AULT INCORPORATED
7105 Northland Terrace |Minneapolis, Minnesota 55428 |USATEL (1) 763-592-1900
FAX (1) 763-592-1911 |E-MAIL info@aultinc.com |WEB www.aultinc.com
© Copyright 2002 Ault Inc.
Specify the Input Configuration Code in your order.
Connector
Input Configuration
MW125
MEDICAL 15-40 WATT SERIES
N. America/
Pinout Code Center contact: positive
Description Switchcraft 760 plug or equivalent
Suggested Mating Switchcraft 712A jack or equivalent
Other Connectors are available by special order
Input
Voltage 100-240 VAC -10%, +10%
Line Frequency 47-63 Hz
Current 1.0 Amp max. at 90 VAC input
Protection Internal primary current fuse, Inrush
limiting,Thermal shutdown
Output
Combined Line and Load ±1% excluding cord
Voltage Regulation
Ripple See table
Transient Response 0.5msec. for 50% load change typical
Holdup Time 18 msec min. @ 120 VAC
80 msec min. @ 240 VAC
Protection Overcurrent protection,
Short circuit Protection
Cord and Connector 6 ft, 2 conductor, 18 AWG
#3 connector: Other connectors also
available.
Safety Approvals
Agency Listings cURus2601, TUV/IEC EN60601-1,
Australian, Japanese certification
available - extra fees apply
Electrical
Topology Switching-fixed frequency flyback
Efficiency Up to 85%
EMI Complies with EMC Directives
CE CE Compliant
MTBF 100,000 Calculated Hours Minimum
Environmental
Operating Temperature 0º to 40º C with no derating
Storage Temperature -30º to +85º C
Relative Humidity 5% to 95% non-condensing
Altitude 0-10,000 feet
Cooling Convectional-non vented case
Mechanical
Case and Dimensions All Models
3.92L x 2.58W x 1.61H (in)
100L x 66W x 41H (mm)
Case Material 94V0 polycarbonate; black is standard
color; other colors available - extra
charges may apply
Weight Max. 10 ounces, 285 grams
(excluding cord)