Features * * * * * * Regulated Converter Reinforced insulation for 250VAC working voltage Clearance and creepage distance: 8mm 5kVAC I/P to O/P 2MOPP isolation 2A patient leakage current Industry standard pinout 2:1 and 4:1 wide input range Description The REM10 series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute isolation with low 2A leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions. The outputs are short circuit and overload protected. The converters are available in two different pinning options and optionally with an external control pin for standby consumption as low as 12.5mW. The converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry UL mark. Selection Guide Part Number Input Voltage Range [VDC] Output Voltage [VDC] Output Current [mA] Efficiency typ. [%] Max. Capacitive Load [F] REM10-xx3.3S/ (3,4) 5 / 12 / 24 / 48 3.3 REM10-xx05S/ (3,4) 5 / 12 / 24 / 48 5 2500 80 / 83 / 83 / 82.5 3000 2000 84 / 85.5 / 86.5 / 86.5 REM10-xx12S/ (3,4) 2500 5 / 12 / 24 / 48 REM10-xx15S/ (3,4) 5 / 12 / 24 / 48 12 830 86.5 / 88 / 89 / 89 430 15 670 87 / 89 / 89 / 89 350 5 / 12 / 24 / 48 24 416 85.5 / 89 / 89 / 88. 125 REM10-xxx05D/ (3,4) 5 / 12 / 24 / 48 5 1000 83 / 84 / 85 / 85 1440 REM10-xx12D/ (3,4) 5 / 12 / 24 / 48 12 416 85.5 / 89 / 89 / 88 250 REM10-xx15D/ (3,4) 5 / 12 / 24 / 48 15 333 86.5 / 88 / 89 / 88 180 REM10-xx3.3SW/ REM10-xx24S/ (3,4) 24 / 48 3.3 2500 83 / 82.5 3000 REM10-xx05SW/ (3,4) 24 / 48 5 2000 86.5 / 86.5 2500 REM10-xx12SW/ (3,4) 24 / 48 12 830 89 / 89 430 REM10-xx15SW/ (3,4) 24 / 48 15 670 89 / 89 350 (3,4) REM10-xx24SW/ (3,4) 24 / 48 24 416 89 / 88.5 125 REM10-xx05DW/ (3,4) 24 / 48 5 1000 85 / 85 1440 REM10-xx12DW/ (3,4) 24 / 48 12 416 89 / 88 250 REM10-xx15DW/ (3,4) 24 / 48 15 333 88 / 88 180 Model Numbering REM10- __ __ __ __ / __ __ Output Power CTRL Option (4) Input Voltage (1) Pinning (3) Output Voltage Input Type (2) DC/DC Converter REM10 10 Watt 2:1 & 4:1 DIP24 Single and Dual Output E314885 2MOPP 250VAC IEC/EN60601-1 certified CSA/CAN C22.2 60601-01 certified ANSI/AAMI ES60601-1 certified EN55011 certified Single or Dual Notes: Note1: for 4:1 Input Voltage Type add "W", see Note 2. 2:1 nom. Vin 4:1 "W" nom. Vin xx= 4.5-9 Vin = "05" xx= 9-36Vin = "24" xx= 9-18Vin = "12" xx= 18-75Vin = "48" xx= 18-36Vin = "24" xx= 36-75Vin = "48" Note2: Blank for Standard 2:1 Input Voltage Range; W" suffix for 4:1 Input Voltage Range Note3: A" suffix for A pinning; C" suffix for C pinning, for more details refer to Package Style and Pinning Note4: CTRL" suffix for control pin option, for A pinning only, for C pinning not available Examples: REM10-0512D/A REM10-1215S/C REM10-4815SW/A/CTRL REM10-243.3SW/C www.recom-power.com = = = = 2:1 Input, 2:1 Input, 4:1 Input, 4:1 Input, 4.5-9Vin, 9-18Vin, 36-75Vin, 9-36Vin, 12Vout, 15Vout, 15Vout, 3.3Vout, pinout A", pinout C", pinout A" pinout C", without control pin without control pin with control pin without control pin REV.: 10/2018 MED-1 REM10 DC/DC Converter Series Specifications (measured @ Ta= 25C, nominal input voltage, full load and after warm-up) BASIC CHARACTERISTICS Parameter Condition Absolute Maximum Input Voltage ( 3sec max.) Typ. Max. 2:1 16VDC 25VDC 50VDC 100VDC 4:1 24Vin nom. 48Vin nom. 50VDC 100VDC 2:1 5Vin nom. 12Vin nom. 24Vin nom. 48Vin nom. 4VDC 8VDC 16VDC 33VDC 4.5VDC 9VDC 18VDC 36VDC 4:1 24Vin nom. 48Vin nom. 8VDC 16VDC 9VDC 18VDC Under Voltage Lockout Start-up Time Min. 5Vin nom. 12Vin nom. 24Vin nom. 48Vin nom. constant resistive load, Power up or Remote ON/OFF 30ms DC-DC ON DC-DC OFF Remote ON/OFF (referenced to -Vin Pin) Open or 0-1.2VDC 2.2-12VDC Current of CTRL Pin -0.5mA Remote OFF Input Current 1mA 2.5mA Internal Operating Frequency 270kHz 10F/25V X7R MLCC for 3.3, 5Vout 10F/25V X7R MLCC for 12, 15Vout 4.7F/50V X7R MLCC for 24Vout Output Ripple and Noise (20MHz BW limited) Efficiency vs. Load 300kHz 30mVp-p 40mVp-p 50mVp-p 330kHz REM10-1212D 100 90 Efficiency [%] 80 70 60 50 40 30 20 10 0 10 20 30 40 50 60 70 80 90 100 Load [%] REGULATIONS Parameter Condition Type Output Accuracy Line Regulation Value 1.0% low line to high line Single Dual 0.2% 0.5% Load Regulation no load to full load Single Dual 0.2% 1.0% Cross Regulation asymmetrical load 25% / Full Load only Dual Output Transient Response www.recom-power.com 25% load step change REV.: 10/2018 5.0% 250s MED-2 REM10 DC/DC Converter Series Specifications (measured @ Ta= 25C, nominal input voltage, full load and after warm-up) PROTECTIONS Parameter Condition Type Value Short Circuit Protection (SCP) continuous, auto-recovery Over Load Protection (OLP) % of Iout rated Hiccup mode, 150% typ. 3.3Vout 5Vout 12Vout 15Vout 24Vout 3.7VDC min. / 5VDC max. 5.6VDC min. / 7VDC max. 13.5VDC min. / 16VDC max. 18.3VDC min. / 22VDC max. 29.1VDC min. / 34.5VDC max. 5Vout Dual 12Vout 15Vout 5.6VDC min. / 7VDC max. 13.5VDC min. / 18.2VDC max. 17VDC min. / 22VDC max. 5kVAC / 1 minute 250VAC / continuous 12pF typ. / 17pF max. Single Output Over Voltage Protection (OVP) Isolation Voltage I/P to O/P working voltage Isolation Capacitance Leakage Current Insulation Grade 240VAC, 60Hz 2A reinforced Means of Protection 2MOPP Medical Device Classification built-in power supply Internal Clearance and Creepage External Clearance and Creepage Notes: Note5: 8mm I/P to O/P "C" Pinning "A" Pinning I/P to O/P >19.72mm >14.64mm This Power module is not internally fused. A input line fuse must be always used Recomended Fuse: 2:1 Input Voltage Fuse (slow blow) 4:1 Input Voltage Fuse (slow blow) 5V T5A 24V T2A 12V T2A 48V T1A 24V T1A 48V T0.5A ENVIRONMENTAL Parameter Condition Value Maximum Case Temperature Range Maximum Ambient Temperature Range -40C to +105C see thermal calculation below Temperature Coefficient Thermal Impedance 0.02%/K typ. natural convection 0.1m/s 18K/W Operating Altitude Operating Humidity 5000m non-condensing 5% - 95% RH max. Pollution Degree MTBF PD2 according to MIL-HDBK-217F, G.B. +25C 3849 x 10 hours 3 Thermal Shock according to MIL-STD-810F standard Vibration according to MIL-STD-810F standard continued on next page www.recom-power.com REV.: 10/2018 MED-3 REM10 DC/DC Converter Series Specifications (measured @ Ta= 25C, nominal input voltage, full load and after warm-up) Thermal Calculation: Tcase hset = hfull load x fh Pdiss = Pout set hset - Pout set Tover = Rth x Pdiss Tamb = Tcase - Tover = baseplate temperature Tover = temperature losses [C] Tamb = ambient temperature [C] Pout nom. = nom. output power [W] Pout set = output power set [W] Pdiss = internal losses [W] Rth = thermal impedance [K/W] hset hfull load = efficiency set [%] = efficiency @ full load [%] fh = efficiency factor [] Efficiency Crosstable (%) @ full Load Efficiency Factor vs. Load 1.2 5 80 84 86.5 87 85.5 83 855 86.5 12 83 85.5 88 89 89 84 89 88 24 83 86.5 89 89 89 85 89 88 48 82.5 86.5 89 89 88.5 85 88 88 24W 83 86.5 89 89 89 85 89 88 48W 82.5 86.5 89 89 88.5 85 88 88 Efficiency Factor [ ] Output Voltage Input Voltage 3.3S 05S 12S 15S 24S 05D 12D 15D [C] 1.0 0.8 0.6 0.4 0.2 0.0 0 10 20 30 40 50 60 70 80 90 100 Load [%] Practical Example: Take the REM10-1212D with 12V Input Voltage, 50% load. What is the maximum ambient operating temperature? Tcase = 105C Pout nom. Pout set Rth hfull load = 89% (Crosstable) fh = 0.991 (Graph) hset = 89 x 0.991 88.19% = 10W = 10 x 0.5 = 5W = 18K/W Pdiss = 5 0.882 -5= 0.67W Tover = 18 x 0.67 = +12.1C Tamb = 105 - 12.1 = +92.9C SAFETY AND CERTIFICATIONS Certificate Type (Safety) Report / File Number Standard Medical Electric Equipment, General Requirements for Safety and Essential Performance E314885-A6-CB-1 CAN/CSA-C22.2 No. 60601-1:08 ANSI/AAMI ES60601-1:2005 Medical Electric Equipment, General Requirements for Safety and Essential Performance (CB Scheme) E314885-A6-CB-1 IEC60601-1:2005 + C2:2007 3rd Edition EN60601-1:2006 EAC RU-AT.49.09571 RoHS2+ TP TC 004/2011 TP TC 004/2011 RoHS-2011/65/EU + AM-2015/863 continued on next page www.recom-power.com REV.: 10/2018 MED-4 REM10 DC/DC Converter Series Specifications (measured @ Ta= 25C, nominal input voltage, full load and after warm-up) Certificate Type (Others) Conditions Standard / Criterion Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests EN60601-1-2:2015 Industrial, scientific and medical equipment - Radio frequency disturbance characteristics - Limits and methods of measurement (7) EN55011:2009 + A1:2010 Class A & B ESD Electrostatic discharge immunity test Air 15kV; Contact 8kV EN61000-4-2:2008 10V/m (80-2500MHz) 27V/m (385MHz) 28V/m (450MHz) Radiated, radio-frequency, electromagnetic field immunity test EN61000-4-3:2006 + A2:2010 Fast Transient and Burst Immunity (6) DC Port: 2kV EN61000-4-4:2012 Surge Immunity (6) DC Port: 2kV EN61000-4-5:2005 Immunity to conducted disturbances, induced by radio-frequency fields 6Vr.m.s EN61000-4-6:2013 Power Frequency Magnetic Field 30A/m EN61000-4-8:2009 Notes: Note6: An external input filter capacitor is required if the model has to meet EN61000-4-4 or/and EN61000-4-5 Recommended components: 5Vin 12Vin, 24Vin 48Vin Note7: aluminium capacitor (Nippon Chemi-con KY series, 1000F/25V) and a reverse diode (Vishay V10P45) to connect in parallel aluminium capacitor (Nippon Chemi-con KY series, 470F/50V) aluminium capacitor (Nippon Chemi-con KY series, 330F/100V) The whole REM10 series can meet EMI Class A with no external filter. And Class B only with external components EMC Filter Suggestion for Class B (8) L1 +Vin L2 C2 C1 +Vin +Vout -Vin -Vout C3 -Vin MODEL +Vout -Vout C1 (8) C2 (8) C3 (8) L1 (8) L2 (8) N/A 22F/16V MLCC 22F/16V MLCC 3.3H; 3.3A SMD Inductor 52H CMC REM10-12xxS_D REM10-24xxS_D(/W) 4.7F/50V MLCC 4.7F/50V MLCC N/A 10H; 2.3A SMD Inductor 175H CMC REM10-48xxS_D(/W) 1F/100V MLCC 4.7F/100V MLCC N/A 10H; 2.3A SMD Inductor 419H CMC REM10-05xxS_D Notes: Note8: The component values can be adapted according to customers application DIMENSION and PHYSICAL CHARACTERISTICS Parameter Type Value Material case PCB potting non-conductive black plastic (UL94-V2) FR4 (UL94-V1) silicone (UL94-V0) Dimension (LxWxH) 31.80 x 20.30 x 10.40mm Weight 14g continued on next page www.recom-power.com REV.: 10/2018 MED-5 REM10 DC/DC Converter Series Specifications (measured @ Ta= 25C, nominal input voltage, full load and after warm-up) Dimension Drawing (mm) "A" Pinning (Standard) embossed logo 20.3 10.4 31.8 Pin Connections Pin # Single Dual 1 CTRL* CTRL* 2 -Vin -Vin 11 NC -Vout 14 +Vout +Vout 16 -Vout Com 22 +Vin +Vin 23 +Vin +Vin * If dont choose CTRL option, there is no pin on the corresponding pin number 3.80 Recommended Footprint Details O0.90 +0.10/-0 2.54 O0.60 +0.10 / -0.05 11 Bottom View 23 22 16 14 16 14 Top View 1 2 11 2.54 2.54 clearance = 14.64 2.00 23 22 6x2.54=15.24 1 2 NC= not connected Tolerance: xx.x= 0.5mm xx.xx= 0.25mm "C" Pinning embossed logo 20.3 10.4 31.8 Pin Connections Pin # Single Dual 1 +Vin +Vin 11 No Pin Com 12 -Vout No Pin 13 +Vout -Vout 15 No Pin +Vout 23 -Vin -Vin 24 -Vin -Vin 3.80 O0.60+0.10 / -0.05 Recommended Footprint Details 11 12 Bottom View 24 23 2.00 www.recom-power.com 15 13 clearance = 19.72 2.54 1 6x2.54=15.24 2.54 O0.90 Tolerance: xx.x= 0.5mm xx.xx= 0.25mm +0.10/-0 24 23 15 13 Top View 1 11 12 2.54 REV.: 10/2018 MED-6 REM10 DC/DC Converter Series Specifications (measured @ Ta= 25C, nominal input voltage, full load and after warm-up) PACKAGING INFORMATION Parameter Type Value Packaging Dimension (LxWxH) tube 255 x 21.8 x 16.5mm Packaging Quantity 7pcs Storage Temperature Range Stoarge Humidity -55C to +125C non-condensing 5% to 95% RH max. 16.50.75 11.50.75 0.50.2 Tube Dimension Drawing (mm) 110.75 21.80.75 The product information and specifications may be subject to changes even without prior written notice.The product has been designed for various applications; its suitability lies in the responsibility of each customer. The products are not authorized for use in safety-critical applications without RECOM's explicit written consent. A safety-critical application is an application where a failure may reasonably be expected to endanger or cause loss of life, inflict bodily harm or damage property. The applicant shall indemnify and hold harmless RECOM, its affiliated companies and its representatives against any damage claims in connection with the unauthorized use of RECOM products in such safety-critical applications. www.recom-power.com REV.: 10/2018 MED-7