Switch with IEC Connector Compact Size
Two function power entry module combining a DPST switch
and an IEC 320 power line connector. Compact packaging
allows a smaller panel cutout with choice of snap-in or
flange mounting. Available with general purpose filtering (F
models - CFS, CFE), medical application filtering (H models
-CHS, CHE) or unfiltered (CS, CE). Terminals allow for easy
connection to any wiring scheme. Preconnected terminals
from line to switch are available (B Models - CBS, CBE) for
reducing OEM wiring time.
The medical versions (H models) are designed specifically
for health care applications. Their extremely low leakage
current allows compliance with UL544 and UL2601 as well
as CSA22.2 health care specifications for use with both
patient and non-patient care equipment. For more
information on UL544 or UL2601, see Appendix C.
Note: When using the models with an IEC connector,
remember that the leakage current of the companion line
cord, GA400, is more significant than that of the filter -
which may disqualify this line cord from use in patient care
applications.
Specifications
Maximum leakage current, each line to ground: †
FModels H Models
@120 VAC 60 Hz: .25 mA 2 µA
@250 VAC 50 Hz: .40 mA 5µA
Hipot rating (one minute):
line-to-ground 2250 VDC
line-to-line 1450 VDC
Operating frequency: 50/60 Hz
Rated voltage: 250 VAC
Operating voltages: 115/230 VAC
Switch: Double-insulated rated
for 10,000 operations
at full load.
51 Amp inrush capability.
Unfiltered Push-on Force: 18 lb/80 N
Minimum insertion loss in dB:
Line-to-ground in 50 ohm circuit
Current Frequency MHz
Rating 0.05 0.15 0.5 1 5 10 30
FModels
01A 10 26 46 48 46 47 46
03A 8 16 32 36 43 48 50
06A 4 11 22 27 36 41 50
10A 1 4 14 18 27 33 42
HModels
01A 16 21 37 44 26 21 10
03A 9 14 31 32 26 24 14
06A 4 10 22 23 19 18 13
10A 2 6 10 15 11 11 9
Line-to-line in 50 ohm circuit
Current Frequency MHz
Rating 0.05 0.15 0.5 1 5 10 30
FModels
01A 1 3 13 28 62 67 42
03A 2 6 14 23 65 65 67
06A 2 6 14 27 46 48 58
10A 1 7 14 23 42 44 62
HModels
01A 1 6 13 29 38 42 26
03A 1 5 10 22 36 34 36
06A 1 5 14 20 31 33 37
10A 1 4 11 19 32 37 38
110 Catalog 1654001 Dimensions are in inches and Dimensions are shown for USA Cust. Svc.: 1-800-468-2023 www.corcom.com
Revised 05-07 millimeters unless otherwise reference purposes only. CORCOM Prods: 1-847-680-7400 www.tycoelectronics.com
specified. Values in italics Specifications subject Germany: 49-89-6089-0
are metric equivalents. to change.
C Series
C Series
CORCOM Product Guide
UL Recognized
CSA Certified
VDE Approved CFS CS
Electrical Schematics
Resistor location for reference only.
F Models (CFS, CFE)
H Models (CHS, CHE)
B Models (CBS, CBE)
N
Power On/Off
L
G
L
G
N
Note 1: Jumpers provided on CBS and CBE versions only.
Note 2: Location of optional filter
Note 1
Note 1
Note 2
Note 2
†Applies to Filtered models only.