DTM65-C8 Series
Features
Meets DOE Level VI Effi ciency
60601-1 & 60601-1-11 Medical Certifi cations
Class II Input
< 0.21W Off -load Power Draw
40W to 65W Medical Class II AC-DC External Power Supplies
Model
Input Voltage range VAC 90 - 264VAC (47 - 63Hz)
Inrush Current A <100A at 230VAC input, 25°C ambient cold start
Input Current A 2A at 90VAC, Full Load
Power Factor - Meets EN61000-3-2 Class A
Hold Up Time (Typ) ms 10ms at 115VAC input
Temperature Coeffi cient - ±0.05%/°C
Voltage Accuracy % ±1%
Adjustment Range - None
Minimum Load A None
Total Regulation - 5V to 15V Models:±5%, >15V models ±3%
Ripple & Noise % 1% pk-pk
Overload & Short Circuit Protection - Continuous - hiccup mode
Overvoltage Protection V 110 - 150% of nominal (Cycle input power to reset)
Overtemperature Protection - Yes (Cycle input power to reset)
Effi ciency % >88%, average effi ciency (5V model: >83.8%)
Operating Temperature - 0 to +60°C derate linearly to 50% load from 41 to 60°C
Storage Temperature °C -10 to +70°C
Humidity (non condensing) %RH 10 - 90% RH
Cooling - Convection
Withstand Voltage VAC Input to Ground 1.5kVAC, Input to Output 4kVAC
Vibration (non operating) - 23.52m/s2 (10 - 55Hz: constant sweep 1 min X, Y, Z for 1 hour)
Shock - < 196.1 m/s2 (20G)
Safety Agency Certifi cations - ANSI/AAMI ES60601-1, EN/IEC60601-1, CSA-C22.2 No.60601 3rd edition (2 x MOPP)
Offl oad Power Consumption W < 0.21W
EMC - EN55011B, EN60601-1-2
IP Rating - IP22
Weight (Typ) g 240g
Size (WxLxH) in (mm) 2.36 x 4.17 x 1.22” (60 x 106 x 31mm)
Cable Length mm 1050mm
AC Input Connector - IEC 320-C8 (2 prong)
Output Connector - Kycon KPPX-4P or equivalent
Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Warranty Years Three Years
Specifi cations
Key Market Segments & Applications