http://www.tracopower.com
DC/DC Converters
THB 3 Series, 3 Watt
Features
Supplementary and reinforced insulation
I/O isolation 4800 VACrms rated for
1000 Vrms (1410 Vpk) working voltage
Medical safety to ES 60601-1 and
IEC/EN 60601-1 3rd edition, 2 x MOOP
Industrial safety to IEC/EN/UL 60950-1
Isolation test voltage 6000 Vpk
Wide 2:1 input voltage ranges
Extended operating temperature range
–40°C to 85°C max.
Input filter meets EN55022, class A
Continuous short-circuit protection
High reliability
3-year product warranty
The THB-3 series is a range of high performance, regulated DC/DC converters in
a DIP-24 plastic package. A reinforced I/O-isolation system and a wide 2:1 input
voltage range make this product the best choice for many demanding applications
like transportation systems, industrial controls, medical equipments, instrumenta-
tions, everywhere where high basic-, supplementary- or reinforced insulation is
required to meet requested safety standards.
A high efficiency allows safe operation in a temperature range of –40°C to +85°C.
Other features of this product are over voltage protection and internal EMI-input
filter to meet EN 55022, class A without additional components. Full SMD-design
with exclusive use of ceramic capacitors ensure a very high reliability and a long
product lifetime.
Order code Input voltage range Output voltage Output current max. Efficiency typ.
THB 3-0511 5 VDC 600 mA 70 %
THB 3-0512 12 VDC 250 mA 75 %
THB 3-0515 4.5 – 9 VDC 24 VDC 125 mA 76 %
THB 3-0522 (5 VDC nominal) ±12 VDC ±125 mA 75 %
THB 3-0523 ±15 VDC ±100 mA 75 %
THB 3-1211 5 VDC 600 mA 74 %
THB 3-1212 12 VDC 250 mA 80 %
THB 3-1215 9 – 18 VDC 24 VDC 125 mA 81 %
THB 3-1222 (12 VDC nominal) ±12 VDC ±125 mA 80 %
THB 3-1223 ±15 VDC ±100 mA 80 %
THB 3-2411 5 VDC 600 mA 78 %
THB 3-2412 12 VDC 250 mA 83 %
THB 3-2415 18 – 36 VDC 24 VDC 125 mA 84 %
THB 3-2422 (24 VDC nominal) ±12 VDC ±125 mA 83 %
THB 3-2423 ±15 VDC ±100 mA 83 %
THB 3-4811 5 VDC 600 mA 78 %
THB 3-4812 12 VDC 250 mA 83 %
THB 3-4815 36 – 75 VDC 24 VDC 125 mA 84 %
THB 3-4822 (48 VDC nominal) ±12 VDC ±125 mA 83 %
THB 3-4823 ±15 VDC ±100 mA 83 %
Models
Page 1 of 3
C B
Scheme
UL 60950 -1
UL 60601-1
UL 60950 -1
IEC 60950 -1
IEC 60601-1
http://www.tracopower.com
DC/DC Converters
THB 3 Series 3 Watt
Input Specifications
Input current at no load 5 Vin models: 40 mA typ.
12 Vin models: 30 mA typ.
24 Vin models: 20 mA typ.
48 Vin models: 10 mA typ.
Start-up voltage / under voltage shut down 5 Vin models: 4.5 VDC / 4.0 VDC (or lower)
12 Vin models: 9 VDC / 8.5 VDC (or lower)
24 Vin models: 18 VDC / 17 VDC (or lower)
48 Vin models: 36 VDC / 34 VDC (or lower)
Recommended external input fuse (slow blow) 5 Vin models: 2.0 A
12 Vin models: 1.0 A
24 Vin models: 0.5 A
48 Vin models: 0.25 A
Surge voltage (1 sec. max.) 5 Vin models: 11 VDC max.
12 Vin models: 25 VDC max.
24 Vin models: 50 VDC max.
48 Vin models: 100 VDC max.
Input filter EN 55022 class A, FCC part 15, class A
Output Specifications
Voltage set accuracy ±1 %
Regulation – Input variation Vin min. to Vin max. 0.5 % max.
– Load variation 25 – 100 %: single output models: 1.0 % max..
dual output models: 2.0 % max. balanced load
Minimum load 15 % of rated max. output current.
(Operation at lower load is safe but major
deviations to specifyed data may occur)
Ripple and noise (20 MHz Bandwidth) 5 VDC models: 100 mVpk-pk max.
other models: 150 mVpk-pk max.
Transient response (25% load step change) 150 µs typ.
Current limitation >120 % Iout max.
Short circuit protection continuous (automatic recovery)
Capacitive load 5 VDC output models: 1000 µF max.
other single output models: 470 µF max.
dual output models: 220 µF max. (each output)
Isolation / Safety Standards
Isolation test voltage (end of line flash test 1 sec.) 6000 Vpk
I/O isolation voltage – according IEC/EN 60601-1 3000 VACrms, rated for 300 Vrms working
(50Hz, 60sec.) voltage, 2 x MOOP
– according IEC/EN 60950-1 4800 VACrms, rated for 1000 Vrms (1410 Vpk)
working voltage
Leakage current 2 µA (at 240VAC, 60Hz)
Isolation capacitance – Input/Output 13 pF max. (at 100KHz, 1V)
Isolation resistance – Input/Output >10 Gohm (at 500VDC)
Safety standards IEC 60950-1:2005 (2nd ed.) +A1:2009 and/or
EN 60950-1:2006 +A1:2010 +A11:2009
+12:2011, UL 60950-1 (2nd ed.)
CSA C22.2 No. 60950-1-03
IEC 60601-1 3rd edition, 2 x MOOP,
EN 60601-1 + A11:2011
ES 60601-1, CSA C22.2 No. 601.1
– Certification documents www.tracopower.com/overview/thb3
All specifications valid at nominal input voltage, full load and +25°C after warm-up time unless otherwise stated.
Page 2 of 3
Specifications can be changed without notice! Make sure you are using the latest documentation, downloadable at www.tracopower.com
www.tracopower.com
DC/DC Converters
THB 3 Series 3 Watt
Outline Dimensions
Rev. July 11, 2018
Pin Single Dual
1+Vin (Vcc) +Vin (Vcc)
11 No pin Common
12 –Vout No pin
13 +Vout –Vout
15 No pin +Vout
23 –Vin (GND) –Vin (GND)
24 –Vin (GND) –Vin (GND)
Pin-Out
Dimensions in [mm], () = Inch
Pin diameter ø 0.6 ±0.05 (0.024 ±0.002)
Tolerances ±0.25 (±0.001)
Pin pich tolerances ±0.13 (±0.005)
General Specifications
Temperature ranges Operating –40°C to +85°C
– Casing +95°C max.
– Storage –50°C to +125°C
Derating 3.3 %/K above 70°C
Humidity (non condensing) 95 % rel H max.
Temperature coefficient ±0.02 %/K typ.
Reliability, calculated MTBF (MIL-HDBK-217F, at +25°C, ground benign) >1 Mio. h
Switching frequency 150 kHz typ. (puls width modulation)
Casing material non conductive plastic (UL 94V-0-rated)
Potting material Silicone TSE 3331 (UL 94V-0-rated)
Pin material Copper alloy with goldplated nickel subplate
Weight 13.0 g (0.46 oz)
Soldering temperature max. 26C / 10 sec.
Altitude during operation up to 5‘000 m (16400 ft) approved
Environmental compliance – Reach www.tracopower.com/products/reach-declaration.pdf
– RoHS according RoHS directive 2011/65/EU
Page 3 of 3
31.8
2x2.54
20.322.54 2.0
0.2
10.53.9
(1.25)
(0.8)
(0.1) (2x0.1) (0.08)
(0.41)
(0.15)
(0.008)
2.54
20.3
15.2
(0.8)
(0.6)
(0.1)
Bottom view
111 12
13152324
Supporting documents : www.tracopower.com/overview/thb3
- The component is not be used in an oxygen rich environment.
- The component is not to be used in conjunction with flammable anaesthetics and agents.
- The component has to be disposed appropriately. Please refer to local regulations (Waste Electrical and Electronic Equipment).
- A modification of the component is not allowed.