COOLX®600
FANLESS, NATURAL CONVECTION-COOLED MODULAR POWER SUPPLY
Advanced Energy's CoolX®600 series, part of our low voltage
solutions, is the world's first fanless, natural convection-cooled
modular power supply. The CoolX600 delivers an incredible 600 W
without fan-assisted cooling from a very compact package. The CoolX600
offers system designers best-in-class efficiency and reliability in addition
to the most comprehensive feature set and specifications available. CX06S CX06M
Power
600 W 600 W
Slots
4 4
Cooling
No fan featured,
convection-cooled
Parameters
215.9 mm x 114.3 mm x 39.1 mm
(8.5 in x 4.5 in x 1 U)
Certifications
Medical
IEC60601-1 3rd edition,
IEC60601-1-2 4th edition (EMC)
2 MOPP
Dual fused
Suitable for type B- and
BF-rated applications
ISO13485
Industrial
IEC60950, IEC62368-1
ISO9000
SEMI F47
Defense/Aero
MIL-STD-810G
AT A GLANCE
PRODUCT HIGHLIGHTS
No Fan Featured
600 W with 100% natural convection
cooling
No base plate needed
No acoustic noise or vibrations
Reliability
MTBF > 400,000 hours, 25% better than
today’s leading solutions
High input surge protection — 4 KV line
to PE for harsh environments
Reverse energy protection — no
blocking diodes required
24 W always ON auxiliary power output
Safety approved to 5000 m altitude
> 94% efficiency
Five-year warranty
Flexibility
Analog and digital management
— PMBus™ monitoring and control
capability
Field-configurable — plug and play
power
Series and parallel outputs — higher
voltages/currents
Mounting options — base/side and
DIN-Rail mounting
TYPICAL APPLICATIONS
Medical
Clinical diagnostic equipment, medical
lasers, dialysis equipment, radiological
imaging, clinical chemistry
Industrial
Test and measurement, industrial
machines, automation equipment, printing,
telecommunications, audio equipment
Hi Rel
Harsh industrial electronics, radar (naval-
and ground-based), communications,
test and measurement
2 advancedenergy.com
COOLX600 SERIES
MODULES
CoolX CoolMods Table
CoolMod Vnom(V) Set Point Adjust Range (V) Imax(A) Power (W)
CmA15 2.5-6.0 21.0 105
CmB112 6.0-15.0 15.0 180
CmC124 15.0-28.0 8.3 200
CmD148 28.0-58.0 4.2 200
High Power Modules (3 Slot)
CmE 24 22.8-25.2 25.0 600
CmF 48 45.6-50.4 12.5 600
Dual Output Modules (1 Slot)
CmG2 V1 24 3.0-30.0 3.0 90
V2 24 3.0-30.0 3.0 90
CmH3 V1 5 3.0-6.0 6.0 36
V2 24 3.0-30.0 3.0 90
Wide Trim Modules (1 Slot)
CmM 5 1.0-6.0 21 105
CmN 12 1.0-15.0 15 180
CmP 24 1.0-28.0 8.33 200
CmQ 48 3.0-58.0 4.17 200
1 Full dynamic specifications may not be met at full load when output voltage is trimmed above 13 V.
2 For the CmG module the max combined power of both outputs is 120 W.
3 For the CmH module the max combined power of both outputs is 100 W.
ELECTRICAL SPECFICATIONS
Input
Parameter Conditions/Description Min Nom 264 Units
AC Operating Input Range 85 264 VAC
Nominal Input Voltage Range Universal Input 47-440 Hz 100 240 VAC
Extended AC Operating Range Maximum for 5 seconds 300 VAC
DC Input Voltage Range 120 300 VDC
Input Current 90 VAC @ 420 W 6 A
Inrush Current 230 VAC @ 600 W 25 A
Power Factor 120 VAC @ 500 W 0.98
Undervoltage Lockout Shutdown 65 74 VAC
Input Fuses Rating Dual Fused (Line and Neutral) 250 VAC 8 A
Efficiency 230 VAC, 600 W with 3 x CmC CoolMods 93 %
230 VAC, 600 W with 1 x CmE CoolMod 94 %
Output
Power Rating CX06: See derating curves 600 W
Minimum Load 0 A
Line Regulation For ±10% change from nominal line ±0.1
CmF, CmG, CmH ±0.5 %
Load and Cross Regulation For 25% to 75% load change ±0.2 %
advancedenergy.com 3
COOLX600 SERIES
Output (Continued)
Parameter Conditions/Description Min Nom 264 Units
Transient Response For 25% to 75% load change 0.5 A/uS: voltage deviation 4 (4) %
*CmE and CmF, CmG, CmH Figures in () Settling Time 500 (1000) µS
Ripple and Noise 100 mV or 1.0% pk-pk. 20 MHz BW 1 %
CmF 1.5 %
Overvoltage Protection Tracking OVP Level (N/A in CmE and CmF, CmG, CmH) 105 125 %
Latching OVP Level 110 160 %
Remote Sense Max line drop compensation (N/A in CmG and CmH) 0.5 VDC
Rise Time Monotonic CmG, CmH Figures in () 10 (20) ms
Turn-On Delay From AC in *CmE and CmF, CmG, CmH Figures in () 800 ms
From Global Enable 10 (20) ms
From CoolMod Enable 10 (20) ms
Hold-Up Time For nominal output voltages at full load 20 ms
CoolMod Power As per CoolMod table ————
Output Adjustment Range Manual: Multi-turn potentiometer, as per CoolMod table
Vtrim: As per CoolMod table
————
Overcurrent Protection Straight line with hiccup activation @ 35% Vo nom
CmE and CmF, CmG, CmH: Current limit hiccup autorecovery
CmM, CmN, CmP and CmQ: Straight line with hiccup
activation at 70% of Vo min
110 130 150 %
Short Circuit Protection Yes, Autorecovery ————
Over Temperature Protection Yes, Autorecovery (CmG, CmH latch off ) ————
Capacitive Load Nominal output voltages at full load 10 mF
CmG, CmH 270 uF
Auxiliary Output
Parameter Conditions/Description Min Nom Max Units
Nominal Output Voltage Aux voltage option A 11.76 12 12.24 V
Aux voltage option B 4.75 5 5.25 V
Load Regulation ±2 %
Line Regulation For ±10% change from nominal line ±0.5 %
Maximum Output Current Aux voltage option A 1.96 A
Aux voltage option B 4.7 A
Maximum Output
Capacitive Load
1000 uF
Output Overcurrent Protection Hiccup 110 140 %
Short Circuit Protection Yes, autorecovery ————
Galvanic Isolation
Parameter Conditions/Description Min Nom Max Units
Input to Output Reinforced (2 x MOPP); contact Advanced Energy for
Hi-Pot instructions
4000 VAC
Input to Case Basic (1 x MOPP) 1850 VAC
Output to Case Basic (1 x MOPP) 1850 VAC
Output to Output Basic (1 x MOPP) 1850 VAC
Output to Output (Dual) CmG, CmH V1-V2 500 VDC
ELECTRICAL SPECIFICATIONS (CONTINUED)
4 advancedenergy.com
COOLX600 SERIES
Reliabilty
Parameter Conditions/Description Min Nom Max Units
Reliability and MTBF MTBF of > 400 kHours, Telecordia SR-332, Issue 1 CoolMod 0.52 Fpmh
CoolPac 1.08 Fpmh
Warranty 5 years ————
Environmental
Parameter Conditions/Description Min Nom Max Units
Operating Temperature Operates to specification below -20°C after 10 min warmup -40 85 °C
Storage Temperature -40 85 °C
Derating See Derating Curves including note 5
Relative Humidity Non-condensing 5 95 %RH
Shock and Vibration MIL-STD-810G Method 514.6
Altitude 5000 m
Leakage Currents
Parameter Conditions/Description Nom Units
AC Leakage Current Input to Earth Ground
Normal Condition (High Line) Mains Voltage 264 VAC / 60 Hz 248 µA
Single Fault Condition (High Line) Mains Voltage 264 VAC / 60 Hz 393 µA
Touch Current
Normal Condition Mains Voltage 264 VAC / 60 Hz 4.7 µA
Single Fault Condition Mains Voltage 264 VAC / 60 Hz 247 µA
EMC
Parameter Conditions/Description Units
Radiated Emissions4EN 55011, EN 55022 and FCC, Class B Compliant
Conducted Emissions4EN 55011, EN 55022 and FCC, Class B Compliant
Power Line Harmonics EN 61000-3-2, Class A Compliant
Voltage Flicker EN 61000-3-3 Compliant
ESD EN 61000-4-2, level 4, 8 kV contact, 15 kV air A
Radiated Immunity EN 61000-4-3, level 2, 3 V/m A
Electrical Fast Transient EN 61000-4-4, level 4, ±4 kV B
Surge Immunity EN 61000-4-5, level 4, 2 kV DM, 4 kV CM B
Conducted RF Immunity EN 61000-4-6, level 3, 10 Vemf 150 KHz-80 MHz A
Power Frequency Magnetic Field EN 61000-4-8, level 4, 30 A/m A
4 Consult AE applications for system level compliance
Standards and Directives4
Parameter Conditions/Description
Safety Agency Approvals EN60601-1 3rd Edition, UL60601-1, CSA601
EN60950 2nd Edition, CSA C22.2 No. 60950-1
IEC/EN 60950-1, Edition 2 and
All National Deviations
UL 60950-1/CSA 22.2 No 60950-1, Edition 2
5000 m (16,400 ft) altitude, 100 V to 240 ±10% VAC
IEC/EN 60601-1, Edition 3 and
All National Deviations
IEC 60601-1(2005), EN60601-1 (2006)
ANSI/AAMI ES 60601-1 (2005)
CAN/CSA C22.2 No. 60601-1 (2008)
5,000 m (16,400 ft) altitude, 100 V to 240 VAC ±10%
ELECTRICAL SPECIFICATIONS (CONTINUED)
advancedenergy.com 5
COOLX600 SERIES
CoolX can be mounted on its base, vertically, or on its side. CoolX can also be mounted on
the DIN Rail Accessory (Z744).
ELECTRICAL SPECIFICATIONS (CONTINUED)
Standards and Directives (Continued)4
Parameter Conditions/Description
IEC 62368 Edition 2 IEC 62368-1 (2014) Edition 2
5000 m (16,400 ft) altitude, 100 V to 240 ±10% VAC
IEC 60601-1-2 Edition 4 IEC 60601-1-2 (2014)
Protection Class Class I
WEEE Waste Electrical and Electronic Equipment Directive (WEEE) 2002/96/EC
ROHS EU DIRECTIVE 2011/65/EC RoHS compliancy
REACH Compliant
4 Designed to support Type B and Type BF Applied Part End Product Requirements. BF rating achieved with CoolX600 configured without a cover.
MECHANICAL SPECIFICATIONS
Mechanica Data
Parameter Description Min
Dimensions (L x W x H) 215.9 mm x 114.3 mm x 39.1 mm (8.5 in x 4.5 in x 1 U)
Weight Nominal Weight: CoolPac + 4 x CoolMods 1 Kg
Connectors Description Mating Connectors (if applicable)
AC/DC input terminal block TE 2-1437667-S, DINKLE DT-35-B07W-03
AC/DC IEC input (Option) IEC 320 Inlet
Main DC output terminal block
(CmA-CmF, CmM-CmQ)
M4 Screws
Main DC output terminal block
(CmG, CmH)
Camden - CTB9350/4A
Output Signal Connector
(CmG, CmH)
Molex - 87833-0831
System Signal Connector J1005 Molex 87833-0831 8-way Locking Molex 51110-0860; Non Locking Molex 51110-0850;
Crimp Terminal: Molex p/n 50394 or Molex 51110-0856 which
includes locking tab and polarization keying
Output Signal Connectors
J1001-1004
Molex 87833-0631 6-way Locking Molex 51110-0660; Non Locking Molex 51110-0650;
Crimp Terminal: Molex p/n 50394 or Molex 51110-0656 which
includes locking tab and polarization keying
Output Sense Connectors J3 JST-S2BPH-K(LF)(SN) JST PHR2. Crimp Terminal JST BPH-002T-P.0.5S or SPH-002T-P.05S
Auxiliary Output Connector J1 Molex 1041880210 2pin
Mechanical Drawings
215.90
3.00
39.10
6.00
162.00
19.55
122.00 31.50
36.15 42.00
24.25 24.25 24.25
114.30
6 advancedenergy.com
COOLX600 SERIES
INTERFACE
Standard (Screw Terminal) Option 1 (IEC Terminal)
5 Enhanced thermal performance with system fans and base plate cooling. Contact Advanced Energy for details.
Input Connectors
CoolPac Connectors
advancedenergy.com 7
COOLX600 SERIES
CoolPac
CX06-600W
S/M
S = Standard
M = Medical
Aux Voltage
A = 12 V Aux (Standard)
B = 5 V Aux
Seings + CoolPac Options
N = Unconfigured
C = Conformal Coating
S = Ruggedized + Conformal Coating
P = Configured
X = Internal Use Only
CoolMods
0 = Unpopulated
# = Unavailable
A = CmA
B = CmB
C = CmC
D = CmD
E = CmE
F = CmF
G = CmG
H = CmH
M = CmM
N = CmN
P = CmP
Q = CmQ
Connectors & Leakage Current
= Screw Terminal, Normal Leakage
1 = IEC Terminal, Normal Leakage
4 = Screw Terminal, Low Leakage (150 uA)
5 = IEC Terminal, Low Leakage (150 uA)
Open Frame / Enclosed
Not used or = Standard Model
L = Cover
Logistic
Use Only
INTERFACE (CONTINUED)
CoolMod Connectors
CONFIGURATION
-= Standard indicates that all voltages are set to the nominal setpoint of each module and there are no parallel/series links fitted to the power supply
ABOUT ADVANCED ENERGY
Advanced Energy (AE) has devoted more than three decades
to perfecting power for its global customers. AE designs and
manufactures highly engineered, precision power conversion,
measurement and control solutions for mission-critical appli-
cations and processes.
AE’s power solutions enable customer innovation in complex
semiconductor and industrial thin film plasma manufacturing
processes, demanding high and low voltage applications,
and temperature-critical thermal processes.
With deep applications know-how and responsive service and
support across the globe, AE builds collaborative partnerships
to meet rapid technological developments, propel growth for
its customers and power the future of technology.
Specifications are subject to change without notice. Not responsible for errors
or omissions. ©2019 Advanced Energy Industries, Inc. All rights reserved.
Advanced Energy® and AE® are U.S. trademarks, and CoolX® is a U.S. and
European trademark of Advanced Energy Industries, Inc.
For international contact information,
visit advancedenergy.com
ENG-LV-COOLX600-235-06 09.19