Model Encoding
Description
2.06" .07" x1 compact size
Medical safety approved (2 x MOPP) to according
ANSI/AAMI ES60601-1 and IEC/EN60601-1
Suitable for BF application with appropriate system consideration
No load power consumption<0.075W (0.09W for 3.3V)
Extremely low leakage current
Wide operating temp. range -35 ~ +85
EMI class B for class configuration
Protections:
Short circuit / Overload / Over voltage / Over temperature
No minimum load required
3 years warranty
Features
MPM-20 series
Output voltage
Rated wattage
Series name
File Name:MPM-20-SPEC 2019-12-13
Portable medical device
Mobile clinical workstation
Medical computer monitor
Medical examination instrument
Applications
-5
MPM - 20
MPM-20 is a 20W high density and small size (52 4 27 2 24mm) AC/DC module type medical power supply .* .*
series offered in pin type. It features the operation for 80~264VAC, a low no load power consumption less
than 0.075W , a high efficiency up to 87%, Class (no FG) double insulation, outstanding
(0.09W for 3.3V)
dissipation and high lifespan thanks to the interior potting, 5G anti-vibration, high EMC performance, 4KVAC
isolation, etc. The design observes IEC EN60601 1 and ANSI/AAMI ES60601 1version three with 2xMOPP /- -
relevant equipment.
level and ultra-low leakage current (<80 A). It is very suitable for BF (patient contact) type medical device or μ
EN 60601-1
IEC 60601-1
8
IEC60601-1EN60601-1ANSI/AAMI ES60601-1
++
20W High Reliable Green Medical Encapsulated Type
OPERATING ALTITUDE Note.8 5000 meters
SPECIFICATION
MODEL
DC VOLTAGE
RATED CURRENT
CURRENT RANGE Note.2
RATED POWER
OUTPUT
LINE REGULATION
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME (Typ.)
VOLTAGE RANGE Note.6
FREQUENCY RANGE
EFFICIENCY (Typ.)
INPUT
INRUSH CURRENT (Typ.)
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
WORKING HUMIDITY
WORKING TEMP.
OVER TEMPERATURE
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSIONOTHERS
NOTE
PACKING
OVERLOAD
OVER VOLTAGE
AC CURRENT (Typ.)
20 ~ 90% RH non-condensing
-35 ~ +85 (Refer to "Derating Curve")
-40 ~ +85 , 10 ~ 95% RH non-condensing
±℃ 0.03%/ (0 ~ 55 )
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
1210Khrs min. MIL-HDBK-217F (25 )
52.4*27.2*24mm (L*W*H) 2.06 *1.07 *0.94 inch
or ″″″
SAFETY STANDARDS
RIPPLE & NOISE (max.) Note.4
VOLTAGE TOLERANCE Note.5
ENVIRONMENT
SAFETY &
EMC
(Note.9)
PROTECTION
Protection type : Shut off o/p voltage, clamping by zener diode
Protection type : Shut down o/p voltage, recovers automatically after temperature goes down
I/P-O/P:4KVAC
I/P-O/P:100M Ohms / 500VDC / 25 / 70% RH
LEAKAGE CURRENT (max.) Note.7
ISOLATION LEVEL Primary-Secondary: 2xMOPP
EMC IMMUNITY
EMC EMISSION
SOLDERING TEMPERATURE 260 5 /10sec.max.℃±
1500ms, 30ms/230VAC 1500ms, 30ms/115VAC at full load
40ms/230VAC 10ms/115VAC at full load
80 ~ 264VAC
47 ~ 440Hz
0.75A/115VAC 0.5A/230VAC
COLD START 20A/115VAC 45A/230VAC
110% ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
MPM-20-3.3
Touch current <80 A/264VAC
μ
±0.5%
±0.5%
±2.0%
150mVp-p
3.3V
4.5A
0 ~ 4.5A
14.9W
±0.5%
±0.3%
±0.5%
±0.3%
±0.5%
±0.3%
±0.5%
±0.5%
±2.0% ±2.0% ±2.0% ±2.0%
150mVp-p 150mVp-p 180mVp-p 180mVp-p
12V 15V 24V
4A 1.8A 1.4A 0.9A
0 ~ 4A 0 ~ 1.8A 0 ~ 1.4A 0 ~ 0.9A
20W 21.6W 21W 21.6W
5V
85%81% 85.5% 87% 87%
13.8 ~ 16.2V 17.3 ~ 20.3V 27.6 ~ 32.4V
5.8 ~ 6.8V3.8 ~ 5V
MPM-20-5 MPM-20-12 MPM-20-15 MPM-20-24
4.95A 4.4A 1.98A 1.54A 0.99A
MPM-20 series
0.056Kg; 240pcs/14.4Kg/0.97CUFT
PEAK CURRENT
20W High Reliable Green Medical Encapsulated Type
PEAK LOAD(10sec.)
Note.3
16.3W 22W 23.8W 23.1W 23.8W
IEC60601-1, EN60601-1, , ULANSI/AAMI ES60601-1(3.1 version), CAN/CSA-C22 3 Edition approved; Design
rd
EAC TP TC 004
refer to EN60335-1
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. No minimum load required.
3. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power.
4. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.μμ
5. Tolerance : includes set up tolerance, line regulation and load regulation.
6. Derating may be needed under low input voltages. Please check the derating curve for more details.
7. Touch current was measured from primary input to DC output.
8. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
9. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still
meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
Parameter
Parameter
ESD
Standard
Standard
EN55011 (CISPR11)
EN55011 (CISPR11)
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class B
Class A
-----
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
Level 4, 15KV air ; Level 4, 8KV contact
Level 3, 1KV/Line-Line
EN60601-1-2
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
Level 3, 10V/m( 80MHz~2.7GHz )
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
Voltage dip, interruption
Table 9, 9~28V/m( 385MHz~5.78GHz )
File Name:MPM-20-SPEC 2019-12-13
LEAD TEMPERATURE 260 5 ,5s (max.)±
Block Diagram fosc : 100KHzfosc : 100KHz
MPM-20 series
20W High Reliable Green Medical Encapsulated Type
FILTER
I/P EMI
FILTER
RECTIFIERS
&
RECTIFIERS
CONTROL
PWM DETECTION
CIRCUIT
&
FILTER
+V
-V
SWITCHING
POWER
O.L.P.
Output Derating VS Input Voltage
INPUT VOLTAGE (VAC) 60Hz
80 10095 120115 160140 200180 240220 264
LOAD (%)
90
100
80
70
60
50
40
Derating Curve
AMBIENT TEMPERATURE ( )
LOAD (%)
3.3V,5V
12V~24V
50
100
30
70
(HORIZONTAL)
-30 -20 -10 0 10 20 30 40 50 60 70 80 85
-35
55
File Name:MPM-20-SPEC 2019-12-13
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
MPM-20 series
Mechanical Specification Case No.219A Unit:(mm)
P/N diameter:1.04
24
1.04
3.5 1mm±
45
52.4
BOTTOM VIEW
AC/N -V
+V
AC/L
3.4
3.2
3.2 8
20.8
27.2
SIDE VIEW
20W High Reliable Green Medical Encapsulated Type
File Name:MPM-20-SPEC 2019-12-13