DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
1
of 5
DC Voltage Simple Logger®
Models L410 & L430
ORDERING INFORMATION CATALOG NO.
Simple Logger®Model L410 (DC Voltage, 0 to 100mVDC Input) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2114.05
Simple Logger®Model L430 (DC Voltage, 0 to 10VDC Input). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2114.07
All DC voltage Simple Loggers®are supplied with software (CD-ROM), 6 ft DB-9 RS-232 serial cable, 9V Alkaline battery
and user manual
Accessories
Voltage Leads, (2) with clips, 5 ft . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2118.51
Features
Two models to choose from:
Model L410
0 to 100mVDC
Model L430
0 to 10VDC
Programmable scales via software
Programmable units via software
Inputs via color-coded safety
banana jacks
RS-232 output
One button operation
Accuracy : ±1% of Reading +
Resolution
Double Insulation
CE Mark
Applications
Circuit design troubleshooting
Sensor monitoring
Battery testing
Power supply profiling
Specifications
ELECTRICAL
Number of Channels: One
Measurement Range:
0 to 100mVDC (L410)
0 to 10VDC (L430)
Input Impedance: L410 and L430: 1MΩ
Input Connection:
Recessed safety 4mm banana jacks
Resolution: 8 Bit
L410: 50μV max
L430: 5mV max
Accuracy*: ±1% of Reading + Resolution
(see inside back cover for details)
Sample Rate: 4096/hr maximum
Storage Technique:
Time Extension Recording(TXR)
Data Storage: 8192 readings
Power: 9V Alkaline battery
Battery Life:
Up to one year of recording @ 25°C
Output: RS-232 via DB-9 connector
Indicators: Red LED flashes once
every two seconds in STANDBY mode,
twice every two seconds in LOGGING
mode, and is off in OFF mode
Controls: One button used to start and
stop recording sessions and to turn the
data logger on and off
Working Voltage: 30VAC maximum
MECHANICAL
Size: 27/8x 25/16 x 15/8"
(73 x 59 x 41mm)
Weight: (with battery) 5 oz (140g)
Mounting: Base plate mounting holes
or Velcro®pads supplied loose
ENVIRONMENTAL
Operating Temperature:
-4 to +158°F (-20 to +70°C)
Storage Temperature:
-4 to +174°F (-20 to +80°C)
Relative Humidity:
5 to 95% non-condensing
Safety: EN 61010, 30V Cat. III
DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
2
of 5
Simple Logger®Software
Features
Very easy to setup and use —
no programming required
Statistics appear on
screen: Max, Min, Avg,
Date and Time
On-Line display of data
logger values
Zoom capability with automatic
statistics update
User keyed text annotations
Automatically displays values
at the cursor position
Export data to spreadsheet
and database software
Print both graphs and
tabular listings
Paste graphs into other
applications
Programmable scale and units
for all DC loggers. Automatic
selection for AC loggers
User programmable High and
Low limits (Model L215)
User programmable record
modes – inside or outside of
limits (Model L215)
Automatic baud rate selection
(version 6.0 or later)
Computer
Requirements
486 or higher
8MB of RAM
Windows®95/98/2000/ME/XP
or Windows®NT
One serial port not in use by
any other device
8MB hard drive space for
program and 400kB max
per file stored
Plot graphs and view statistics quickly and easily when you download
data from Simple Logger ®to your computer. Scale and time axis
update automatically.
Click and drag your mouse to zoom up the desired area of the graph
for better viewing. The Min, Max and Average statistics at the bottom
of the graph update automatically with zoom window.
Click the Annotate button to add a title and descriptive information to
your graph. This information can be stored and printed with a graph.
DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
3
of 5
Plot area with value and time data
shown on the X and Y axis
Click the Annotate button to add a title and descriptive
information to your graph. This information can be stored
and printed with a graph.
On-line display of data
Annotate
button
Automatic update of
Y data at
cursor position
Graph
Statistics
DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
4
of 5
Simple Logger®Software
Click and drag your mouse to zoom up the desired area of the
graph for better viewing. The Min, Max and Average statistics
at the bottom of the graph update automatically when the
graph is zoomed.
Click the scale button in the file menu to
create custom scales and engineering units
for DC voltage (Models L410 & L430)
and DC current loggers (Model L320).
DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
5
of 6
Accessories
Current Probe and Simple Logger®Compatibility
Series Probe Ratio Measurement Range Output Signal Output Catalog Logger
Model AC DC Current Voltage Connection No. Model
MN211 1000:1 0.5 to 240A 1mAAC/AAC Lead 2115.73 L110
MN213 1000:1 0.5 to 240A 1mAAC/AAC Lead 2115.75 L110
MN251 0.1 to 240A 1mVAC/AAC Lead 2115.77 L100
MD301 2 to 500A 1mVDC/AAC Lead 1201.07 L430
MD303 1000:1 4 to 500A 1mAAC/AAC Lead 1201.21 L110
MD305 1000:1 1 to 600A 1mAAC/AAC Lead 1201.36 L110
MD314 4 to 500A 1mVAC/AAC Lead 2110.75 L100
MR410* 1 to 400A 1mV/AAC Lead 1200.70 L100 (AC)
1 to 600A 1mV/ADC Lead 1200.70 L430 (DC)
0.2 to 40A 10mV/AAC Lead 1200.68 L100 (AC)
0.5 to 400A 0.5 to 600A 1mV/AAC
MR411*
0.4 to 60A 10mV/ADC Lead 1200.68 L430 (DC)
0.5 to 600A 1mV/ADC
MR520* 1 to 1000A 1mV/AAC/DC Lead 1200.71 L100 (AC)
1 to 1500A 1mV/AAC/DC L430 (DC)
0.2 to 100A 0.4 to 150A 10m V/AAC Lead 1200.69 L100 (AC)
0.5 to 1000A 0.5 to 1500A 1mV/AAC
MR521*
0.2 to 100A 0.4 to 1500A 10mV/ADC Lead 1200.69 L430 (DC)
0.5 to 1000A 0.5 to 1500A 1mV/ADC
SR604 1000:1 0.1 to 1000A 1mAAC/AAC Lead 2113.44 L110
SR652 0.1 to 1000A 1mVAC/AAC Lead 2113.46 L100
SR704 1000:1 1mA to 1000A 1mAAC/AAC Lead 2116.30 L110
SR752 0.1 to 1000A 1mVAC/AAC Lead 2116.32 L100
*Caution: Do not use DC loggers with AC probes or AC loggers with DC Probes
Not all models are UL approved; please consult factory.
ø = 0.78"
ø = 1.18"
ø = 1.6"
ø = 2.05"
Split Core AC and DC Current Probes
ø = 1.18"
DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
6
of 7
2111.31
Safety Grip Probes, set of two
2114.26
Simple Logger®Software Upgrade for all
models (Receive software upgrade
notices when warranty registration
is on file)
DC Voltage Simple Logger®Models L410 & L430 (pdf) Rev. 03 12/05
Technical Assistance (800) 343-1391 www.aemc.com
7
of 7
Contact Us
United States & Canada:
Chauvin Arnoux®, Inc.
d.b.a. AEMC®Instruments
200 Foxborough Blvd.
Foxborough, MA 02035 USA
(508) 698-2115 • Fax (508) 698-2118
www.aemc.com
Customer Support – for placing an order, obtaining price & delivery:
customerservice@aemc.com
Sales Department – for general sales information:
sales@aemc.com
Repair and Calibration Service – for information on repair & calibration, obtaining a user manual:
repair@aemc.com
Technical and Product Application Support – for technical and application support:
techinfo@aemc.com
Webmaster – for information regarding www.aemc.com:
webmaster@aemc.com
South America, Central America, Mexico, Caribbean, Australia & New Zealand:
Chauvin Arnoux®, Inc.
d.b.a. AEMC®Instruments
15 Faraday Drive
Dover, NH 03820 USA
(978) 526-7667 • Fax (978) 526-7605
export@aemc.com
www.aemc.com
All other countries:
Chauvin Arnoux SCA
190, rue Championnet
75876 Paris Cedex 18, France
33 1 44 85 45 28 • Fax 33 1 46 27 73 89
info@chauvin-arnoux.com
www.chauvin-arnoux.com