This supplement contains information necessary to ensure the accuracy of
the above manual. Enter the corrections in the manual if either one of the
following conditions exist:
1. The revision letter stamped on the indicated PCA is equal to or higher
than that given with ea ch change.
2. No revision letter is indicated at the beginning of the change.
1520 Calibration Manual Su pp lem ent
10/01 1
Change #1
On page 15, in the
Replacement Parts and Optional Accessori es
table,
Replace:
Fuse, 6 mm x 32 mm (0.25 x 1.25 in), 0.5 A, 750 V, Fast Acting 1556096
With:
W Fuse, 6 mm x 32 mm (0.25 x 1.25 in), 0.5 A, 660 V, Fast Acting, 50 A
Minimum Inter rupt r atin g OR
NA
W Fuse, 6 mm x 32 mm (0.25 x 1.25 in), 0.5 A, 750 V, Fast Acting 1556096
W For safety, use exact replacement only.
Change #2
On page 8, under
Discharge Circuit Test
, replace the last sentence in step 3 with the
following:
Check that the DMM reading for each position is between 2600 and 3600e.