MWA120 Series - ICCNexergy Page 1 of 4 MWA120 Series 120 Medical Desktop Power Supply * * * * * * * * 5-Year Limited Warranty High Efficiency: Level V High Power Density Lifetime Expectation >5 years Hold-up Time > 25ms at full load EISA, CEC Compliant Floating output Medical Approval- EN60601-1 Class I Elpac Part Number Output Voltage Output Current Peak Current Total Regulation Typical Efficiency MWA120012A-12A 12.0V 10.0A 12.0A 5% 89% MWA120015A-12A 15.0A 8.0A 10.0A 5% 90% MWA120018A-12A 18.0A 6.7A 8.0A 5% 90% MWA120024A-12A 24.0V 5.0A 6.0A 5% 91% MWA120048A-11A 48.0V 2.5A 3.0A 5% 92% Input Input Voltage 85 - 264VAC; 100 - 240VAC Nominal Input Frequency 47 - 63Hz Input Current <2A rms Inrush Current <37A at 230VAC cold start Power Factor >0.97 Zero Load Power Consumption <0.5W Output Output Voltage See Table Total Regulation +/-5% Minimum Load No minimum load required Start-Up Delay <1.5s Hold-Up Time >25ms at any input voltage http://www.iccnexergy.com/elpac/medical-power-supplies/external/mwa120-series/?ms=0 5/31/2013 MWA120 Series - ICCNexergy Page 2 of 4 Ripple & Noise <1% pk-pk * Over Voltage Protection 110-135% Over Temperature Protection Active - Recoverable; plus Passive - Non Recoverable Over Current Protetion 120 - 180% Short Circuit Protection Shutdown, auto-restart (hiccup mode) General Efficiency Avg Efficiency 90% @ 115VAC; 92% @ 230VAC MTBF min. 200,000 hours demonstrated Size 7.56" (192mm) x 2.45" (62.2mm) x 1.52" (38.7mm) Weight 1.55 lbs (0.70 kg) Environmental Operating Temperature 0 - 60C (Full load to 40C, derate linearly to 50% load at 60C) Storage Temperature -40C to +85C Relative Humidity 5-95%, non-condensing Cooling Natural Convection Vibration All units production tested to 19.6m/s2 EMC & Safety Emissions FCC class B, CISPR11 class B EN61000-3-2, -3 Immunity EN61000-4-2, -3, -4, -5, -6, -8, -11 Certified by TUV to the following: cTUVus UL 60601-1 CAN/CSA-22.2 No.601.1-M90 CB per IEC60601-1 CE marked to LVD Input Configuration Standard Input Cable Not Provided Connection on Power Supply Body IEC 320 C14 Receptacle http://www.iccnexergy.com/elpac/medical-power-supplies/external/mwa120-series/?ms=0 5/31/2013 MWA120 Series - ICCNexergy Page 3 of 4 Output Configuration (12V, 15V) Standard Output Cable 4 ft. Cord Size 2x16awg Connector (PSU side) Switchcraft DIN-8, P/N 15BL8M (male pins) Mating Connector Switchcraft 62GB8FX (8 pin) or equivalent Output Configuration (18V, 24V) Standard Output Cable 6 ft. Cord Size 2x16awg Connector (PSU side) Switchcraft DIN-8, P/N 15BL8M (male pins) Mating Connector Switchcraft 62GB8FX (8 pin) or equivalent Output Configuration (48V) Standard Output Cable 6 ft. Cord Size 2x18awg Connector (PSU side) Switchcraft DIN-5, P/N 05GM5M (male pins) Mating Connector Switchcraft 57GB5FX (5 pin) or equivalent Output Pin Assignments Pin 1 Return Pin 2 Return Pin 3 +V1 Pin 4 Return Pin 5 +V1 Output Pin Assignments http://www.iccnexergy.com/elpac/medical-power-supplies/external/mwa120-series/?ms=0 5/31/2013 MWA120 Series - ICCNexergy Page 4 of 4 Pin 1 +V1 Pin 2 +V1 Pin 3 Return Pin 4 +V1 Pin 5 Return Pin 6 +V1 Pin 7 Return Pin 8 Return MWA120 SERIES | ICCNEXERGY | (c) 2012 | http://www.iccnexergy.com/elpac/medical-power-supplies/external/mwa120-series/?ms=0 5/31/2013