2Honeywell•SensingandControl
Model 7541
GENERAL SPECIFICATIONS
Characteristic Measure
Straingageinput Any80ohmto2Kohmtransducer,di-
rectlywiredortransformercoupled.4,6,
or 7 wire circuits are accommodated
Transducer excitation,
ac models
3Vrms,3.28kHz±0.01%sinewave.
Regulatedandshortcircuitprotected
Transducer excitation,
dc models
5 V or 10 V dc, user selectable, regu-
lated
Sensitivity 1.0mV/Vto4.5mV/Vwith50%over-
range;automaticallyscaled
Input resistance 100mohminparallelwith33pF
Automaticnull InPhase:±10%offullscale(with50%
overrange),±60%offullscale(with0%
overrange).Quadrature:±1mV/V
Autocalibration Dual polarity shunt calibration with provi-
sionforCALresistorfeedback
Spurioussignalrejection 60Hz:120dBcommonmode,100dB
normalmode.Carrierquadrature:60dB
Anti-aliaslter 200Hz,7-poleBesselresponselter
Low-passltering Four-poleBesselresponsedigitallter
with11cutofffrequenciesfrom0.1Hzto
200Hzin1-2-5steps
ENVIRONMENTAL SPECIFICATIONS
Characteristic Measure
Temperatureeffects,zero ±0.001%offullscale/°Fmax.
Temperature effect, span ±0.001%offullscale/°Fmax.
Frequencyinput Anyunidirectionalorbidirectional
(quadrature) source including self-
generatingandzerovelocitymagnetic
pickups, optical encoders, flowmeters,
etc. When used with bidirectional sen-
sors, the conditioner outputs both direc-
tion and magnitude
Input impedance and
conguration
Differential or single ended inputs. 100
kohm differential, 50 kohm single ended
Input threshold
(keypad selectable)
10, 20, 50, 100 or 200 mV pk-pk (be-
tweeninputs)orTTL
Maximumvoltage ±130Vdcor130Vrms
Input signal bandwidth 0.001kHzto200kHz(10mVto200mV
pk-pkthreshold)0.001kHzto400kHz
(TTLthreshold)
Display ranges and
resolution
Rangeless(useanyfullscaleengineer-
ing unit value) with 50 % overrange.
Resolutionis0.01%offullscale
Lowpasslter
(keypad selectable)
20kHz(-3dB)ornone(thislterisnot
availableforTTLinputs)
Responsetime Greaterof:1ms,typ(2msworstcase)
or the input pulse length
Commonmoderejection 80dB(60Hz),55dB(0kHzto10kHz)
Lowpasslteringof
sampled data
Unlteredorfour-poleBessellter.Cut-
offfrequenciesfrom0.1Hzto100Hzin
1-2-5 steps
Overall accuracy 0.01%offullscale@25°C[77°F],
0.015%offullscale@5°Cto50°C[41
°Fto122°F]
Excitation supplies 12V@125mA2or5V@250mA2.
Shortcircuit(currentlimitandovervolt-
age (fuses) protected
Maximumtransducer
cable length
500 ft, except 200 ft for 100 ohm or
lower strain gage transducers
ELECTRICAL SPECIFICATIONS
Characteristic Measure
Signal-to-noiseratio1With1/10/100/200Hzlters86/76/66/62
dB@1mV/Vfullscaleand86/80/72/66
dB@5mV/VFS
Resolution 0.01 % of full scale
Overall accuracy @
25°C[77ºF]
0.02 % of full scale, worst case.
DISPLAY SPECIFICATIONS
Characteristic Measure
Systemdisplay Twolineby16alphanumericcharac-
ters,each0.2inwideby0.3inhigh.
BacklitLCDwithadjustablecontrast
Views Selecteithertwochannels,oneChannel
with limit status or one channel with I/O
status
Data displayed Selectfromcurrent,max.,min.,spread,
held data and tare value
Data format Engineering units with six digits (1-2-5
format)andvecharacter,upperor
lower case, user-entered legend/de-
scriptor
SYSTEM RESPONSE (PER CHANNEL)
SPECIFICATIONS
Characteristic Measure
Data samp. & max./min.
update rates
2000Hz(hardwarechannels),50Hz
(CH3calculation)
Limitcheckingrate 1000Hz(hardwarechannels),50Hz
(CH3calculation)
LogicI/Oresponsetime 1ms(hardwarechannels),20ms(CH3
calculation)
Update rate for each
analog output
1000Hz