DC POWER SUPPLIES
A Greater Measure of Condence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
Series 2280S
Precision Measurement, Low Noise,
Programmable DC Power Supplies
The Series 2280S Precision Measurement, Low
Noise, Programmable DC Power Supplies are
much more than just sources of clean power;
they are also precision measurement instru-
ments. They can source stable, low noise voltag-
es as well as monitor load currents over a wide
dynamic range from amps to nanoamps. The
Model 2280S-32-6 can output up to 32V at up to
6A; the Model 2280S-60-3 can output up to 60V
at up to 3.2A.
Both supplies use linear regulation to ensure
low output noise and superior load current
measurement sensitivity. A high resolution color
Thin Film Transistor (TFT) screen displays a
wide range of information on measurements.
Soft-key buttons and a navigation wheel combine
with the TFT display to provide an easy-to-navigate user interface that speeds instrument setup and
operation. In addition, built-in plotting functions allow monitoring trends such as drift. These sup-
plies provide the flexibility required for both benchtop and automated test system applications. For
example, they provide a list mode, triggers, and other speed optimization functions to minimize test
time in automated testing applications.
DMM-Quality Low Current Measurements with High Resolution
Unlike conventional power supplies, Series 2280S supplies can also make measurements with up to
6½ digits of resolution. Voltage output measurements can be resolved down to 100µV. These supplies
measure load currents from 100nA to amps and can resolve down to 10nA. Four load current mea-
surement ranges (10A, 1A, 100mA, and 10mA) support measuring a device’s full load current, standby
mode current, and small sleep mode currents with DMM-quality accuracy. The high resolution allows
characterizing small changes in load currents with confidence. It also makes it possible to make a
broad range of measurements on a single range with excellent accuracy across both low and high
current values.
Measure Rapidly Changing Loads
To monitor fast-changing and pulse-like load currents properly, Series 2280S supplies offer the speed
necessary to capture load changes that occur at intervals as short as 140µs. This capability allows
designers and manufacturers of portable, battery-operated devices to monitor load currents easily in
all of a device’s operating modes so they can determine the device’s total power consumption.
This high speed measurement capability allows measuring each state of a power-up load sequence
and a power-down sequence. Measurements can be made as fast as 2500 readings per second, mak-
ing it possible to characterize and test the current draw at each of the start-up states.
Trigger
Delay Measure
Trigger
Delay Measure
Figure 1. Make time-critical measurements on fast-changing or pulse-like loads. An external
trigger initiates the acquisition. Programmable delay and measure times enable measurements
at a specific time on the load current pulse.
Monitor load currents from
100nA to 6A with high accuracy
Measure voltage and current
with 6½-digit resolution
Capture dynamic load currents
as short as 140µs
Output up to 192W of low
noise, linear regulated power
Programmable rise and fall
times eliminate voltage
overshoot and undershoot
transients
Built-in graphing simplifies
analyzing trends or displaying
voltage or current waveforms
High resolution TFT display
and soft-key/icon-based user
interface simplify power supply
operation
Programmable output
sequences reduce test times
Sink up to 0.45A to discharge
voltage quickly
Digital I/O for direct
communication with other
devices and instruments
GPIB, USB, and LAN interfaces
Built-in web page simplifies
automated control/monitoring
Automate tests easily with
KickStart start-up software
Precision measurement, low noise, programmable DC power supplies
Precision measurement, low noise, programmable DC power supplies
DC POWER SUPPLIES
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Condence A Tektronix Company
Figure 2. Measure the start-up load currents as a device powers up.
View and Control Every Parameter of the Supply for Maximum Performance
The bright, 4.3-inch TFT display shows voltage and current readings, the source settings and many
additional settings in large, easy-to-read characters. The icon-based main menu provides all the
functions users can control and program for fast access to source setup, measurement setup, display
formats, trigger options, and system settings. Menus are short, and the menu options you need are
easy to find and clearly described, so that you can set up test parameters quickly by using the naviga-
tion wheel, keypad, or soft-keys. Many set-up parameters, such as for voltage and current settings, can
be entered directly from the home screen; less complex tests don’t even require accessing the main
menu to make adjustments. Whether your test’s requirements are uncomplicated or complex, the
Series 2280S supplies provide a simple way to set up all the required parameters.
Ordering Information
2280S-32-6 Precision
Measurement DC
Power Supply, 32V, 6A
2280S-60-3 Precision
Measurement DC
Power Supply,
60V, 3.2A
Accessories Supplied
Quick Start Guide
KickStart Quick Start Guide
User Documentation CD
LAN Crossover Cable
Power Cord
Rear Panel Mating
Connector with Cover
Series 2280S
Precision Measurement, Low Noise,
Programmable DC Power Supplies
ACCESSORIES AVAILABLE
2280-001 Rear Panel Mating Connector and Cover
CA-180-3A LAN Crossover Cable
USB-B-1 USB Cable Type A to B, 1m (3.3 ft)
2450-TLINK Trigger Link cable to connect 2280S digital
I/O to Trigger Link I/O on other Keithley
instruments
4299-8 Single Fixed Rack-Mount Kit
4299-9 Dual Fixed Rack-Mount Kit
4299-10 Dual Fixed Rack-Mount Kit for one 2U
Graphical Display Instrument and one
Series 26xx Instrument
4299-11 Dual Fixed Rack-Mount Kit for one 2U
Graphical Display Instrument and one
Series 24xx, Series 2000, or 2U Agilent
Instrument
KPCI-488LPA IEEE-488.2 Interface Board for the PCI Bus
KUSB-488B IEEE-488.2 USB-GPIB Interface Adapter for
USB Port with 2m (6.6 ft) cable
7007-05 Double Shielded Premium IEEE-488
Interface Cables, 0.5m (1.6 ft)
7007-1 Double Shielded Premium IEEE-488
Interface Cables, 1m (3.2 ft)
7007-2 Double Shielded Premium IEEE-488
Interface Cables, 2m (6.5 ft)
7007-3 Double Shielded Premium IEEE-488
Interface Cables, 3m (10 ft)
7007-4 Double Shielded Premium IEEE-488
Interface Cables, 4m (13 ft)
Figure 3. Adjust voltage,
current, the current range,
the measurement mode,
protection levels, and
other functions from
the home screen.
Figure 4. Access the full
functionality of the Series
2280S supplies from the
main menu.
External Triggers
start measurements
sleep
standby
active
Measurement
Complete signals
Precision measurement, low noise, programmable DC power supplies
Precision measurement, low noise, programmable DC power supplies
DC POWER SUPPLIES
A Greater Measure of Condence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
Series 2280S Precision Measurement, Low Noise,
Programmable DC Power Supplies
Figure 5. Configure the readback measurements exactly the way
you need them using a Measure Sub-screen.
Controlled Rise and Fall Times Protect Devices Under Test
Are you working on a device or a system with a high in-rush current? Series
2280S supplies allow programming the voltage output’s rise time to slow
the voltage ramp and avoid voltage overshoot, which could potentially
damage the DUT. The voltage fall time can also be controlled to prevent a
fast ramp down of the output voltage. Voltage rise and fall times can be set
as slow as 10V/s or as fast as 100V/s. Small voltage transitions can be pro-
grammed to rise and fall as fast as 1000V/s.
Figure 6. Programmable slew rates permit controlling voltage rise and
fall rates, preventing undesirable transients during power up
of capacitive loads.
Perform Semi-Automated Testing Right on Your Bench
Test your design over its operating voltage range automatically or study
how it responds to DC output changes by using the Series 2280S supplies’
built-in List Mode function. Create and save up to 10 lists of sequenced
voltage levels with up to 99 distinct voltages in each list. The duration of
each voltage level can be set as short as 1ms or as long as 60s. Create sim-
ple linear ramps or any custom configuration from either the front panel
or an interface bus. A single trigger will automatically execute the list once
or multiple times.
Figure 7A. Create an output sequence using the List setup screen.
Figure 7B. Use the list mode to step the output voltage automatically
through a series of levels. The voltages were created by the list shown
in Fig. 7A.
Model 2280S with
slew rate control
power supply with no
slew rate control
Precision measurement, low noise, programmable DC power supplies
Precision measurement, low noise, programmable DC power supplies
DC POWER SUPPLIES
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Condence A Tektronix Company
Series 2280S Precision Measurement, Low Noise,
Programmable DC Power Supplies
Trend Analysis and Load Current Characterization
Use the built-in graphing function to monitor the stability of the load
current or capture and display a dynamic load current. Or, use it to view
a start-up or turn-off load current. Series 2280S supplies can take mea-
surements quickly and store up to 2,500 measurement points. In addition
to displaying a voltage or current waveform or both, they can compute
statistics on the stored data. Statistical calculation options include average,
maximum, minimum, peak-peak, and standard deviation. All the informa-
tion you need is accessible through a few display menus.
Figure 8. Monitor voltage, current, or both with the graphing function.
Automate a test quickly and easily with KickStart software
Set up an automated test for the Series 2280S supplies in four simple steps
with the KickStart start-up software. KickStart is available for installation
on your PC via a free download from www.keithley.com. Once the program
is open, perform these steps: create a test file, select the 2280S power sup-
ply, select the power supply application, and enter the test parameters. Run
the test, and get results without having to write a line of code. KickStart
provides both graph and spreadsheet displays. Data is saved to the test file,
and the data can be exported to data analysis programs. Or take a screen
shot of the plot for reproduction in test reports. Up to 15,000 data points
can be stored. KickStart may be all the software you ever need.
Figure 9. KickStart DC power main screen.
Figure 10. KickStart plot with measurement table.
Optimized Performance for Production Test
Series 2280S supplies are as powerful in a production test system as
they are flexible on the R&D benchtop. In addition to their List Mode
function, several other Series 2280S features help minimize test time in
automated systems. For example, an external trigger input allows hardware
synchronization and control by other instruments in the test system. To
eliminate many system delays, Series 2280S supplies produce a “mea-
surement complete output” to signal the test system when the supplies
have completed their required actions. To reduce measurement time, the
reading speed can be increased by reducing the acquisition time from
power line cycle integration times, 16ms at 60 Hz (20 ms at 50 Hz), to
33µs (40µs). Furthermore, the Series 2280S supplies can sink up to 0.45A.
Thus, these supplies can quickly discharge a voltage level and change to a
different voltage.
Figure 11. Series 2280S rear panel showing the rear output connector
with remote sense inputs, digital I/O, and the three interfaces (GPIB,
USB, and LAN).
A choice of front or rear panel terminals provides enhanced connection
flexibility. For maximum voltage accuracy, 4-wire remote sensing ensures
that the output voltage programmed is actually the level applied to the
load. In addition, the sense lines are monitored in order to detect any
breaks in them. These features ensure any production problems can be
quickly identified and corrected. Four additional digital I/O pins can pro-
Precision measurement, low noise, programmable DC power supplies
Precision measurement, low noise, programmable DC power supplies
DC POWER SUPPLIES
A Greater Measure of Condence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
Series 2280S Precision Measurement, Low Noise,
Programmable DC Power Supplies
Specifications
23°C ±5°C with 1-hour instrument warm-up.
DC OUTPUT RATINGS
2280S-32-6 2280S-60-3
Voltage 0 to 32 V 0 to 60 V
Current 0 to 6 A 0 to 3.2 A
Maximum Power 192 W 192 W
VOLTAGE 1
SOURCE SETTING
2280S-32-6 2280S-60-3
Accuracy ± (0.02% + 3 mV) ± (0.02% + 6 mV)
Resolution 1 mV 1 mV
MEASUREMENT
2
(0.5V over-range)
2280S-32-6 2280S-60-3
Accuracy ± (0.02% + 2 mV) ± (0.02% + 4 mV)
Resolution 0.1 mV 0.1 mV
ADDITIONAL OFFSET AT FASTER MEASUREMENT SETTINGS
2280S-32-6 2280S-60-3
5½ (0.1 PLC) 0.21 mV 0.40 mV
4½ (0.01 PLC) 1.44 mV 2.7 mV
3½ (0.002 PLC) 7.60 mV 14.25 mV
REGULATION
2280S-32-6 2280S-60-3
Load ±(0.01% + 2 mV) ±(0.01% + 2 mV)
Line ±(0.01% + 1 mV) ±(0.01% + 1 mV)
OUTPUT RIPPLE AND NOISE
2280S-32-6 2280S-60-3
Bandwidth 20Hz–20MHz < 1mVRMS < 2mVRMS
< 5mV p-p < 7mV p-p
LOAD TRANSIENT RECOVERY TIME: Resistive load change 50% load to 100% load or 100% load to 50% load: <50µs to within
15mV of V-set.
SLEW RATE: Rising voltage and falling voltage: 10V/s to 100V/s. Up to 1000V/s under limited conditions 3. 100V/s (default).
MAXIMUM SOURCE VOLTAGE DROP PER LEAD: To maintain specified voltage accuracy: 1V.
MAXIMUM SENSE HI AND SENSE LO LEAD RESISTANCE: To maintain specified voltage accuracy: 2.
CURRENT
CURRENT LIMIT SETTING
2280S-32-6 2280S-60-3
Full-scale Amps 6 A 3.2 A
Accuracy ±(0.05% + 5 mA) ±(0.05% + 5 mA)
Resolution 0.1 mA 0.1 mA
MEASUREMENT 4 (120% over-range except 10A)
Range Resolution
2280S-32-6 2280S-60-3
10 mA 10 nA ±(0.05% + 10 µA) ±(0.05% + 10 µA)
100 mA 100 nA ±(0.05% + 10 µA) ±(0.05% + 10 µA)
1 A 1 µA ±(0.05% + 250 µA) ±(0.05% + 250 µA)
10 A 10 µA ±(0.05% + 250 µA) ±(0.05% + 250 µA)
vide fault status outputs or control an external
relay or status lamp. Series 2280S supplies can
be controlled via their built-in GPIB, USB, or
LAN interfaces. The USB interface is test and
measurement system (TMC) compliant. The LXI
Core compliant LAN interface supports con-
trolling and monitoring a Series 2280S supply
remotely, so test engineers can always access
the power supply and view measurements, even
if theyre located on a different continent than
their test systems.
To facilitate production test software develop-
ment, native National Instruments LabVIEW™
drivers, IVI-C and IVI-COM drivers are available
on the Keithley website, www.keithley.com.
Series 2280 specifications
Series 2280S specifications
DC POWER SUPPLIES
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Condence A Tektronix Company
Series 2280S Precision Measurement, Low Noise,
Programmable DC Power Supplies
ADDITIONAL OFFSET AT FASTER MEASUREMENT SETTINGS10
Measure Resolution
and (NPLC) Range
2280S-32-6 2280S-60-3
5½ (0.1 PLC)
10 mA 5.0 µA 5.0 µA
100 mA 20 µA 20 µA
1 A 80 µA 80 µA
10 A 2.0 mA 2.0 mA
4½ (0.01 PLC)
10 mA 20 µA 20 µA
100 mA 40 µA 40 µA
1 A 500 µA 500 µA
10 A 10 mA 10 mA
3½ (0.002 PLC)
10 mA 30 µA 30 µA
100 mA 250 µA 250 µA
1 A 25 mA 25 mA
10 A 75 mA 75 mA
CURRENT PULSE MEASUREMENT 5
2280S-32-6 2280S-60-3
Minimum Pulse Width (10mA and 100mA range) 6 2.1 ms 2.1 ms
Minimum Pulse Width (1A and 10A range) 6 140 µs 140 µs
Minimum time to capture two consecutive pulses 0.5 ms 0.5 ms
REGULATION
2280S-32-6 2280S-60-3
Load ±(0.01% + 0.25 mA) ±(0.01% + 0.25 mA)
Line ±(0.01% ± 0.25 mA) ±(0.01% ± 0.25 mA)
OUTPUT RIPPLE AND NOISE
2280S-32-6 2280S-60-3
Bandwidth 20Hz–20MHz < 3mA RMS < 3mA RMS
MAXIMUM CONTINUOUS AVERAGE SINK CURRENT
2280S-32-6 2280S-60-3
Non-programmable 0.45 A ±15% 0.45 A ±15%
SYSTEM MEASUREMENT SPEEDS
Settings Concurrent ( V+I ) Current or Voltage (I or V)
Readings/s Measure Resolution
and (NPLC)
Autozero
On
60 Hz (50 Hz)
Autozero
OFF
60 Hz (50 Hz)
Autozero
On
60 Hz (50 Hz)
Autozero
OFF
60 Hz (50 Hz)
‘Read?’ with BUS Transfer
6½ (5 PLC) 2.0 (1.5) 5.4 (4.5) 2.5 (2.3) 9.0 (8.5)
6½ (1 PLC) 9.0 (8.0) 20 (18) 11.5 (9.5) 30.0 (28)
5½ (0.1 PLC) 48 (38) 50 (48) 50.0 (48.0) 95.0 (85.0)
‘*TRG and TRACe:DATa?’
into Memory
4½ (0.01 PLC) 7 440 (430) 1915 (1820)
3½ (0.002 PLC) 8 510 (510) 2668 (2650)
SERVICES AVAILABLE
2280S-32-6-EW 1 Additional Year of Factory
Warranty (total of 4 years)
2280S-32-6-5Y-EW 2 additional years of factory warran-
ty beyond the 3-Year factory warran-
ty (total of 5 years)
C/2280S-32-6-3Y-STD 3 Calibrations Within 3 Years of
Purchase
C/2280S-32-6-3Y-DATA 3 (ANSI-Z540-1 compliant)
Calibrations Within 3 Years of
Purchase
C/2280S-32-6-3Y-1702 3 (ISO 17025 Accredited)
Calibrations within 3 Years of
Purchase
C/2280S-32-6-5Y-STD 5 Calibrations Within 5 Years of
Purchase
C/2280S-32-6-5Y-DATA 5 (ANSI-Z540-1 compliant)
Calibrations Within 5 Years of
Purchase
C/2280S-32-6-5Y-1702 5 (ISO 17025 Accredited)
Calibrations within 5 Years of
Purchase
2280S-60-3-EW 1 Additional Year of Factory
Warranty (total of 4 years)
2280S-60-3-5Y-EW 2 additional years of factory warran-
ty beyond the 3-Year factory warran-
ty (total of 5 years)
C/2280S-60-3-3Y-STD 3 Calibrations Within 3 Years of
Purchase
C/2280S-60-3-3Y-DATA 3 (ANSI-Z540-1 compliant)
Calibrations Within 3 Years of
Purchase
C/2280S-60-3-3Y-1702 3 (ISO 17025 Accredited)
Calibrations within 3 Years of
Purchase
C/2280S-60-3-5Y-STD 5 Calibrations Within 5 Years of
Purchase
C/2280S-60-3-5Y-DATA 5 (ANSI-Z540-1 compliant)
Calibrations Within 5 Years of
Purchase
C/2280S-60-3-5Y-1702 5 (ISO 17025 Accredited)
Calibrations within 5 Years of
Purchase
Series 2280S specifications
DC POWER SUPPLIES
A Greater Measure of Condence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
Series 2280S Precision Measurement, Low Noise,
Programmable DC Power Supplies
PROTECTION
Overvoltage Protection (OVP) 2280S-32-6 2280S-60-3
Setting Accuracy ±(0.25% + 0.25 V) ±(0.25% + 0.5 V)
Resolution 125 mV 250 mV
Response Time 6 ms (typical)96 ms (typical)9
Overcurrent Protection (OCP) 2280S-32-6 2280S-60-3
Setting Accuracy ±(0.25% + 0.10 A) ±(0.25% + 0.10 A)
Resolution 25 mA 12.5 mA
Response Time 6 ms (typical)96 ms (typical)9
Overtemperature Protection (OTP) 2280S-32-6 2280S-60-3
Output turn-off Temperature >93°C (typical) >93°C (typical)
Response Time 6 ms (typical)96 ms (typical)9
OTHER TIMING DATA
CV TO CC TRANSITION TIME (V-Set = 5V, I-limit = 0.5A, Resistive Load change 25 to 2.5): 2.4ms.
CC TO CV TRANSITION TIME (V-Set = 5V, I-limit = 0.5A, Resistive Load change 2.5 to 25: 1.1ms.
FUNCTION CHANGE (from detection of bus command to function change completed): 10ms (typical).
OUTPUT OFF/ON (from detection of bus command to voltage beginning to decrease): 5ms (typical).
REVERSE LEADS ACTUATION: 6ms (typical)9.
GENERAL
COMMON MODE CURRENT: < 6µA peak-peak.
CHASSIS ISOLATION: ±240V, any terminal to chassis. >1G in parallel
with <6.8nF.
TEMPERATURE COEFFICIENT: Add the following to all accuracy specifications when
outside the range, 23°C ±5°C: (0.15 × specification)/°C for 0° to 18°C and 28° to 40°C.
MEASUREMENT DISPLAY MODES: Voltage and current, voltage only, current only.
MEASUREMENT ACQUISITION CONTROL: Continuous, Manual, External digital input,
PC bus.
LIST MODE: Maximum number of stored lists: 9.
Number of points in a list: 2 – 99.
List Storage Location: Internal memory or USB memory stick.
MATH AND FILTER FUNCTIONS:
REL: Removes offset from current reading display, Range: –1×106 to +1×10 6.
Mx+b: Reading = x, M = –1×106 to +1×10 6, b = –1×106 to +1×10 6.
Filter: Moving average, Count: 2–100, Window: 0.01% to 100%.
MEMORY BUFFER:
2500 locations; each location contains: Voltage measurement, current measurement,
CV/CC Mode, and time stamp, NVRAM.
DISPLAY: 4.3 in. front panel color display, resolution: 480 pixels × 272 pixels.
Display Modes:
Real time voltage and current readings and settings.
Plots of stored data: voltage vs data point, current vs data point, voltage and
current vs. data point, 100 point resolution.
Plots can also display statistics: mean, maximum, minimum, peak-peak,
standard deviation.
Table of stored data: time/date, voltage, current.
Soft button and navigation wheel control.
COMMUNICATIONS:
GPIB: IEEE-488.2 compliant and status model topology.
LAN: RJ-45 connector, 10/100BT, Auto MDIX.
IP Configuration: Static or DHCP.
LXI Core 2011, version 1.4,
USB: USB2.0 device (rear panel, type B), USBTMC compliant.
USB2.0 host (front panel, type A), full speed, support USB Flash Drives.
INPUT CONNECTIONS:
Front: (2-wire). Adjustable supporting, safety shrouded banana, spade lug, or wire.
Rear: (4-wire sense). 6-pin removal screw terminal, safety shrouded cover, removable
local sense jumpers.
REAL-TIME CLOCK: Capacitive charged, 20 days between next power on cycle at 23°C
and 50%RH.
DIGITAL I/O: 9-pin female D-sub. 6 Input/Output pins.
Input Signal Levels:
0.7V (maximum logic low).
3.7V (minimum logic high).
Input Voltage Limits:
0.25V (Absolute minimum).
+5.25V (Absolute maximum).
Maximum Source Current: +2.0mA@ >2.7V (per pin).
Maximum Sink Current: –50mA @ 0.7V (per pin, solid-state fuse protected).
5V power supply, limited to 0.5A @>4V (solid-state fuse protected).
Trig In minimum pulse s, Logic Low pulse.
Meter Ready Pulse, 15–3s, Logic Low Pulse.
EMC: Conforms to European Union EMC directive.
SAFETY:
U.S. NRTL Listing: UL61010-1 3rd ed 2012 and UL61010-2-030:2012.
Canadian Certification: CAN/CSA C22.2 No. 61010-1 3rd ed 2012.
European Union Compliance: Low Voltage Directive .
COOLING: Forced air, side intake and rear exhaust.
POWER SUPPLY: 100V/120V/220V/240V ±10%.
POWER LINE FREQUENCY: 50/60Hz ±3Hz, automatically sensed at power-on.
POWER CONSUMPTION: 630VA peak.
OPERATING ENVIRONMENT: 0° to 40°C, 80% RH up to 35°C, non-condensing.
Altitude: up to 2000 meters.
STORAGE ENVIRONMENT: –25° to 70°C.
LXI WEB BROWSER COMPATIBLE OPERATING SYSTEM AND SOFTWARE: Windows 2000,
Win 7 and XP compatible, supports Web browsers with Java plug-in (requires Java plug-in
1.7 or higher). Web page served by Model 2280S.
RACK DIMENSIONS: (D), without boot: 213.8 × 88.4 × 383.3mm (8.42 × 3.48 × 15.1 in.).
BENCH DIMENSIONS: (W×H×D) with boot: 255.3 × 107.2 × 415.0mm (10.1 × 4.22 × 16.34 in.)
SHIPPING WEIGHT: 13.29kg (29.3 lbs.).
NET WEIGHT: 10.85kg (23.9 lbs.).
WARR ANT Y: 3 years.
NOTES
1. Specifications based on using remote sense connections. For 2-wire
connections, add an offset of 0.5mV/A (Front terminals).
2. 6½ digit resolution, 1 PLC reading rate, filter on, autozero on.
3. 100V/s to 1000V/s rise and fall slew rates:
2280S-32-6: limited to 5V changes at a maximum of 3A; between 0V to
5V, the load current can be up to full load, 6A
2280S-60-3: limited to 10V changes at a maximum of 2A; between 0V
and 10V, the load current can be up to full load, 3.2A.
4. 6 ½ digit resolution, 1PLC integration time, Autozero: on, Filter: on
10mA and 100mA ranges: Source Delay: 2ms
1A and 10A ranges: Source Delay: 1ms.
5. Settings: Autozero: off, 0.002 PLC, Arm Source: External, Trigger
Source: Immediate, Filter: off
10mA and 100mA ranges: Source Delay: 2ms
1A and 10A ranges: Source Delay: 0ms or off
6. Time includes trigger detection, jitter, and 0.002PLC integration time.
7. Settings: Autozero: off, Output Delay: off, Source Delay: off, voltage
output is constant.
8. Settings: Autozero: off, Output Delay: off, Source Delay: off, voltage
output is constant, Measure Count: 1000.
9. Time defined as from detection of condition to start of output turn-off.
10. Filter on, 10mA and 100mA ranges: Source Delay: 2ms
1A and10A ranges: Source Delay: 1ms
Note: Specifications are subject to change without notice.
Series 2280S specifications
DC POWER SUPPLIES
Specifications are subject to change without notice. All Keithley trademarks and trade names are the property of Keithley Instruments.
All other trademarks and trade names are the property of their respective companies.
KEITHLEY INSTRUMENTS 28775 AURORA RD. CLEVELAND, OH 44139-1891 440-248-0400 Fax: 440-248-6168 1-888-KEITHLEY www.keithley.com
A Greater Measure of Condence
A Tektronix Company
BENELUX
+31-40-267-5506
www.keithley.nl
BRAZIL
55-11-4058-0229
www.keithley.com
CHINA
86-10-8447-5556
www.keithley.com.cn
FRANCE
+33-01-69-86-83-60
www.keithley.fr
GERMANY
+49-89-84-93-07-40
www.keithley.de
INDIA
080-30792600
www.keithley.in
ITALY
+39-049-762-3950
www.keithley.it
JAPAN
81-120-441-046
www.keithley.jp
KOREA
82-2-6917-5000
www.keithley.co.kr
MALAYSIA
60-4-643-9679
www.keithley.com
MEXICO
52-55-5424-7907
www.keithley.com
RUSSIA
+7-495-664-7564
www.keithley.ru
SINGAPORE
01-800-8255-2835
www.keithley.com.sg
TAIWAN
886-3-572-9077
www.keithley.com.tw
UNITED KINGDOM
+44-1344-39-2450
www.keithley.co.uk
For further information on how to purchase or to locate a sales partner please visit www.keithley.com/buy
© Copyright 2014 Keithley Instruments Printed in the U.S.A No. 3231 7.7.14