temp. Coefficient: Typically <100ppm/oC.
Output Protection: Outputs will withstand forward voltages of up to 20V
above the rated output voltage. Reverse protection
by diode clamp for currents up to 3A.
Status Indication: Output ON lamps. Constant current mode lamps.
Output Switch: Electronic. Preset voltage and current displayed when
output is off.
Output Terminals: 4mm terminals on 19mm (0.75") pitch (see below).
METERS (main outputs)
Meter Types: Separate 3 digit meters for voltage and current with
14mm (0.56") LED displays. Reading rate 4/sec.
Meter Resolutions: 100mV, 10mA.
Meter Accuracies: Voltage 0.3% ±1 digit. Current 0.6% ±1 digit.
LOGIC OUTPUT (EL302T only)
Voltage: 5V or 3.3V, selectable by front panel switch.
Voltage accuracy: ±3%
Current limit: 1A minimum.
Load regulation: <0.5% for a 90% load change.
Line regulation: <0.1% for a 10% line voltage change.
Ripple & Noise
(20MHz bandwidth): Typically <1mV rms.
Output Protection: Output will withstand up to 16V forward voltage.
Diode clamp reverse protection for currents to 3A.
Output Terminals: 4mm terminals on 19mm (0.75") pitch (see below).
Status Indication: Current limit lamp.
Thurlby Thandar Instruments Ltd. operates a policy of continuous development and
reserves the right to alter specifications without prior notice.
GENERAL SPECIFICATIONS
MODEL SIZE (W X H X D) mm WEIGHT (kg) Input VA max.
EL301 140 x 160 x 295 3.4 85
EL183 140 x 160 x 295 4.4 160
EL302 140 x 160 x 295 4.4 160
EL561 140 x 160 x 295 4.4 160
EX355 140 x 160 x 320 3.0 350
EL302D 260 x 160 x 295 7.5 320
EL302T 260 x 160 x 295 7.5 320
EX354D 260 x 160 x 320 4.3 600
EX354T 260 x 160 x 320 4.3 600
EX752M 260 x 160 x 320 4.4 600
OUTPUT TERMINALS
Standard Type: 4mm "binding post" terminals suitable for plugs or wires.
EX752M only: As above but with fully "touch proof" construction.
Optional (all models): 4mm safety sockets, suitable for shrouded plugs.
POWER
AC Input: EL series: 230V or 115V ±10% AC, 50/60Hz.
EX355, EX752M: 99V to 262V AC, 50/60Hz.
EX354D, EX354T: 230V ±14% (115V to order).
Installation Category II.
Consumption: See table for maximum VA figures for each model.
ENVIRONMENTAL
Operating Range: +5oC to +40oC, 20% to 80% RH.
Storage Range: -40oC to + 70oC.
Environmental: Indoor use at altitudes to 2000m, Pollution Degree 1.
Safety: Complies with EN61010-1.
EMC: Complies with EN50081-1 and EN50082-1.
Designed and built in the U.K. by:
Thurlby Thandar Instruments Ltd.
Glebe Road, Huntingdon. Cambs. PE18 7DX England
Tel: 01480 412451 Fax: 01480 450409 Email: sales@ttinst.co.uk
Web: http://www.ttinst.co.uk
voltage ranges from 0-18V to 0-150V
powers from 30 watts to over 300 watts