Page 1 of 2, Document #3023
725 Multifunciton Calibrator
©2000 Fluke Corporation
Rev. C-7/2000
Simultaneous Function Capability Channel A Channel B
24.000 mA DC M M or S
24.000 mA DC with 24V loop supply M
100.00 mV DC M or S
30.000V DC Measure M
20.000V DC Measure 10.000V DC Source M or S
15 to 3200 Ohms M or S
Thermocouple J, K, T, E, R, S, B, L, U, N M or S
RTD Ni120; Pt100 (392); Pt100 (JIS);
Pt100, 200, 500, 1000 (385) M or S
Pressure (requires Fluke 700PXX Modules) M M used as S
Frequency; Squarewave, 1 CPM to10 kHz;
fixed amplitude 5V p-p M or S
Fluke 725 Multifunction
Calibrator
Simply Powerful!
The new Fluke 725 Multifunction Process
Calibrator is a powerful yet easy-to-use field
calibrator. Use the measure and source functions
to test and calibrate almost any process
parameter.
Small, streamlined shape makes it easy
to carry
Rugged, reliable design stands up to field use
Easy to read measure/source screen lets you
view input and output simultaneously
Measure volts, mA, RTDs, thermocouples,
frequency, and ohms to test sensors and
transmitters
Source/simulate volts, mA, thermocouples, RTDs,
frequency, and ohms to calibrate transmitters
Measure/source pressure using any of 28
Fluke 700Pxx Pressure Modules
Source mA with simultaneous pressure
measurement to conduct valve and I/P tests
Support flow meter testing with frequency
and CPM functions
Perform fast linearity tests with auto step and
auto ramp features
Power transmitters during test using loop
supply with simultaneous mA measurement
Store frequently-used test setups for later use
Backlight lets you work in poor light
Remote interface allows benchtop automated
operations
Large battery capacity of four AA cells
Battery door for easy changes
Ordering information
Fluke 725 Multifunction Process Calibrator
Each calibrator includes: TL75 Test Leads, AC70A Test
Clips, one pair of stackable test leads, Users Manuals
appropriate to country of destination (English, plus three
of: French, German, Spanish, Italian, Dutch, Norwegian,
Danish, Swedish, Finnish, Portuguese, Korean, Chinese,
and Japanese), Statement of Quality Assurance Practices;
CE and CSA markings.
V mA
LOOP
100%
25%
25%
RECALL
C / F
MEAS
SOURCE
STORE
SETUP
TC RTD
V mA
PROCESS CALIBRATOR
MULTIFUNCTION
725
0%
M = Measure S = Source/Simulate
Fluke. Keeping your world
up and running.
Page 2 of 2, Document #3023
725 Multifunciton Calibrator
©2000 Fluke Corporation
Rev. C-7/2000
Resolution
J, K, T, E, L, N, U 0.1°C, 0.1°F
B, R, S 1°C, 1°F
Notes
Accuracy specifications include 0.2°C cold
junction uncertainty.
Thermocouple Measure or Source
J -200 to 0°C1.0°C
0 to 1200°C0.7°C
K -200 to 0°C1.2°C
0 to 1370°C0.8°C
T -200 to 0°C1.0°C
0 to 400°C0.8°C
E -200 to 0°C0.9°C
0 to 950°C0.7°C
R -20 to 0°C2.5°C
0 to 500°C1.8°C
500 to 1750°C1.4°C
S -20 to 0°C2.5°C
0 to 500°C1.8°C
500 to 1750°C1.5°C
B 600 to 800°C2.2°C
800 to 1000°C1.8°C
1000 to 1800°C1.4°C
L -200 to 0°C0.85°C
0 to 900°C0.7°C
U -200 to 0°C1.1°C
0 to 400°C0.75°C
N -200 to 0°C1.5°C
0 to 400°C0.9°C
Function Range Resolution Accuracy Notes
Measure or Source
Voltage 0 to 100 mV 0.01 mV .02% Rdg + 2 LSD Max load, 1 mA
0 to 10V (source) 0.01V
0 to 30V (measure) 0.01V
mA 0 to 24 0.001 mA .02% Rdg + 2 LSD Max load, 1000
mV (TC terminals) -10.00 mV to +75.00 mV .01 mV .025% of range + 1 LSD
Resistance 15 to 32000.01 to 0.10.10 to 1.0
Frequency 2.0 to 1000.0 CPM 0.1 CPM ±.05% For frequency source,
1 to 1000 Hz 1 Hz ±.05% waveform is 5V p-p
1.0 to 10.0 kHz 0.1 kHz ±.25% squarewave, -0.1V offset
Loop Supply 24V dc N/A 10%
Specifications
General specifications
Maximum voltage: 30V
Storage temperature: -40°C to 71°C
Operating temperature: 10°C to 55°C
Relative humidity: 95% (10°C to 35°C); 75% (30°C to 40°C);
45% (40°C to 50°C); 35% (50°C to 55°C)
Shock: 30g, 11ms, half-sine shock (or 1meter drop test)
Vibration: Random, 2g, 5-500 Hz
Safety: CSA C22.2 No. 1010.1:1992
EMC: EN50082-1:1992 and EN55022:1994 Class B
Size/weight: 96 x 200 x 47 mm (3.8 x 7.9 x 1.9 inches)
650g (23 oz)
Battery: Four AA alkaline batteries. Battery life: 25 hours typical
Warranty: Three years
Thermocouple accuracy specifications
Summary specifications (18°C to 28°C for one year)
Temperature coefficient, -10°C to 18°C, 28°C to 55°C, ±.005% of range per °C.
RTD ranges and accuracy specifications
RTD Types, Ranges and Accuracies
Measure
(4 wire) Source
Ni 120 -80°C to 260°C0.2°C0.2°C
Pt 100 - 385 -200°C to 800°C0.33°C0.33°C
Pt 100 - 3926 -200°C to 630°C0.3°C0.3°C
Pt 100 - 3916 (JIS) -200°C to 630°C0.3°C0.3°C
Pt 200 - 385 -200°C to 250°C0.2°C0.2°C
250°C to 630°C0.8°C0.8°C
Pt 500 - 385 -200°C to 500°C0.3°C0.3°C
500 to 630°C0.4°C0.4°C
Pt 1000 - 385 -200°C to 100°C0.2°C0.2°C
100°C to 630°C0.3°C0.2°C
Resolution
RTD 0.1°C, 0.1°F
Mouser Electronics
Authorized Distributor
Click to View Pricing, Inventory, Delivery & Lifecycle Information:
Fluke:
FLUKE-725 US FLUKE-725EX