MIT400 Series
Industrial Insulation Testers
SPECIFICATION
All quoted accuracies are at +20°C.
Insulation
Nominal test voltages
MIT400 250 V, 500 V, 1000 V
MIT410, 420,430 50 V, 100 V, 250 V, 500 V, 1000 V
MIT40X 10 V to 100 V variable (1 V increments)
Insulation resistance range
MIT400 20 GΩ
MIT410 100 GΩ
MIT420, 430 200 GΩ
MIT40X 200 GΩ
Range Full Scale Accuracy
All ranges ±2% ±2 digits up to 100 MΩ.
Then:
1000 volts ±3% ±2 digits ±0.2% per GΩ
500 volts. ±3% ±2 digits ±0.4% per GΩ
250 volts. ±3% ±2 digits ±0.8% per GΩ
100 volts. ±3% ±2 digits ±2.0% per GΩ
50 volts. ±3% ±2 digits ±4.0% per GΩ
10 volts ±3% ±2 digits ±2.0% per 100 MΩ
Analogue range: 1 GΩfull scale
Short Circuit Current: 1 mA ±0.2 mA
Terminal voltage: -0% +20% ±1 V
Test Current on load:
1 mA at min. pass value of insulation specified in BS7671, HD384
and IEC364, 2 mA max.
EN61557 Operating range: 0,10 MΩto 1,00 GΩ
Leakage current range 10 µA 2000 µA
Leakage current: 10% ±3 digits
Voltage display: 3% ±3 digits ±0.5% of rated
voltage
Polarisation Index (PI): 10 min / 1minute ratio
Dielectric Absorption Ratio (DAR): 60 sec / 30 sec ratio
Notes:
(1) All ranges measure from 0,00 MΩupwards.
(2) Above specifications only apply when high quality silicone leads
are being used.
Continuity
Measurement: 0,01 Ωto 99,9 Ω(0 to 100 Ωon
analogue scale)
Accuracy: ±2% ±2 digits (0 to 100 Ω)
Open circuit voltage: 5 V ±1 V
Test current: 205 mA (±5 mA)
(0.01 Ωto 9.99 Ω)
20mA (±1 mA )
(10.0 Ωto 99.9 Ω)
Zero offset at probe tips: 0,10 Ωtypical
Lead resistance zeroing: Up to 9.99 Ω
Buzzer: Variable limit 1 Ω, 2 Ω, 5 Ω,
10 Ω, 20 Ω
Resistance
Measurement: 0.01 kΩto 1000 kΩ(0 to 1 MΩ
on analogue scale)
Accuracy: ±3% ±2 digits up to 50 kΩ
then ±5% ±2 digits
Open circuit voltage: 5 V ±1 V
Short circuit current: 1.5 mA ±0.2 mA
Voltage range
0 to 600 V d.c. ±2% ±2 digits
10 mV to 600 V TRMS sinusoidal (40 to 400 Hz) ±2% ±2 digits
0 to 1000 V on Analogue scale
Unspecified input level 0 - 10 mV (40 to 400 Hz)
For non-sinusoidal waveforms additional specification
apply:
±3% ±2 digits 101 mV to 600 V TRMS and ±8% ±2 digits 10 mV to
100 mV TRMS
Default Voltmeter: Operates at >25 V a.c. or d.c.
on any range except OFF
Frequency: 40-450 Hz (40 Hz - 99.9 Hz)
±0.5% ±1 digit (100 Hz to
450 Hz)
Capacitance measurement
MIT420, MIT430
Measurement range: 100 pF to 10 µF
Accuracy: ± 5.0% ±2 digits
Distance by capacitance:
MIT420, MIT430
Arithmetic conversion from capacitance measurement on
Default capacitance measurement: 50nF/km
Capacitance range: 40 nF/km to 60 nF/km
Result storage
Capacity: >1000 test results
Download: Bluetooth wireless
Bluetooth Class: Class II
Range: up to 10 m
Power Supply:
5 x 1,5 V cells type IEC LR6 (AA, MN1500, HP7, AM3 R6HP) Alkaline
NiMH rechargeable cells may be used.
Battery life: 2200 insulation tests with duty cycle of 5 sec ON /55
sec OFF @ 1000 V into 1 MΩ
Dimensions
Instrument: 220 x 92 x 50 mm (8.66 in. x 3.63 in. x 1.97 in.)
Instrument + case: 456 x 178 x 89 mm (18 in. x 7 in. x 3.5 in.)
Weight
Instrument only: 590 gms, 775 gms with boot (20.73 oz.
27.22 oz.)
Instrument plus case: 1.75kg (3.86 lb)