DTM110-C8 Series
Features
Meets DoE Level VI & EU Tier 2 Effi ciency
Medical & ITE Certifi cations
Class II Input (No Earth Ground Required)
< 0.15W Off -load Power Draw
Meets IEC60601-1-2 Ed4 & 60601-1-11
Suitable for B & BF Rated Equipment
110W Medical/ITE Class II External Power Supplies
Model DTM110-C8
Input Voltage range VAC 90 - 264VAC (47 - 63Hz)
Inrush Current A <120A at 230VAC input, 25°C ambient cold start
Input Current A 2A @90VAC; 1.25A @264VAC Maximum
Touch Current - <100uA @ 264VAC
Power Factor - Meets EN61000-3-2, >0.9 at 230VAC full load
Hold Up Time (Typ) ms 10ms at 115VAC input
Temperature Coeffi cient %/°C ±0.05%/°C
Adjustment Range - None
Minimum Load A None
Total Regulation % 12-19V: ±5%, 24-48V: ±3%
Ripple & Noise (pk-pk) mV 150mV
Overload & Short Circuit Protection - 105 - 150% - hiccup mode
Overvoltage Protection V Cycle input power to reset
Effi ciency % >89%, average effi ciency
Operating Temperature °C -20 to +60°C derate linearly to 50% load from 40 to 60°C
Derate linearly to 80% load from 0 to -20°C
Storage Temperature °C -20 to +85°C
Humidity (non condensing) %RH 0 - 95%RH
Cooling - Convection
Withstand Voltage VAC Input to Output 4kVAC (2xMoPP)
Safety Agency Certifi cation - IEC/ES/CSA/EN60601-1, IEC60601-1-11, IEC/UL/CSA/EN62368-1, 60950-1, CE Mark
Offl oad Power Consumption W < 0.15W
EMC - IEC60601-1-2 Ed4:2014, EN55011 Class B, EN55032 Class B,
FCC Part 15 Class B, FCC Part 18 Class B
IP Rating - IP41
Altitude m 5,000m maximum
Weight (Typ) g 780g
Size (LxWxH) mm 140 x 63 x 33mm
Size (LxWxH) in 5.5 x 2.48 x 1.29”
Cable Length mm 1000mm
AC Input Connector - IEC 320-C8 (2 prong)
Output Connector - Kycon KPPX-4P or equivalent; Pins 1 & 2: -Vout, Pins 3 & 4: +Vout
Warranty yrs Three Years
Specifi cations
Key Market Segments & Applications
Notes:
Other connectors are available. Contact factory.