GLD150 GOLD PERFORMANCE MEDICAL SWITCHERS 150 Watt Single Output Features: * Compact 4.5" x 7" x 1.7" size * Power factor corrected to IEC 1000-3-2 Class A * Less than 300 A leakage * EMI compliance to CISPR11, FCC Class B * Power fail and remote sense standard * Approved to UL2601, IEC601-1 and CSA22.2 No. 601.1-M90 and 60601: 1990 * 2 year warranty * marked to LVD SPECIFICATIONS Ac Input 85-264 Vac, 47-63 Hz single phase. Input Current 2.8 A line current maximum, at 90 Vac, 60 Hz with full rated load, power factor .99 typical, .96 minimum. Input current harmonic content meets the requirements of IEC1000-3-2. Output Power 150 W with convection cooling, 180 W with fan cooling. Efficiency Minimum 80% at full rated load with 230 Vac Input. Approximately 3% less at 115 Vac. Hold-Up Time Outputs will remain within regulation limits for 25 ms minimum from loss of ac input at full load, 10 ms before Power Fail indication. Dc Ouput Total regulation is the maximum deviation from the nominal voltage for all steady state loading conditions. Peak ratings are for 60 s maximum duration, 10% duty cycle. Overload Protection Fully protected against short circuit and output overload. Short circuit protection is cycling type power limit. Minimum Load No minimum load required to maintain output specifications. Output Noise 0.5% rms, 1% pk-pk, 20 MHz Bandwidth, differential mode. Measured with noise probe directly across output terminals of the power supply. Transient Response Main Output - 500 s typical response time for return to within 0.5% of final value for a 50% load step change, i/t< 0.2 A s. Maximum voltage deviation is 3%. Remote Sense Provided as a standard feature on all models. Overvoltage Protection Built in on all models. Voltage Adjustment Output Voltage is adjustable +/- 5% with user adjustable potentiometer. Temperature Coefficient 0.03% / C typical on all outputs. Turn On Time Less than 1 second. Overshoot Less than 2% overshoot at turn-on under all conditions, less than 1% overshoot at turn-off under all conditions. Internal ac fuses provided on both line and neutal on all models. Inhibit Inhibit signal is pulled to the V1 output common to reduce average output voltage to less than 5% of nominal. 78 EMI/EMC Compliance All models include built-in EMI filtering to meet the EMC requirements of IEC601-1-2. Unless otherwise stated, all tests are done at full load and 115 and 230 Vac input. Conducted Emissions Static Discharge RF Field Susceptibility Fast Transients / Bursts Surge Susceptibility Conducted RF Susceptibility Voltage Sags & Surges EN55011, Class B; FCC Class B EN61000-4-2, Level 3 EN61000-4-3, Level 3 EN61000-4-4, Level 3 EN61000-4-5, Level 3 EN61000-4-6 EN61000-4-115% Inrush Current Inrush 240 Vac is less than 37 A, averaged over the first ac halfcycle under cold start conditions. Limiting provided by internal thermistors. Fan Output An additional 12 Vdc, 250 mA output suitable for powering a dc fan is included in all models. The fan output is both current limited and thermally protected. Thermal Shutdown Provided as a standard feature. Designed to protect unit from prolonged over temperature. Power Fail TTL / CMOS compatible output goes low (<0.5 V) 8 ms before output voltage drops more than 4% below nominal voltage upon loss of ac power. Power Good TTL / CMOS compatible output goes high more than 100 ms after V1 reaches regulation and should assure that sufficient energy is stored in the input section to provide normal power fail/shutdown. Medical Approvals All models are Certified to be in compliance with the applicable requirements of UL2601, CSA 22.2 No. 601.1-M90, IEC 601-1 (1988), EN 60601-1: 1990. Leakage Current 70 A, 132 Vac @ 60 Hz normal conditions. Single fault conditions, 130 A, 254 Vac @ 50 Hz. Design Verification Documents The "Gold" series has undergone rigorous review and design analysis. The following product documentation is available upon request; 1. MTBF study 2. DVT Data 3. EMC / Susceptibility test results Condor D.C. Power Supplies, Inc./A subsidiary of SL Industries, Inc. 2311 Statham Parkway, Oxnard, CA 93033 * 800-235-5929 * 805-486-4565 * FAX 805-487-8911 * www.condorpower.com GLD150 Medical Switchers 150 Watt Multiple Output Med ical Mod el Output Vol t ag e Output Cu rren t (A) Output Cu r r en t ( B ) Vol t ag e Ad j u st men t Tot al Reg u l at ion O VP S et p oin t GLD150-12 12 V 12.5 A 15 A 5% 2% 14 + 1.1 V GLD150-15 15 V 10 A 12 A 5% 2% 18.5 + 1.5 V 1% GLD150-24 24 V 6.2 A 7.5 A 5% 2% 28 + 2.5 V 1% GLD150-28 28 V 5.3 A 6.4 A 5% 2% 34 + 2.8 V 1% GLD150-48 48 V 3.2 A 3.75 A 5% 2% 55 + 4.0 V 1% Rip p l e/ Nois e 1% Notes: A. Maximum continuous current rating for unrestricted convection cooling. B. Maximum continuous current rating with 150 LFM air or peak rating. C. Add "C" suffix for cover option and derate convection rating to 130 W. GLD150 MECHANICAL SPECIFICATIONS INPUT J1 MOLEX P.C.B. HEADER P/N: 39-30-2056 PIN 1) AC GROUND PIN 2) N/C PIN 3) AC NEUTRAL PIN 4) N/C PIN 5) AC LINE MATING CONNECTOR MOLEX P/N HOUSING 39-01-4051 CONTACT 39-00-0164 SIGNALS J2 AMP P.C.B. HEADER P/N 641215-6 PIN 1) INHIBIT PIN 2) +SENSE PIN 3) POWER GOOD PIN 4) -SENSE PIN 5) COMMON PIN 6) POWER FAIL MATING CONNECTOR AMP P/N HOUSING 770602-6 CONTACT 770666-6 OUTPUT J3 MOLEX P.C.B. HEADER P/N: 39-29-9085 PINS 3,4,7,8) +Vout PINS 1,2,5,6) RETURN MATING CONNECTOR MOLEX P/N HOUSING 39-01-2080 CONTACT 39-00-0163 FAN AMP P.C.B. HEADER P/N: 641215-2 PINS 1) RTN PINS 2) +12V MATING CONNECTOR AMP P/N HOUSING 770602-2 CONTACT 770666-02 Environmental Specification Temperature (A) Humidity (A) Shock (B) Altitude Vibration (C) Operating WEIGHT: 1.9 LBS [.86 kg] MAX. 1.278 [32.46] 6-32 MTG 2 PLCS EA. SIDE .250 [6.35] 1.70 [43.18] 1.95 [49.53] OPTIONAL COVER 6.500 [165.10] J3 6-32 MTG 4 PLCS OUTPUT 1.360 [34.54] FAN OUTPUT 1.944 [49.38] 1.278 [32.46] 4.50 [114.30] OUTPUT V ADJ. 1.330 [33.78] 1.170 [29.72] J2 SIGNALS 7.00 [177.80] J1 ~ INPUT Non-operating 0 to 50C -40 to +85C 0 to 95% RH 0 to 95% RH 20 gpk 40 gpk -500 to 10,000 ft -500 to 40,000 ft 1.5 grms, 0.003 g2/Hz 5 grms, 0.026 g2/Hz A. Units should be allowed to warm up/operate under non-condensing conditions before application of power. B. Random vibration--10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating. C. Shock testing--half-sinusoidal, 10 3 ms duration, direction, 3 orthogonal axes, total 6 shocks. 4/17/01 Condor D.C. Power Supplies, Inc./A subsidiary of SL Industries, Inc. 2311 Statham Parkway, Oxnard, CA 93033 * 800-235-5929 * 805-486-4565 * FAX 805-487-8911 * www.condorpower.com 79