RPS-75 75W Single Output Medical Type series 1/2 Features : EUniversal AC input / Full range Low leakage current <100gA E Protections: Short circuit / Overload / Over voltage E Free air convection for rated power and 23.5CFM forced air E convection for peak load EMedical safety approved (2 x MOPP between primary to secondary)(Note.8) No load power consumption<0.75W E Fixed switching frequency at 65KHz E 3 years warranty E SPECIFICATION MODEL OUTPUT RPS-75-3.3 RPS-75-5 RPS-75-12 RPS-75-15 RPS-75-24 RPS-75-36 RPS-75-48 DC VOLTAGE 3.3V 5V 12V 15V 24V 36V 48V RATED CURRENT 15A 14A 6.3A 5A 3.2A 2.1A 1.6A CURRENT RANGE 0 ~ 20A 0 ~ 18.7A 0 ~ 8.3A 0 ~ 6.7A 0 ~ 4.2A 0 ~ 2.8A 0 ~ 2.1A RATED POWER 49.5W 70W 75.6W 75W 76.8W 75.6W 76.8W PEAK LOAD (23.5CFM) 66W 93.5W 99.6W 100.5W 100.8W 100.8W 100.8W 80mVp-p 120mVp-p 150mVp-p 240mVp-p 300mVp-p 300mVp-p 4.75 ~ 5.5V 11.4 ~ 13.2V 13.5 ~ 16.5V 22.8 ~ 27.6V 34.2 ~ 39.6V 45.6 ~ 52.8V O2.0% O1.0% O1.0% O1.0% O1.0% O1.0% RIPPLE & NOISE (max.) Note.2 80mVp-p 2.9 ~ 3.6V VOLTAGE ADJ. RANGE VOLTAGE TOLERANCE Note.3 O2.0% INPUT LINE REGULATION O0.5% O0.5% O0.5% O0.5% O0.5% O0.5% O0.5% LOAD REGULATION O1.5% O1.5% O1.0% O1.0% O1.0% O1.0% O1.0% SETUP, RISE TIME 500ms, 30ms/230VAC HOLD UP TIME (Typ.) 80ms/230VAC VOLTAGE RANGE 90 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz 85% 86% 86% 41.4 ~ 48.6V 55.2 ~ 64.8V 82% 73% 78% AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 1.5A/115VAC 1A/230VAC COLD START 25A/115VAC g 83% 50A/230VAC g Note.7 Earth leakage current < 150 A/264VAC , Touch current < 100 A/264VAC OVERLOAD PROTECTION OVER VOLTAGE WORKING TEMP. WORKING HUMIDITY ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT VIBRATION SAFETY STANDARDS SAFETY & WITHSTAND VOLTAGE ISOLATION RESISTANCE EMC (Note 4) EMC EMISSION 140 ~ 180% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed. 3.8 ~ 4.46V 5.75 ~ 6.75V 13.8 ~ 16.2V 17.25 ~ 20.25V 27.6 ~ 32.4V Protection type : Shut down o/p voltage, re-power to recover -20 ~ +70J (Refer to "Derating Curve") 20 ~ 90% RH non-condensing -40 ~ +85J, 10 ~ 95% RH O0.03%/J (0 ~ 45J) 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25J/ 70% RH Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61000-6-2, EN61204-3, heavy industry level, EN61204-3 medical level, criteria A MTBF 446.8K hrs min. MIL-HDBK-217F (25J) 127*76.2*31mm (L*W*H) 0.26Kg; 63pcs/16.3Kg/1.35CUFT DIMENSION PACKING NOTE 20ms/115VAC at full load 127 ~370VDC EFFICIENCY(Typ.) LEAKAGE CURRENT OTHERS 500ms, 30ms/115VAC at full load 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25J of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1uf & 47uf parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on http://www.meanwell.com) 5. Length of set up time is measured at cold first start. Turning ON/OFF the power supply may lead to increase of the set up time. 6. Heat Sink HS1,HS2,HS3 can not be shorted. 7. Touch current was measured from primary input to DC output. 8. Suitable for BF application with appropriate system consideration. File Name:RPS-75-SPEC 2012-11-28 RPS-75 75W Single Output Medical Type Mechanical Specification series Unit:mm FAN 5.7 15cm 23.5CFM min. HS3 CN2 3 2 1 64.75 76.2 1 2 3 4 5 6 7 8 1 CN3 2 HS1 CN1 HS2 AC FUSE T2.5/250V AC Input Connector (CN1) : JST B3P-VH or equivalent Pin No. 1 2 3 Assignment AC/N No Pin AC/L Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent DC Output Connector (CN2) : JST B8P-VH or equivalent FS2 FS1 115.8 5.6 Pin No. 1,2,3,4 5,6,7,8 4 -r 3.8 127 Assignment +V -V Mating Housing Terminal JST VHR or equivalent JST SVH-21T-P1.1 or equivalent Remote Sense(CN3) : JST B2B-XH or equivalent 31 Pin No. 1 2 Assignment RS+ RS- Mating Housing Terminal JST XHP or equivalent JST SXH-001T-P0.6 or equivalent 3max. HS1,HS2,HS3 can not be shorted Block Diagram fosc : 65KHz +S RECTIFIERS & FILTER EMI FILTER I/P RECTIFIERS & FILTER POWER SWITCHING +V -V -S DETECTION CIRCUIT FG O.L.P. CONTROL O.V.P. Derating Curve Output Derating VS Input Voltage 100 133 23.5CFM Forced air Cooling 90 Others 100 80 Convection Cooling 70 3.3V,5V LOAD (%) LOAD (%) 80 50 20 -20 0 10 20 30 40 50 AMBIENT TEMPERATURE (J) 60 70 (HORIZONTAL) 60 50 40 90 95 100 115 120 140 160 180 200 220 240 264 INPUT VOLTAGE (VAC) 60Hz File Name:RPS-75-SPEC 2012-11-28