No. SS2-MGH200-0100 (Rev.3) MagneW 3000 PLUS Smart Electromagnetic Flowmeter HENRI Flowmeter for Pulp Slurry Model MGH14C (Converter) / MGH18 (Detector) OVERVIEW The MagneW 3000 PLUS HENRI (High-Energy Noise-Resistant Instrument) flowmeter is a highperformance and highly reliable flowmeter for Pulp & Paper applications based on Yamatake's proven MagneW 3000 flow measurement technologies. The MagneW 3000 PLUS HENRI offers a stable and accurate measurement for severe Pulp and Paper applications with a unique lining manufacturing technology and high-frequency excitation technology. Size 40 to 400 mm (11/2 to 16 inches) FEATURES Stable and accurate measurement by high-frequency excitation technology * MagneW 3000 PLUS HENRI minimizes the output fluctuation caused by various noises (Slurry noise, AC noise and Electrochemical noise) and realizes stable and accurate flow measurement for pulp slurry applications. High-performance lining * A unique high quality lining technology makes it possible to offer mirror-finish PFA lining. The mirror-finish PFA lining is best suited for scaling applications. * A new material ETFE (harder than PFA) lining is suitable for wearing applications in Pulp & Paper process. ETFE lining extends life-time of lining than (that of) PFA lining. Remote converter APPLICATIONS Applications for pulp production, preparation and papermaking process * Pulpstock flow to head box * DIP (De-inked pulp) application * Pulp from screen * Green liquor * Diluted black liquor * Dye/pigment dosing application * White liquor * High consistency pulp * Pulp from pulper Replaceable face-to-face dimension detector (Optional) * This detector can replace the detector of our existing models and those of other suppliers. Please consult your Yamatake representative for details. Rugged detector structure * A stainless steel housing has been adopted for sizes 40 to 200 mm (11/2 to 8 inches). * A watertight structure is used for the watertight model (Model MGH18). Specifications are subject to change without notice. -1- First issue: Dec. 1999 Rev.3: Jun. 2003 No. SS2-MGH200-0100 (Rev.3) Yamatake Corporation Digital output range/load resistance For converter (Model MGH14C) With SFC communication 3.2mA to 22.4mA (-5% to +115%) External power source for SFC communication/ digital communication 16 to 45 V DC Load resistance = (External power source for Communication 8.5V)/0.025 FUNCTIONAL SPECIFICATIONS Type of protection JIS C 0920 waterproof model NEMA ICS6-110 TYPE 4X IEC PUBL 529 IP66 Input signal Unit of flow indication Flow signal Electromotive force from the detector Volume flow m3, L, cm3, G, mG, kG, B Contact input (optional) Solid-sate contact or no-voltage contact Mass flow t, kg, g, lb Output signal Time d, h, min, s Analog output 4 to 20mADC Damping time Digital output DE Analog or digital output is selectable. Adjustable between 0.5 and 199.9 seconds Low flow cutoff Adjustable between 0 to 10% of setting range. Below selected value, output is driven to the zero flow rate signal level. Contact output (optional) Open collector Contact capacity: 30 V DC max., 200mA max. Drop out Pulse output (optional) Open collector Contact capacity: 30 V DC max., 200mA max. Pulse frequency: 2000 Hz max. Pulse width: 0.3 to 999.9 ms Adjustable from 0.3 to 999.9 ms or fixed at 50% of duty Adjustable between 0 to 10% of setting range. Below selected value, pulse output is cut. Lightning protection 12 kV, 1000A Equipped with the lightning arrester in the power source and external input and output terminals. Power failure Analog output range/load resistance An EEPROM retains data record of totalized value when pulse output is used (retention period approximately 10 years). Without SFC communication 0.8mA to 22.4mA (-20% to +115%) Load resistance: 0 to 600 Power supply With SFC communication 3.2mA to 22.4mA (-5% to +115%) External power source for communication DC16V to 45V Load resistance = (External power source for Communication -8.5 V) / 0.025 90 to 246 V AC Power supply frequency 47 to 63 Hz Power consumption 30 W max. With HART communication Load resistance: 250 minimum Ambient temperature limits -25 to +50C (-13 to +122F) Ambient humidity limits 5 to 100% RH -2- Yamatake Corporation No. SS2-MGH200-0100 (Rev.3) Optional specifications Self-check result output Activates only when a self-diagnostic abnormality occurs. Display Indication by LCD with backlighting Main display 7-segment, 6 digits Empty detection output Activates only when an empty status is detected. Sub display 16 digits, 2 lines High/low limit alarm output Activates when a high/low limit alarm occurs. Display contents Demonstrates three values simultaneously. * Flow volume in percentage * Actual flow volume * Totalized value (when pulse output selected) Two-stage flow rate alarm output (with two contact outputs) Activates when the first high/low limit alarm (H/ L) occurs and the second high/low limit alarm (HH/LL) occurs. Selection of main display and sub display Main display is selectable from the three values. Empty pipe detection When the detector is empty, the analog output, digital output and pulse output are fixed at zero. Display is latched to zero. Data setting device Setting by infrared ray touch sensor Infrared ray touch sensor: Four key switches Pulse output Outputs the totalized value. Must be selected in case of using totalizer. Contact input function External 0% lock input Forces outputs (analog, digital, pulse) to the zero flow rate signal level. Activates by applying a contact input. Built-in counter function Totalizer According to the pulse scale setting, it totals one count at a time, for normal and reverse flows. External automatic zero adjustment input Adjusts zero-point. Activates by applying a contact input. Totalizer with presetting function A preset value (target totalized value) can be set between 0000000000 and 9999999999. The counting method is same as that of the standard totalizer. External range switching input Switches two flow measurement ranges. Two measurement ranges * Dual range for normal directions * Normal/reverse range Activates by applying a contact input. Normal/reverse flow difference totalizer The difference in flow volumes in the normal and reverse directions is calculated and displayed. Built-in counter reset input (for pulse output model) Resets the totalized value of the built-in counter. Activates by applying a contact input. Traceability certificate The following three documents are provided. * Traceability System Chart * Traceability Certificate * Test Report Contact output function Alarm output Outputs an alarm under the following conditions. * Self-diagnostic result * Empty detection * High/low limit alarm Tropicalization treatment Protects the electromagnetic flow meter in harsh environments during transportation and/or storage. The following treatments are applied. * Corrosion protection * Moisture prevention * Mildew proofing Range switching output * Outputs the status of flow range. * Large/small in dual range * Normal/ Reverse Indication other than SI units The following non-SI units are available. Counter preset status output (for pulse output model) Activates when the counter reaches the preset value. Volume unit B (barrel), kG (kilo-gallon), G (gallon), mG (milli-gallon) Mass unit lb (pound) -3- No. SS2-MGH200-0100 (Rev.3) Yamatake Corporation PHYSICAL SPECIFICATION For detector (Model MGH18) Housing material Aluminum alloy FUNCTIONAL SPECIFICATIONS Finish Type of protection Standard Baked acryl paint JIS C 0920 watertight model NEMA ICS6-180 TYPE 4X IEC PUBL 529 IP67 Corrosion-resistant Baked acryl paint Size Wafer style 40, 50, 80, 100, 150, 200 mm (11/2, 2, 3, 4, 6, 8 inches) Corrosion-proof Baked epoxy paint Color Light beige (Munsell 4Y7.2/1.3) Flange style 40, 50, 80, 100, 150, 200, 250, 300, 350, 400 mm (11/2, 2, 3, 4, 6, 8, 10, 12, 14, 16 inches) Display cover material Tempered glass, 5 mm thick Flange rating Weight JIS 10K, JIS 16K, JIS 20K, JIS 30K, JPI1 50, JPI 300, ANSI 150, ANSI 300, DIN PN10, DIN PN16, DIN PN25, DIN PN40 (Size 40 to 200 mm (11/2 to 8 inches)) 3.7 kg (8.2 lb) INSTALLATION Model MGH18 detector must be coupled with model MGH14C converter. Electrical connection G1/2 (PF1/2) internal thread, 1/2NPT internal thread, CM20 internal thread, Pg13.5 internal thread Mounting JIS 10K, JIS 20K, JPI 150, JPI 300 ANSI 150, ANSI 300, DIN PN10, DIN PN16, DIN PN25 (Size 250 to 400 mm (10 to 16 inches)) Ambient temperature limits -30 to +80C (-22 to +176F) Wall mounting, 2-inch pipe mounting Ambient humidity limits Grounding 5 to 100% RH Resistance less than 100 Measurable electrical conductivity Coupled with model MGH14C converter 20 S/cm or greater (In case of less than 20 S/cm, consult your Yamatake representative.) -4- Yamatake Corporation No. SS2-MGH200-0100 (Rev.3) Temperature range pressure range process fluid PFA lining ETFE lining Size: 80 to 200 mm (3 to 8 inches) Size: 40 to 200 mm (11/2 to 8 inches) 2.94 MPa (426 psi) 2.94 MPa (426 psi) 0.98 MPa (142 psi) 0.98 MPa (142 psi) -0.098 MPa (-14.2 psi) -40 (-40) 80 (176) 120 (248) -0.098 MPa (-14.2 psi) -40 (-40) 160 C (320 F) 80 (176) 120 C (248 F) Size: 250 to 400 mm (10 to 16 inches) Size: 250 to 400 mm (10 to 16 inches) 1.96 MPa (284 psi) 1.96 MPa (284 psi) 0.98 MPa (142 psi) 0.98 MPa (142 psi) -0.098 MPa (-14.2 psi)-40 (-40) -0.098 MPa (-14.2 psi) -40 (-40) 100 120 C (212) (248 F) 100 120 C (212)(248 F) : Standard : Standard : Refer to factory basis Consult your Yamatake representative : Refer to factory basis Consult your Yamatake representative Measurement flow range Minimum flow velocity range is 0 Maximum flow velocity range is 0 to 0.3 m/s (0 to 0.98 ft/s) to 10m/s (0 to 32.8 ft/s) Size Minimum range Maximum range Conversion factor K mm inches m3/h GPM m3/h GPM 40 11/2 0 to 1.358 0 to 5.976 0 to 45.23 0 to 199.1 0.2210 50 2 0 to 2.121 0 to 9.337 0 to 70.68 0 to 311.2 0.1415 65 21/2 0 to 3.584 0 to 15.78 0 to 119.4 0 to 525.9 0.08371 80 3 0 to 5.429 0 to 23.90 0 to 180.9 0 to 796.7 0.05526 100 4 0 to 8.483 0 to 37.35 0 to 282.7 0 to 1244 0.03537 150 6 0 to 19.09 0 to 84.04 0 to 636.1 0 to 2801 0.01572 200 8 0 to 33.93 0 to 149.4 0 to 1130 0 to 4979 0.008842 250 10 0 to 53.02 0 to 233.5 0 to 1767 0 to 7781 0.005659 300 12 0 to 76.35 0 to 336.2 0 to 2544 0 to 11205 0.003930 350 14 0 to 104.0 0 to 457.6 0 to 3463 0 to 15251 0.002887 400 16 0 to 135.8 0 to 597.7 0 to 4523 0 to 19920 0.002210 Velocity V(m/s) = K x Q K = Flow conversion factor = 1/3600 x 4/(D2) Q = Flow rate (m3/h) -5- No. SS2-MGH200-0100 (Rev.3) Yamatake Corporation PHYSICAL SPECIFICATIONS Optional specifications Test report Calibration certificate, withstand voltage test, insulation resistant, hydrostatic pressure test, physical inspection are included. Main body material Measuring pipe material SUS304 stainless steel Traceability certificate The following three documents are included. * Traceability System Chart * Traceability Certificate * Test Report Flange SUS304 stainless steel (Size 40 to 50 mm (11/2 to 2 inches)) Carbon steel + corrosion-preventive painting (Size 80 to 400 mm (3 to 16 inches)) Material certificate Material certificate for electrode/grounding ring Housing SUS304 stainless steel (Size 40 to 200 mm (11/2 to 8 inches)) Carbon steel (Size 250 to 400 mm (10 to 16 inches)) Gasket for plastic piping When the detector is being mounted on plastic pipe, attach this gasket between the lining and the grounding ring, and between the grounding ring and the plastic pipe flange. Terminal box Aluminum alloy Attaching the tag number to the terminal box Stamp the tag with the specified number and attach to the terminal box. The maximum number of characters of the tag number is 8. Attaching the tag number to the neck section Stamp the tag with the specified number and attach to the neck section of the detector with stainless wire. The maximum number of characters of the tag number is 16. Terminal box Baked acryl paint Case (size 250 to 400 mm (10 to 16 inches)) Polyurethane paint Corrosion-proof Terminal box Baked epoxy paint Vs = velocity of setting range 1.0 < Vs 10 0.5% of rate 0.3 < Vs 1.0 (0.1/Vs+0.4)% of rate Terminal box Baked acryl paint Corrosion-resistant Accuracy (model MGH14C with model MGH18 detector) Velocity during measurement > Vs x 20% Standard Case (size 250 to 400 mm (10 to 16 inches)) Polyurethane paint PERFORMANCE SPECIFICATIONS Vs (m/s) Finish Velocity during measurements < Vs x 20% Case (size 250 to 400 mm (10 to 16 inches)) Epoxy paint Color Light beige (Munsell 4Y7.2/1.3) 0.1% of Vs 0.2(0.1/ Vs+0.4)% of Vs Process wetted materials Lining PFA (Size 40 to 400 mm (11/2 to 16 inches)) ETFE (Size 80 to 400 mm (3 to 16 inches)) Electrode SUS316L, Hastelloy C, titanium, zirconium, tantalum, tungsten-carbide, platinum-iridium Grounding ring SUS316, Hastelloy C, titanium, zirconium, tantalum, platinum Gasket PTFE (if grounding ring material is exotic.) -6- Yamatake Corporation No. SS2-MGH200-0100 (Rev.3) INSTALLATION Electrical connection G1/2 (PF1/2) internal thread 1/2NPT internal thread CM20 internal thread Pg13.5 internal thread Pipe connection Wafer (Size 40 to 200 mm (11/2 to 8 inches)) Flange (Size 40 to 400 mm (11/2 to 16 inches)) Grounding Resistance less than 100 Length of straight pipe Upstream side A minimum five straight pipe diameters A minimum 10 straight pipe diameters is required if a diffuser/valve/pump is installed upstream side. Downstream side Two straight pipe diameters is recommended. Cable (between remote detector and converter) Maximum length 300 m (984 ft) (depends on fluid conductivity) Outer diameter 10 to 12 mm (0.4 to 0.47 inch) Signal cable Dedicated cable: MGA12W (O.D. 11.4 mm, 0.75 mm2) or equivalent (CVVS, CEEV etc.) Excitation cable Dedicated cable: MGA12W (O.D. 10.5 mm, 2 mm2) or equivalent (CVV and others) -7- No. SS2-MGH200-0100 (Rev.3) Yamatake Corporation MODEL SELECTION Type / Basic model No. Lining material Pipe connection Size mm (inches) Page Waterproof converter model MGH14C page 8 Watertight detector model MGH18D PFA / ETFE Wafer 40 to 200 11/2 to 8 page 9 Watertight detector model MGH18F PFA / ETFE Flange 40 to 400 11/2 to 16 page 10 HENRI converter Model MGH14C converter must be coupled with model MGH18 HENRI type detector Model MGH14C - I II III IV - V VI VII VIII - Y / Options (some options can be selected per each model) Option "Y" must be specified for Yamatake version. Basic model no. MGH14C I II - 90-264 V AC 50/60 Hz M Volume flow 4-20mA DC output / without communication Volume flow 4-20mA DC output / with communication Volume flow DE output / with communication Volume flow HART protocol output / with communication Electrical connec- G1/2 internal thread / without watertight gland tion / watertight G1/2 internal thread / with brass (Ni-plated) watertight gland gland G1/2 internal thread / with plastic watertight gland 1/2NPT internal thread / without watertight gland CM20 internal thread / without watertight gland Pg13.5 internal thread / without watertight gland G1/2 internal thread / with SUS304 watertight gland Installation / wir- Remote type Wall mounting with standard bracket ing direction 2-inch pipe mounting with standard bracket Without mounting bracket IV V Finish VI Display VII Contact inputs / outputs VIII Style code Optional selections - Power supply output signal / communication III Options Selections A B C H 1 2 3 4 5 6 7 G H X Standard X Corrosion-resistant finish 1 Corrosion-proof finish 2 None Main display: Instantaneous indication of flow volume in % Main display: Instantaneous indication of actual flow volume Main display: Indication of integrated flow volume (need pulse output board) Note 1 None 1 input and 1 output (ranging function, warning for contact input/output, etc.) 2 input (ranging function, external automatic zero adjustment input, etc.) 2 output (ranging function, warning for contact output, etc.) None Yamatake version (must be selected) Empty pipe detection function Pulse output (open collector) Traceability certificate for converter Indication other than SI units Attachment of the TAG number to the terminal box for converter Note) 1. In case of this code, option "B" must be selected. -8- Y A B C H J X A B C X 1 2 3 X Yamatake Corporation No. SS2-MGH200-0100 (Rev.3) HENRI detector Wafer type (40 to 200 mm (11/2 to 8 inches)) PFA / ETFE liner Model MGH18D detector must be coupled with MGH14C converter Model MGH18D - I II III IV V VI VII VIII IX - X XI - Y/Options (some options can be selected per each model) Option "Y" must be specified for Yamatake version Basic model no. Selections MGH18D I - Line size II III IV V VI VII VIII IX 40 mm (11/2 inches) 040 50 mm (2 inches) 050 80 mm (3 inches) 080 100 mm (4 inches) 100 150 mm (6 inches) 150 200 mm (8 inches) 200 Linear ETFE (Size 80 to 200 mm (3 to 8 inches)) E PFA P Piping connection Wafer JIS 10K 11 Wafer JIS 20K 12 Wafer JIS 30K 13 Wafer ANSI 150 21 Wafer ANSI 300 22 Wafer DIN PN10 41 Wafer DIN PN16 42 Wafer DIN PN25 43 Wafer DIN PN40 44 Wafer JPI 150 61 Wafer JPI 300 62 Electrode SUS316L L Hastelloy C C Titanium K Zirconium H Tantalum T Tungsten carbide W Platinum iridium P Others ! Grounding ring SUS316 Hastelloy C Titanium Zirconium Tantalum Platinum iridium Others Electrical connection / Remote G1/2 internal thread / without watertight gland G1/2 internal thread / with brass (Ni-plated) watertight gland watertight gland type G1/2 internal thread / with plastic watertight gland 1/2NPT internal thread / without watertight gland CM20 internal thread / without watertight gland Pg13.5 internal thread / without watertight gland G1/2 internal thread / with SUS304 watertight gland Face-to-face dimenStandard Others sions Installation / wiring Remote Upstream side (horizontal / vertical piping mounting) Downstream side (horizontal / vertical piping mounting) direction type Horizontal piping mounting / left side viewed from upstream Horizontal piping mounting / right side viewed from upstream Calibration Standard Others X Finish XI Bolt / nut Options Optional selections - Standard Corrosion-resistant finish Corrosion-proof finish None Carbon steel SUS304 S C K H T P ! 2 3 4 5 6 7 8 A ! A B C D A ! X 1 2 X 1 2 Yamatake version (must be selected) Calibration certificate (sent to ordering location separately) Traceability certificate for detector Material certificate (only for electrodes and ground rings) With gasket for plastic piping Attachment of the TAG number to the terminal box for detector (Note 1) Attachment of the TAG number plate to neck section for detector (Note 1) Note) 1. Must be selected for Tag no. requirement -9- Y A B C J K L No. SS2-MGH200-0100 (Rev.3) Yamatake Corporation HENRI detector Flange type (40 to 400 mm (11/2 to 16 inches)) PFA / ETFE liner Model MGH18F - I II III IV V VI VII VIII IX X - XI - Y/Options (some options can be selected per each model) Option "Y" must be specified for Yamatake version Basic model no. Selections MGH18F I III IV V VI VI VIII X Options XI - Line size II IX Optional selections - 40 mm (11/2 inches) 040 50 mm (2 inches) 050 80 mm (3 inches) 080 100 mm (4 inches) 100 150 mm (6 inches) 150 200 mm (8 inches) 200 250 mm (10 inches) 250 300 mm (12 inches) 300 350 mm (14 inches) 350 400 mm (16 inches) 400 Linear ETFE (Size 80 to 400 mm (3 to 16 inches)) E PFA P Piping connection Flange JIS 10K J1 Flange JIS 20K J2 Flange JIS 30K J3 Flange ANSI 150 A1 Flange ANSI 300 A2 Flange DIN PN10 D1 Flange DIN PN16 D2 Flange DIN PN25 D3 Flange DIN PN40 D4 Flange JPI 150 P1 Flange JPI 300 P2 Flange material Standard 1 Electrode SUS316L L Hastelloy C C Titanium K Zirconium H Tantalum T Tungsten carbide W Platinum iridium P Others ! Grounding ring SUS316 Hastelloy C Titanium Tantalum Platinum iridium Others Electrical connection / Remote G1/2 internal thread / without watertight gland G1/2 internal thread / with brass (Ni-plated) watertight gland watertight gland type G1/2 internal thread / with plastic watertight gland 1/2NPT internal thread / without watertight gland CM20 internal thread / without watertight gland Pg13.5 internal thread / without watertight gland G1/2 internal thread / with SUS304 watertight gland Face-to-face dimensions Standard Others Installation / wiring direc- Remote Upstream side (horizontal / vertical piping mounting) Downstream side (horizontal / vertical piping mounting) tion type Horizontal piping mounting / left side viewed from upstream Horizontal piping mounting / right side viewed from upstream Calibration Standard Others Finish Standard Corrosion-resistant finish Corrosion-proof finish S C K T P ! 2 3 4 5 6 7 8 A ! A B C D A ! X 1 2 Yamatake version (must be selected) Calibration certificate (sent to ordering location separately) Traceability certificate for detector Material certificate (only for electrodes and ground rings) With gasket for plastic piping Attachment of the TAG number to the terminal box for detector (Note 1) Attachment of the TAG number plate to neck section for detector - 10 - Y A B C J K L Note) 1. Must be selected for Tag no. requirement Yamatake Corporation No. SS2-MGH200-0100 (Rev.3) DIMENSIONS All dimensions are in millimeters, dimensions in brackets ( ) are in inches (inch). Wafer type Size: 40 to 200 mm (11/2 to 8 inches) Size (1.19 max.) (3.31) Face to face dimension (3.31) L H1 (1.85) H2 Height H3 H4 Housing W width Housing outer D diameter Weight mm (inches) mm (inches) mm (inches) mm (inches) mm (inches) mm (inches) mm (inches) mm (inches) kg (lb) 40 (11/2) 80 (3.15) 84 (3.31) 43.5 (1.71) 174.5 (6.87) 109 (4.29) 125 (4.92) 87 (3.43) 2.8 (6.2) 50 (2) 86 (3.39) 93 (3.66) 52 (2.05) 192 (7.56) 118 (4.65) 135 (5.31) 104 (4.09) 3.4 (7.5) 80 (3) 106 (4.17) 108 (4.25) 67 (2.64) 222 (8.74) 133 (5.24) 164 (6.46) 134 (5.28) 5.2 (11.5) 100 (4) 120 (4.72) 120.5 (4.74) 79.5 (3.13) 247 (9.72) 145.5 (5.73) 189 (7.44) 159 (6.26) 6.7 (14.8) 150 (6) 160 (6.30) 160 (6.30) 110 (4.33) 317 (12.48) 185 (7.28) 240 (9.45) 220 (8.66) 13.6 (30.0) 200 (8) 200 (7.87) 185 (7.28) 135 (5.31) 367 (14.45) 210 (8.27) 290 (11.42) 270 (10.63) 22.0 (48.5) Flange type Size: 40 to 400 mm (11/2 to 16 inches) (1.19 max.) (3.31) (3.31) (1.85) (1.19 max.) (3.31) (3.31) (1.85) Size Face to face dimension L H1 H2 Height H3 H4 Weight mm (inches) mm (inches) mm (inches) mm (inches) mm (inches) mm (inches) kg (lb) 40 (11/2) 200 (7.87) 84 (3.31) 77 (3.03) 208 (8.19) 109 (4.29) 6.0 (13.2) 50 (2) 200 (7.87) 93 (3.66) 88 (3.46) 228 (8.98) 118 (4.65) 10.5 (23.1) 80 (3) 200 (7.87) 108 (4.25) 113 (4.45) 268 (10.55) 133 (5.24) 15.5 (34.2) Note) The table indicated dimensions for ANSI 150 flange - 11 - 100 (4) 250 (9.84) 120.5 (4.74) 131 (5.16) 299 (11.77) 145.5 (5.73) 23.3 (51.4) 150 (6) 300 (11.81) 160 (6.30) 174 (6.85) 381 (15.0) 185 (7.28) 35.4 (78.0) 200 (8) 350 (13.78) 185 (7.28) 204 (8.03) 436 (17.17) 210 (8.27) 60.0 (132) 250 (10) 450 (17.72) 235 (9.25) 224 (8.82) 506 (19.92) 260 (10.24) 68.0 (150) 300 (12) 500 (19.69) 258 (10.16) 271 (10.67) 576 (22.68) 283 (11.14) 97.0 (214) 350 (14) 550 (21.65) 282 (11.10) 297 (11.09) 626 (24.65) 307 (12.09) 128.0 (282) 400 (16) 600 (23.62) 310 (12.20) 340 (13.39) 697 (27.44) 335 (13.19) 163.0 (359) No. SS2-MGH200-0100 (Rev.3) Yamatake Corporation Wall mount type (5.04) (5.12) (0.55) (0.61) (3.23) (8.50) (5.91) (4.33) (9.76) (0.13) (1.18) (1.34) (4.06) (0.98) 2-inch pipe mount type (1.74) (5.04) (0.55) (0.57) (3.23) (8.50) (5.12) (5.91) (1.18) (1.34) (4.06) (0.98) Savemation Saving through Automation Totate international Building 2-12-19 Shibuya Shibuya-ku, Tokyo 150-8316 Japan Tel : 81-3-3486-2310 Fax : 81-3-3486-2593 - 12 This has been printed on recycled paper. http://www.yamatake.com/