Hans Turck GmbH & Co.KG • Postfach • D-45466 Mülheim an der Ruhr • Telefon 02 08/49 52-0 • Telefax 02 08/49 52-264
Subject to change without notice • © Hans Turck GmbH & Co.KG 2002 • Edition: • 13.04.2004
Subject to change without notice • © Hans Turck GmbH & Co.KG 2002 • Edition:
Hans Turck GmbH & Co.KG • Postfach • D-45466 Mülheim an der Ruhr • Telefon 02 08/49 52-0 • Telefax 02 08/49 52-264
non-invasive filling level control sensor
levelprox
LPRE-T50-UP6X3-H1151
detection of liquids through the wall
of a container
echo or reverberation mode adju-
stable via DIP switches
suited for a large variety of contai-
ner materials such as metal and
glass (plastic and other materials on
request)
non-invasive, without medium con-
tact, thus especially hygienic
simple mounting with welding or ad-
hesive Tri-Clamp adapter
alarm output and alarm indication in
case of errors or insufficient cou-
pling
automatic teach-in of ambient con-
ditions via push button or teach line
Wiring diagram
Function principles
Non-invasive ultrasonic sensors detect li-
quids through the metal wall of a container
or tank by means of high frequency ultraso-
nic pulses. In the reverberation mode, the
reflection of the interior container wall adja-
cent to the sensor are evaluated. Contami-
nation of or air bubbles in the medium
hardly effect operation, whereas residues
on the container inside will lead to unwan-
ted switching. In the echo mode, the sound
is evaluated that it reflected by the oppo-
sing interior wall. Residues on the container
inside will not cause any problems; but if
the container contains built-in components
such as stirring devices, or if the medium is
contaminated or contains air bubbles, this
operation mode may not be used. Electri-
cal or acoustic interference or a sensor
failure due to insufficient coupling are indi-
cated by an LED and an alarm output.
+
3 BU
4 BK
2 WH
1 BN
5 GR
Alarm
Teach
Type LPRE-T50-UP6X3-H1151
Ident-No. 1606201
Container 1... 15 mm wall thickness
Material Steel, Stainless steel, Glass
Min. container diameter 100 mm
Max. container diameter 2000 mm (echo mode)
no limit (reverberation mode)
Adjustment to ambient conditions via push button or teach line
Min. absolute repeat accuracy ± 5 mm
Switching delay 1...10 sec (adjustable via DIP-switches)
Rated operational voltage (DC) UB15... 30 VDC
Function of switching element programmable
Alarm output PNP
No-load current I0 60 mA
Rated operational current (DC) Ie 200 mA
Overload trip point >270 mA
Max. voltage drop at Ie 2,5 V
Switching frequency 1 Hz
Time delay before availability 2 s
Short-circuit protection yes
Reverse polarity protection complete
Wire breakage protection yes
Housing style cylindrical/smooth; T50 (Tri-Clamp Gehäuse)
Housing material metal, A4 1.4404 (AISI 316L)
Wiring connector, M12 x 1
Operating temperature -25 ...+ 70 °C
Degree of protection IP67
(cover screw connections tightened)
Protection class III, SELV
Supply voltage indication LED green
Switching status indication LED yellow
Error indication LED red flashing and alarm output on
Alarm indication LED red on and alarm output on
ø 64
12
29
44
67
M12 x 1
ø 25
ø 50
ø 72
67
Hans Turck GmbH & Co.KG • Postfach D-45466 Mülheim an der Ruhr • Telefon 02 08/49 52-0 • Telefax 02 08/49 52-264
Subject to change without notice • © Hans Turck GmbH & Co.KG 2002 • Edition: • 13.04.2004
Accessories
Subject to change without notice • © Hans Turck GmbH & Co.KG 2002 • Edition:
Hans Turck GmbH & Co.KG • Postfach • D-45466 Mülheim an der Ruhr • Telefon 02 08/49 52-0 • Telefax 02 08/49 52-264
non-invasive filling level control sensor
levelprox
LPRE-T50-UP6X3-H1151
Typ Ident-
No.
Description Dimension drawing
LP-MS-T50-S 6900253 Tri-Clamp adapter for weld mounting. seal and clam-
ping collar are included in delivery.
LP-MS-T50-K 6900255 Tri-Clamp adapter for adhesive mounting. Seal, clam-
ping collar and two-component adhesive are inclu-
ded in delivery.
21,5
ø 29ø 50,4
ø 26
ø 58
21,5
ø 29 ø 50,4
ø 26