40 Watt Medical Features l 3 x 2 x 0.75 Inches Form factor l 40 Watts Convection l Approval to EN60601 3rd Edition l Efficiencies 85% Typical -40 to 70 degree operating temperature* l Dual fusing l 2m Hours, Telcordia -SR332-issue 3 MTBF l No Load Power < 0.3W l Medical (BF) Safety Approvals l Meets standard IEC60601-1-2 : 2014 (4th Edition) l Electrical Specifications Input Voltage Input Frequency Input Current No Load Power Inrush Current Leakage Current Efficiency Hold-up Time Output Power Line Regulation Load Regulation Transient Response Rise Time Set Point Tolerance Over Current Protection Over Voltage Protection Short Circuit Protection Switching Frequency Operating Temperature4 Storage Temperature Relative Humidity Altitude MTBF Isolation Voltage 85-264 VAC/390 VDC, Universal 47-63 Hz 115 VAC: 0.8 A max. 230 VAC: 0.4 A max. less than 0.3W typical 115 VAC - 25 A, 230 VAC - 45 A, 264 VAC - 75 A 300 uA Typical, (N.A. For Class II Option) Touch current <100uA 85% Typical 40W:60 ms @230VAC 40W Convection Cooling +/-0.5% +/-1% 25% step load change, at 0.1A/uS slew rate, 50% duty cycle, 50/60Hz,max excursion 4%, recovery time 5mS. 50ms typical 2% (3% for 5V model) >110% 110 to 140% Hiccup mode 65 KHz Typical - 40 to +70C, * -40 to 0C startup is guaranteed with spec deviation -40 to +85C 5% to 95%, noncondensing Operating: 16,000 ft.; Nonoperating: 40,000 ft. 2m Hours, Telcordia -SR332-issue 3 Input to Output - 4000 VAC medical applications. Input to GND - 1500 VAC (Not Applicable For Class II Option) Output to GND- 1500VAC for type BF , 500 VAC for type B (Not Applicable For Class II Option) 27 4EM-19-180 39-DE60-44850-002 / A6 Option 1 Model Number Description Voltage Max. Load Min. Load Ripple1 5V 12 V 15 V 24 V 30 V 48 V 58 V 5A 3.33A 2.67A 1.67A 1.33A 0.83A 0.69A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 1.5% 1% 1% 1% 1% 1% 1% Description Voltage Max. Load Min. Load Ripple1 with PCB Mounting with PCB Mounting with PCB Mounting with PCB Mounting with PCB Mounting with PCB Mounting with PCB Mounting 5V 12 V 15 V 24 V 30 V 48 V 58 V 5A 3.33A 2.67A 1.67A 1.33A 0.83A 0.69A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 1.5% 1% 1% 1% 1% 1% 1% Voltage Max. Load Min. Load Ripple1 5V 12 V 15 V 24 V 30 V 48 V 58 V 5A 3.33A 2.67A 1.67A 1.33A 0.83A 0.69A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 0.0 A 1.5% 1% 1% 1% 1% 1% 1% MULP40-1305 with Header MULP40-1312 with Header MULP40-1315 with Header MULP40-1324 with Header MULP40-1330 with Header MULP40-1348 with Header MULP40-1358 with Header MULP40-CK metal cover kit accessory Option 2 Model Number MULP40-1205 MULP40-1212 MULP40-1215 MULP40-1224 MULP40-1230 MULP40-1248 MULP40-1258 Option 3 Model Number Description MULP40-1005 with Screw Terminal MULP40-1012 with Screw Terminal MULP40-1015 with Screw Terminal MULP40-1024 with Screw Terminal MULP40-1030 with Screw Terminal MULP40-1048 with Screw Terminal MULP40-1058 with Screw Terminal MULP40-CK metal cover kit accessory Connectors J1 Option 1 & 3 J2 Option 1 & 3 Pin 1 Pin 2 Pin 3 Pin 1,2 Pin 3,4 AC LINE NOT FITTED AC NEUTRAL V1 +VE V1 -VE 2 Innovations in Power 4EM-19-180 39-DE60-44850-002 / A6 Notes 1. Ripple is peak to peak with 20 MHz bandwidth and 10 F (Tantalum capacitor) in parallel with a 0.1 F capacitor at rated line voltage and load ranges. 2. Class II version available, Add "-II" suffix at the end of the Model Number. 3. Specifications are for nominal input voltage, 25C unless otherwise stated. 4. Output ripple can be more than 10% of the output voltage. 5. When used in Cover Kit, de-rate output power to 70 % under all operating conditions Mechanical Specifications AC Input Connector (J1) Option 1 DC Output Connector (J2) Option 1 Dimensions Weight Tyco: 640445-3 Mating: 647402-3; Pins: 3-647409-1 Tyco: 640445-4 Mating: 647402-4; Pins: 3-647409-1 3 x 2 x 0.75 inches (76.20 x 50.8 x 19.05 mm) 100 gm approx EMC Parameter Conducted Emissions Radiated Emissions Conditions/Description EN 55011-B,CISPR22-B, FCC PART15-B EN 55011 A Input Current Harmonics Voltage Fluctuation and Flicker ESD Immunity Radiated Field Immunity Electrical Fast Transient Immunity Surge Immunity Conducted Immunity Magnetic Field Immunity Voltage dips, interruptions EN 61000-3-2 EN 61000-3-3 EN 61000-4-2 EN 61000-4-3 EN 61000-4-4 EN 61000-4-5 EN 61000-4-6 EN 61000-4-8 EN 61000-4-11 Criteria Pass Pass Level B with external core (King core K5B RC 25x12x15-M in input cable) Class A Pass Level 4, Criterion A Level 3, Criterion A Level 3, Criterion A Level 3, Criterion A Level 3, Criterion A Level 4, Criterion A Criterion A & B Safety CE Mark Complies with LVD Directive Approval Agency Safety Standard(s) Safety File Number(s) Nemko, UL, C-UL EN60601-1, IEC 60601-1 (ed.3), ANSI / AAMI ES 60601 - 1, CSA C22.2 No. 60601-1 Class I : Nemko: P17221669, CB: NO95698, UL: 2017-1-30-E173812 Class II : Nemko: P17221705, CB : NO95847, UL: 2017-01-31-E173812 29 4EM-19-180 39-DE60-44850-002 / A6 Derating Curve De-rate linearly from 100% at 50C to 50% at 70C Mechanical Drawing Option 1 Notes: In case the PCB is mounted in a metal enclosure, using metal hardware ensure the following 1. Stand off, used to mount PCB has OD of 5.4 mm max. 2. Screws, used to fix PCB on stand off, have head dia of 6.0 mm max. 3. Washer, if used, to have dia of 6.5 mm max. 4 Innovations in Power 4EM-19-180 39-DE60-44850-002 / A6 Mechanical Drawing Option 2 Mechanical Drawing Option 3 31 4EM-19-180 39-DE60-44850-002 / A6