200W Reliable Green Medical Power Supply + EN 60601-1 IEC 60601-1 8 ANSI/AAMI ES60601-1 EN60601-1 RPS-200 series + IEC60601-1 Features 4"x2" compact size Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/EN60601-1 Suitable for BF application with appropriate system consideration 140W convention, 200W force air EMI Conduction for Class B Radiation for Class B with FG(Class) and Class A without FG(Class) No load power consumption<0.5W Extremely low leakage current 12V/0.5A fan supply Protections: Short circuit / Overload / Over voltage / Over temperature Lifetime > 65K hours Operating altitude up to 5000 meters 3 years warranty Applications Oral irrigator Hemodialysis machine Medical monitors Sleep apnea devices Pumps machine Electric bed Description 3 RPS-200 is a 200W highly reliable green PCB type medical power supply with a high power density (21.9W/in) on the 4 by 2 footprint. It accepts 80~264VAC input and offers various output voltages between 12V and 48V. The working efficiency is up to 95% and the extremely low no load power consumption is down below 0.5W. RPS-200 is able to be used for both Class(with FG) and Class(no FG) system design. The extremely low leakage current is less than 130A. In addition, it conforms to the international medical regulations (2*MOPP) and EMC EN55011, perfectly fitting all kinds of BF rated patient contact medical system equipment. Model Encoding RPS - 200 - 12 -C Type Output voltage Rated wattage Series name Type Blank C Description PCB Type Enclosed casing Type Note In stock In stock File Name:RPS-200-SPEC 2016-11-09 File Name:RPS-200-SPEC 2020-10-20 RPS-200 200W Reliable Green Medical Power Supply series SPECIFICATION MODEL RPS-200-12 12V RPS-200-15 15V RPS-200-24 24V RPS-200-27 27V RPS-200-48 48V 16.7A 13.4A 8.4A 7.5A 4.2A 9.4A 5.9A 5.3A 3A 200.4W 201W 201.6W 202.5W 201.6W Convection 140.4W DC VOLTAGE CURRENT RATED POWER OUTPUT 10CFM Convection 11.7A 10CFM 141W 141.6W 143.1W 144W RIPPLE & NOISE (max.) Note.2 100mVp-p 100mVp-p 120mVp-p 120mVp-p 120mVp-p VOLTAGE ADJ. RANGE 14.3~15.8V 22.8~25.2V 25.6 ~ 28.4V 45.6 ~50.4V VOLTAGE TOLERANCE Note.3 2.0% 2.0% 1.0% 1.0% 1.0% LINE REGULATION 0.5% 0.5% 0.5% 0.5% 0.5% LOAD REGULATION 1.0% 1.0% 1.0% 1.0% 1.0% SETUP, RISE TIME 700ms, 30ms/230VAC HOLD UP TIME (Typ.) 16ms/230VAC 94% 95% 11.4~12.6V VOLTAGE RANGE Note.4 80 ~ 264VAC INPUT 700ms, 30ms/115VAC at full load 16ms/115VAC at full load 113 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz POWER FACTOR PF>0.94/230VAC PF>0.98/115VAC at full load EFFICIENCY (Typ.) 93% AC CURRENT (Typ.) 2A/115VAC INRUSH CURRENT (Typ.) COLD START 30A/115VAC 93.5% 94% 1A/230VAC 60A/230VAC LEAKAGE CURRENT(max.)Note.5 Earth leakage current < 130A/264VAC , Touch current < 40A/264VAC OVERLOAD PROTECTION 110 ~ 140% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed 16.5 ~ 19.5V 13.2 ~ 15.6V OVER VOLTAGE OVER TEMPERATURE FUNCTION FAN SUPPLY 26.4 ~ 31.2V 29.7 ~ 35V 52.8 ~ 62.4V Protection type : Shut down o/p voltage, re-power on to recover Protection type : Shut down o/p voltage, re-power on to recover 12V@0.5A for driving a fan ; tolerance +15% ~ -15% WORKING TEMP. -30 ~ +70 (Refer to "Derating Curve") WORKING HUMIDITY 20 ~ 90% RH non-condensing ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH non-condensing TEMP. COEFFICIENT 0.03%/ (0 ~ 50) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes OPERATING ALTITUDE Note.6 5000 meters SAFETY STANDARDS IEC60601-1, TUV EN60601-1, EAC TP TC 004 ,UL ANSI / AAMI ES60601-1 (3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335 - 1 ISOLATION RESISTANCE Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP WITHSTAND VOLTAGE I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC ISOLATION RESISTANCE I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25/ 70% RH Parameter Conducted emission EMC EMISSION SAFETY & EMC (Note 7) EMC IMMUNITY OTHERS NOTE Standard EN55011 (CISPR11) EN55011 (CISPR11) EN61000-3-2 EN61000-3-3 Radiated emission Harmonic current Voltage flicker EN60601-1-2 Parameter ESD Standard EN61000-4-2 RF field susceptibility EN61000-4-3 EFT bursts Surge susceptibility Conducted susceptibility Magnetic field immunity EN61000-4-4 EN61000-4-5 EN61000-4-6 EN61000-4-8 Voltage dip, interruption EN61000-4-11 Test Level / Note Class B Class A (for Class II);Class B (for Class I) Class A ----- Test Level / Note Level 4, 15KV air ; Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz ) Level 3, 2KV Level 4, 4KV/Line-FG ; 2KV/Line-Line Level 3, 10V Level 4, 30A/m 100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods MIL-HDBK-217F (25) MTBF 500.2Khrs min. DIMENSION (L*W*H) PCB:101.6*50.8*29mm or 4*2*1.14inch ; Enclosed type:103.4*62*40mm or 4.07*2.44*1.57inch PACKING PCB:0.19Kg; 72pcs/14.7Kg/0.84CUFT ; Enclosed type:0.3Kg; 60pcs/19Kg/1.06CUFT 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1f & 47f parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. Derating may be needed under low input voltages. Please check the derating curve for more details. 5. Touch current was measured from primary input to DC output. 6. The ambient temperature derating of 3.5/1000m with fanless models and of 5/1000m with fan models for operating altitude higher than 2000m(6500ft). 7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) Product Liability DisclaimerFor detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx File Name:RPS-200-SPEC 2020-10-20 RPS-200 200W Reliable Green Medical Power Supply series Block Diagram fosc : 65KHz EMI FILTER & RECTIFIER EMI FILTER I/P RECTIFIERS & FILTER POWER SWITCHING +V -V RECTIFIERS & FILTER FG +12V FAN supply DETECTION CIRCUIT O.T.P. O.L.P. CONTROL O.V.P. Derating Curve Output Derating VS Input Voltage 100 100 With 10CFM Fan for RPS-200(-C) With 10CFM Fan for RPS-200(-C) 90 80 80 70 Convection for RPS-200 24~48V 70 LOAD (%) LOAD (%) 60 40 Convection for RPS-200-C 12 V~ 48 V & RPS-200 12V~15V 20 -30 0 10 20 30 40 45 50 AMBIENT TEMPERATURE () 60 70 (HORIZONTAL) 60 50 40 80 90 115 120 130 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz File Name:RPS-200-SPEC 2020-10-20 RPS-200 200W Reliable Green Medical Power Supply series Mechanical Specification RPS-200 (PCB Type) Top View FAN 5cm 3.175 Note 48 10CFM Air flow direction LED SVR1 44.45 CN1 3 2 1 CN101 FS1 T4A/250V FS2 F4A/250V 3.175 2 1 95.25 1 2 3 4 5 6 4 - 3.3 29 101.6 3 max. 50.8 CN2 Side View File Name:RPS-200-SPEC 2020-10-20 RPS-200 200W Reliable Green Medical Power Supply RPS-200-C (Enclosed Type) Case No.245A series Unit:mm Side View Top View Side View Side View 103.4 78 12.7 100 1.7 2-R1.75 20 25 40 2-M3 L=4 Side View 48 . 9 FAN 10CFM 5cm Air flow direction 30.8 62 CN2 3 2 1 2-M3 L=2 CN1 CN101 2 1 24.2 1 2 3 4 5 6 55 Bottom View File Name:RPS-200-SPEC 2020-10-20 200W Reliable Green Medical Power Supply RPS-200 series AC Input Connector (CN1) : JST B3P-VH or equivalent Pin No. 1 2 3 Assignment AC/L No Pin AC/N Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent DC Output Connector (CN2) : JST B6P-VH or equivalent Pin No. 1,2,3 4,5,6 Assignment +V -V Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent FAN Connector(CN101) : JST B2B-PH-K-S or equivalent Pin No. 1 2 Assignment DC COM +12V Mating Housing Terminal JST PHR-2 or equivalent JST SPH-002T-P0.5S or equivalent Note : 1. The FAN supply is designed to serve as the source of the additive external fan for the cooling of the power supply, enabling the full load delivery and assuring the best life span of the product. Please do not use this FAN supply to drive other devices. 2.The PCB type(Blank type)EMI Conduction for Class B. Radiation for Class B with FG(Class) and Class A without FG(Class) 3.The enclosed type(-C type) model is not suitable for the configuration within a Class (no FG) system but is suggested to used within a Class (with FG) system. Installation Manual Please refer to : http://www.meanwell.com/manual.html File Name:RPS-200-SPEC 2020-10-20