Compare Pressure Transducer I and II Page 1 of 2 honeywell.com/aero Home Products Applications Industries FAQ Links Precision Pressure Transducer Comparison FEATURES Other Languages Contact Us PPT PPT2 Total Digital Error Band Over Temperature 0.10% FS MAX, 0.05%FS TYP 0.075% FS MAX, 0.0375 FS TYP Total Analog Error Band Over Temperature 0.12% FS MAX, 0.06%FS TYP (12-bit DAC) 0.09% FS MAX, 0.045 FS TYP (16 -bit DAC) Operating Temp Range -40 to 85C Standard: -40 to 85C, Extended: -55 to 110C Storage Temp Range -55 to 90C Standard: -50 to 100C, Extended: -60 to 125C Display Resolution e.g. 20 psia: XX.XXX e.g. 20 psia: XX.XXXX ** Digital Noise Resolution 0.005%FS at 120 readings/sec 0.005%FS at 250 readings/sec Analog Resolution 1.22 mV steps (12 bits) 0.15 mV steps, typical (15+ bits) Maximum Sample Rate Up to 120 Readings/Sec Up to 1000 Readings/Sec Minimum Response Delay 17 ms 2 ms Baud Rate Up to 28,800 Up to 115,200 Literature EN 61326-1: 2006, Industrial Level RF Susceptibility EN 61326-1:2006, Basic Level 3 V/m (80 MHz to 1 GHz) 3 V/m (1.4 GHz to 2 GHz) 1 V/m (2.0 GHz to 2.7 GHz) 10 V/m (80 MHz to 1 GHz) 3 V/m (1.4 GHz to 2 GHz) 1 V/m (2.0 GHz to 2.7 GHz) RTCA/DO-160F, Section 20.4, Category R 10 V/m (100 MHz to 400 MHz) 150 V/m (400 MHz to 8 GHz) RoHS Compliant No Yes Electrical Connector Plastic and Metal Metal only Power Supply 5.5 to 30 VDC Pressure Connections Brass Brass and SST Size (excluding connector and fittings) 2.45 x .975 x 1.80 in 2.47 x .97 x 1.56 in Weight (excluding fittings) 5.0 oz. (142gm) 4.4 oz. (125 gm) 6 to 34 VDC (with reverse polarity protection) ** 1 extra digit to right of decimal place for all pressure ranges and display units. Extra digit may be turned off by user for PPT compatibility Honeywell Aero | Terms & Conditions | Privacy Statement | Contact Us | (c) Copyright Honeywell International Inc. 2013 http://pressuresensing.com/ppt2-comparison.php 10/16/2013 Compare Pressure Transducer I and II Page 2 of 2 http://pressuresensing.com/ppt2-comparison.php 10/16/2013