Agilent
U2741A USB Modular 5.5 Digits Digital Multimeter
Data Sheet
2
Features
Makes fast measurements
with up to 100 readings per
second
Measures up to 300 VDC with
5.5 digits resolution
Frequency and temperature
measurement capability
Wide voltage measurement
range
- DC: 1 μVDC to 300 VDC
- AC: 1 μVrms to 250 Vrms
Wide current measurement
range
- DC: 1 μADC to 2 ADC
- AC: 1 μArms to 2 Arms
Compatibility with Hi-Speed
USB 2.0, USBTMC 488.2
standards
Dual-play — standalone and
modular capability
Bundled software — Agilent
Measurement Manager
(AMM)
Wide range of compatible
Agilent Development
Environments (ADEs)
Introduction
The Agilent U2741A is a 5.5 digits digital multimeter (DMM), which is the latest
addition to the Agilent’s USB modular family. It can operate as standalone or as
a modular unit when used with the U2781A USB modular instrument chassis. It
comes with various features and functions to meet your needs today.
U2741A use as standalone device U2741A use as modular device
Figure 1. The U2741A DMM showing dual-play capability
Various U2741A features to meet your demands
Affordable and measures up to ten different measurements
Quick and easy Hi-Speed USB 2.0 connectivity
Dual-play capability allowing fl exible confi guration in simple or
sophisticated test environment
Easy-to-use bundled software to get you started with the DMM without
programming
Compatible with a wide range of ADEs to choose
Portable, accurate, and reliable
Value for money and feature-
rich DMM
The U2741A DMM gives you an
accurate 5.5 digits resolution. It
offers a wide range of measurement
functions and features despite its
value price. The U2741A is capable of
performing DC voltage, AC voltage, DC
current, AC current, 2-wire resistance,
4-wire resistance, continuity,
diode, frequency, and temperature
measurement to meet the industrial
needs.
Quick and easy connectivity
The U2741A allows quick connection
to the PC with Hi-Speed USB 2.0.
The USB interface is also compliant
with USBTMC 488.2 standards. You
can easily connect the device using
USB cable which makes the setup
easy and hassle-free. This simple
and easy-to-use device is ideal for
design validation and educational
teaching labs. Its robust design is
also appropriate for manufacturing
eld, use in consumer electronics and
semiconductor.
3
Dual-play capability
The U2741A has a breakthrough
design which gives the users great
exibility to use the device as a
standalone or as a modular unit
used in the U2781A USB modular
instrument chassis. You can use this
DMM as standalone for simple test
confi gurations or slot the DMM into
the chassis for more sophisticated
applications systems that requires
connection and triggering of other
modular devices.
Supports IVI-COM, SCPI, and
compatible with wide range of
ADEs
All U2741A purchases are shipped
with the IVI-COM drivers. With the
offered IVI-COM drivers, you are able
to program the U2741A with any
popular ADE that is available in the
market. The U2741A is compatible
with a wide range of ADEs that
allows you to set up your devices in
different programming environment
with minimal effort as they are ready
to be programmed directly using SCPI
commands.
The popular development environment
that is compatible with U2741A is
listed below:
Agilent VEE
Microsoft® Visual Studio .NET
C/C++ and Visual Basic®
LabVIEW
Easy-to-use bundled software
The Agilent measurement manager (AMM) bundled application software is
simple and easy to use. This software offers a standard user-friendly graphical
user interface (GUI) that allows you to confi gure and control your U2741A
remotely without performing any programming task.
Data logger function with AMM
The AMM software comes with data logger function that allows you to
customize your data logging needs. Data can be stored for analysis when this
option is enabled. The data capture interval range can be set from 0 s to 100 s.
With this feature, you can start the U2741A to take measurement over duration
and check your results when the test is completed. Data can be stored to a CSV
(comma separated value) fi le format.
Figure 2. Example of U2741A AMM GUI
4
Product Outlook and Dimensions
System Requirements
Processor 1.6 GHz Pentium® IV or higher
Operating system Windows® XP Professional or Home Edition (Service
Pack 1 or later) or Windows® 2000 Professional
(Service Pack 4 or later)
Browser Microsoft® Internet Explorer 5.01 or higher
Available RAM 512 MB or higher recommended
Hard-disk space 1 GB
Video Super VGA 800×600
Prerequisites Agilent IO Libraries Suite 14.2 or higher (version 15.0[1]
recommended),
Agilent T&M Toolkit Runtime version 2.1[2],
Agilent T&M Toolkit Redistributable Package 2.1
patch[2],
Microsoft® .NET Framework version 1.1 and 2.0[2]
[1] Available on Agilent Automation-Ready CD.
[2] Bundled with Agilent Measurement Manager application software installer.
Standard Shipped
Accessory
12 V, 2 A AC/DC adapter
Power cord
Standard test lead
USB Standard-A to Mini-B
interface cable
L-Mount kit (used with the
modular instrument chassis)
Agilent Automation-Ready CD
(contains the Agilent IO Libraries
Suite)
Agilent USB Modular Products
Quick Start Guide
Agilent USB Modular Products
Reference CD-ROM
Agilent Measurement Manager
Quick Reference Card
Certifi cate of Calibration
Optional Accessory
E2308A thermistor temperature probe
Front View Rear View
Top View
41.00 mm
180.00 mm
117.00 mm
8.60 mm
5
General Characteristics
Remote interface Hi-Spe ed USB 2.0
USBTM C 488.2 Class device
Power consumption +12 VDC, 2 A maximum
Isolated ELV power source
Operating environment One year specifi cations full accuracy from 18 °C to 28 °C at 50% RH
Altitude up to 2000 meters
Pollution degree 2
For indoor use only
Storage compliance 20 °C to +70 °C
Relative hum idity at 5% to 90% RH (non-condensing)
Safety compliance Certifi ed with:
IEC 61010-1:2001/EN61010-1:2001 (2nd Edition)
Canada: CAN/CSA-C22.2 No. 61010-1-04
USA: ANSI/UL 610• 10-1:2004
Measurement compliance CAT II 300 V Over-voltage Protection
EMC compliance IEC 61326-2002/EN61326:1997+A1:1998+A2:2001+A3:2003
Canada: ICES-001:2004
Australia/New Zealand: AS/NZS CISPR11:2004
Common mode rejection DC CMRR >120 dB with 1k unbalanced load
AC CMRR >70 dB at 50/60 Hz ±0.1% with 1k unbalanced load
Normal mode rejection
ratio (NMRR)
>60 dB at 50/60 Hz ±0.1%[1]
>0 dB at 50/60 Hz ±0.1%[2]
Shock and vibration Tested to IEC/EN 60068-2
I/O connector Four banana socket terminals
Dimensions (W × D × H) 105.0 0 mm x 175.00 mm x 11.50 mm (without bumpers)
117.00 mm x 180.00 mm x 41.00 mm (with bumpers)
Weight 451 g (without bumpers)
509 g (with bumpers)
Warranty One year
Calibration Calibration interval of one year is highly recommended
[1] Applicable for NPLC >1.
[2] Applicable for NPLC 0.2 and 0.02.
6
Product Specifi cations
DC specifi cations[1]
Function Range Input impedance
Test current /
Burden voltage,
shunt resistance
Accuracy
±(% of reading +
% of range)
Temperature
Coeffi cient
0 °C to 18 ºC
28 °C to 55 °C
Voltage[2] 100.000 mV 10 MΩ 0.015 + 0.008 0.002 + 0.0008
1.00000 V 10 MΩ 0.015 + 0.005 0.001 + 0.0005
10.0000 V 10 MΩ 0.018 + 0.005 0.002 + 0.0005
100.000 V 10 MΩ 0.018 + 0.005 0.002 + 0.0005
300.000 V 10 MΩ 0.018 + 0.005 0.002 + 0.0005
Current[3] 10.0000 mA <0.2 V, 10 Ω0.06 + 0.015 0.005 + 0.0025
100.000 mA <0.2 V, 1 Ω0.06 + 0.005 0.008 + 0.002
1.0000 A <0.3 V, 0.1 Ω0.15 + 0.007 0.005 + 0.002
2.0000 A <0.8 V, 0.1 Ω0.15 + 0.007 0.005 + 0.002
Resistance[4] 100.000 Ω 1.0 mA 0.03 + 0.008 0.006 + 0.0008
1.00000 kΩ 1.0 mA 0.03 + 0.005 0.006 + 0.0005
10.0000 kΩ 100 μA 0.03 + 0.005 0.006 + 0.0005
100.000 kΩ 10.0 μA 0.03 + 0.005 0.006 + 0.0005
1.00000 MΩ 1 μA 0.06 + 0.005 0.01 + 0.0005
10.0000 MΩ 225 nA 0.25 + 0.005 0.025 +0.0005
100.000 MΩ 225 nA // 10MΩ2.0 + 0.005 0.3 + 0.0005
Diode test[5] 1.0000 V 1.00 mA 0.015 + 0.03 0.005 + 0.0005
Continuity Test[6] 1.0000 kΩ 1.00 mA 0.05 + 0.03 0.005 + 0.0005
[1] Specifi cations are based on 30 minutes warm-up time, NPLC 20 resolution, and calibration temperature within 18 °C
to 28 °C. For NPLC 0 and 0.025, add 0.01% of range.
[2] 120% over range on all ranges except 300 VDC. Input protection up to 300 VDC.
[3] Input protected with externally accessible 2 A, 250 V fast blown fuse.
[4] Specifi cations are for 4-wire ohm or 2-wire ohms using null function in AMM software. If without null function in
AMM software, add 0.2 Ω additional error. Input protection up to 300 Vdc. Specifi cations apply for NPLC ≥1.
[5] Specifi cations are for the voltage measured at the input terminals only.
[6] Continuity threshold is fi xed at less than 10 Ω.
DC current settling considerations
Switching current from >1 A to lower current measurement may cause an additional error approximately 0.3% of reading
due to self- heating, and will generally dissipate within a minute.
7
AC specifi cations[1]
AC accuracy for voltage
Function Range
Accuracy input ±(% of reading + % of range)
Frequency (Hz)
20 ~ 45 45 ~ 10k 10k ~ 30k 30k ~ 100k[3]
Voltage[2] 100.000 mVrms 1 + 0.1 0.2 + 0.1 1.5 + 0.3 5.0 + 0.3
1.00000 V 1 + 0.1 0.2 + 0.1 1.0 + 0.1 3.0 + 0.2
10.0000 V 1 + 0.1 0.3 + 0.1 1.0 + 0.1 3.0 + 0.2
100.000 V 1 + 0.1 0.3 + 0.1 1.0 + 0.1 3.0 + 0.2
250.000 V[4] 1 + 0.1 0.3 + 0.1 1.0 + 0.1 3.0 + 0.2
[1] Specifi cations are based on 30 minutes warm-up time and calibration temperature within 18 °C to 28 °C. In manual
range, the settling time is 2.6 seconds while in autorange, the fi rst measurement accuracy is <1% .
[2] Specifi cations are for sine wave inputs more than 5% of range. 120% over range on all ranges except 250 VAC.
Maximum crest factor of 5 at full scale. Input impedance is 1 MΩ in parallel with capacitance less than 120 pF, AC
couple with up to 300 VDC.
[3] Additional error to be added as frequency more than 30 kHz and signal input less than 10% of range. 30 kHz to
100 kHz: add 0.003% of range per kHz.
[4] Input signal has to be more than 50 Vrms.
Temperature coeffi cient for voltage (0 ºC to 18 ºC, 28 ºC to 55 ºC)
Range
Frequency (Hz)
(% of reading + % of range)
20 ~ 45 45 ~ 10k 10k ~ 30k 30k ~ 100k
Temperature
coeffi cient 100.000 mVrms,
1.00000 V,
10.0000 V,
100.000 V,
250.000 V
0.02 + 0.02 0.02 + 0.02 0.05 + 0.02 0.1 + 0.02
AC accuracy for current[1]
Function Range Burden voltage /
Current shunt
Accuracy input (% of reading + % of range)
Frequency (Hz)
20 ~ 45 45 ~ 1k 1k ~ 10k
Current[2] 10.0000 mA <0.2 V, 10 Ω1.5 + 0.1 0.5 + 0.1 2 + 0.2
100.000 mA <0.2 V, 1 Ω1.5 + 0.1 0.5 + 0.1 2 + 0.2
1.00000 A <0.3 V, 0.1 Ω1.5 + 0.1 0.5 + 0.1 2 + 0.2
2.0000 A <0.8 V, 0.1 Ω1.5 + 0.1 0.5 + 0.1 2 + 0.2
[1] In manual range, the settling time is 2.6 seconds while in autorange, the fi rst measurement accuracy is <1%.
[2] Input protected with externally accessible 2 A, 250 V fast blown fuse.
8
Temperature coeffi cient for current (0 ºC to 18 ºC, 28 ºC to 55 ºC)
Range
Frequency (Hz)
(% of reading + % of range)
20 ~ 45 45 ~ 10k 10k ~ 30k
Temperature
coeffi cient 10.0000 mA,
100.000 mA,
1.00000 A,
2.0000 A
0.02 + 0.02 0.02 + 0.02 0.02 + 0.02
Frequency specifi cations[1]
Frequency accuracy
Function Range
Accuracy
(% of reading + % of
range)
Minimum input
frequency
Temperature coeffi cient
(% of range)
Frequency 20 to 300 kHz 0.0200 + 0.003 1 Hz 0.005
Frequency sensitivity for AC voltage
Function Input range
Minimum sensitivity (RMS sine wave)
Frequency (Hz)
20 ~ 100 k 100 ~ 300 k
AC voltage 100 mV[2] 20 mV 20 mV
1 V 100 mV 120 mV
10 V 1 V 1.2 V
100 V 10 V 20 V
250 V 100 V 120 V
[1] Frequency measurement can only be done in auto range mode. Specifi cations are for 30 minutes warm-up time, using
one second aperture. Measuring method is using reciprocal counting technique with AC coupled input at AC voltage
function. Gate time of 0.1 second or 1 second.
[2] Only applicable for square wave measurement.
Temperature specifi cations
Temperature accuracy
Function Thermistor type Range Accuracy Temperature coeffi cient
Temperature 5 kΩ thermistor –80.0 °C to 150 °C
–112 °F to 302 °F Probe accuracy + 0.2 °C 0.002 °C
Remove all doubt
Our repair and calibration services
will get your equipment back to you,
performing like new, when prom-
ised. You will get full value out of
your Agilent equipment through-
out its lifetime. Your equipment
will be serviced by Agilent-trained
technicians using the latest factory
calibration procedures, automated
repair diagnostics and genuine parts.
You will always have the utmost
confi dence in your measurements.
Agilent offers a wide range of ad-
ditional expert test and measure-
ment services for your equipment,
including initial start-up assistance,
onsite education and training, as
well as design, system integration,
and project management.
For more information on repair and
calibration services, go to:
www.agilent.com/fi nd/removealldoubt
Agilent Email Updates
www.agilent.com/fi nd/emailupdates
Get the latest information on the
products and applications you select.
Agilent Direc
t
www.agilent.com/fi nd/agilentdirect
Quickly choose and use your test
equipment solutions with confi dence.
Agilent
Open
www.agilent.com/fi nd/open
Agilent Open simplifi es the process
of connecting and programming
test systems to help engineers
design, validate and manufacture
electronic products. Agilent offers
open connectivity for a broad range
of system-ready instruments, open
industry software, PC-standard I/O
and global support, which are
combined to more easily integrate
test system development.
www.agilent.com
www.agilent.com/fi nd/U2741A
For more information on Agilent Technologies’
products, applications or services, please
contact your local Agilent office. The
complete list is available at:
www.agilent.com/fi nd/contactus
Americas
Canada (877) 894-4414
Latin America 305 269 7500
United States (800) 829-4444
Asia Pacifi c
Australia 1 800 629 485
China 800 810 0189
Hong Kong 800 938 693
India 1 800 112 929
Japan 0120 (421) 345
Korea 080 769 0800
Malaysia 1 800 888 848
Singapore 1 800 375 8100
Taiwan 0800 047 866
Thailand 1 800 226 008
Europe & Middle East
Austria 01 36027 71571
Belgium 32 (0) 2 404 93 40
Denmark 45 70 13 15 15
Finland 358 (0) 10 855 2100
France 0825 010 700*
*0.125 €/minute
Germany 07031 464 6333**
**0.14 €/minute
Ireland 1890 924 204
Israel 972-3-9288-504/544
Italy 39 02 92 60 8484
Netherlands 31 (0) 20 547 2111
Spain 34 (91) 631 3300
Sweden 0200-88 22 55
Switzerland 0800 80 53 53
United Kingdom 44 (0) 118 9276201
Other European Countries:
www.agilent.com/fi nd/contactus
Revised: July 17, 2008
Product specifi cations and descriptions
in this document subject to change
without notice.
© Agilent Technologies, Inc. 2008
Printed in USA, Aug 18, 2008
5989-9361EN