ZTP-115
Thermopile Infrared (IR)
Sensor
Thermometrics Thermopile Infrared (IR)
Sensors are used for non-contact surface
temperature measurement. The ZTP-115 Model
consists of thermo-elements, at IR lter, and
a thermistor for temperature compensation
in a hermetically-sealed small-size package.
Various lters are also available to maximize
performance in specic applications.
Applications
Patient monitoring
Ear & Tympanic thermometers
Occupancy detection
HVAC
Appliance
Features
Non-contact measurement
Wider surface area measurement
Small-size sensor package
Included ambient temperature (thermistor)
sensor for compensation
High sensitivity
Fast response time
Low cost
RoHS compliant
Amphenol
Advanced Sensors
T ambient
(° C)
Rmin (kW) Rcent (kW) Rmax (kW)
-40 267 284.7 303.2
-35 197.2 209.6 222.5
-30 147.1 155.9 165
-25 110.8 117.1 123.6
-20 84.16 88.68 93.35
-15 64.46 67.73 71.11
-10 49.74 52.13 54.59
-5 38.65 40.41 42.21
0 30.24 31.54 32.86
5 23.81 24.77 25.75
10 18.86 19.58 20.3
15 15.03 15.56 16.1
20 12.04 12.44 12.84
25 9.7 10 10.3
30 7.823 8.082 8.342
35 6.342 6.566 6.791
40 5.168 5.361 5.557
45 4.233 4.4 4.569
50 3.484 3.629 3.776
55 2.882 3.007 3.135
60 2.396 2.504 2.615
65 2.0009 2.095 2.192
70 1.679 1.7612 1.8458
75 1.4153 1.4871 1.5612
80 1.1984 1.261 1.3264
85 1.0193 1.0745 1.1317
90 0.8707 0.9193 0.9697
95 0.7469 0.7898 0.8344
100 0.6433 0.6812 0.7208
105 0.5562 0.5899 0.625
Specications
Thermopile Chip
Parameter
Limits
Units Condition
Min Typ Max
Chip Size 1.8 x 1.8 mm2
Diaphragm
Size 1.0 x 1.0 mm2
Active Area 0.5 x 0.5 mm2
Internal
Resistance 35 50 65 kW25°C
Resistance T.C 0.15 %/°C
Responsivity 42 60 78 V/W 500K, 1 Hz
Responsivity
T.C -0.10 %/°C
Noise Voltage 30 nV rms R.M.S.,
25°C
NEP 0.50 nW/Hz1/2 500K, 1 Hz
Detectivity 1.00 E08 cmHz1/2/W 500K, 1 Hz
Time Constant 20 ms
Parameter
Limits
Units Condition
Min Typ Max
Resistance 9.7 10 10.3 kWTol. :3%, @25°C
Beta - Value 3930 3970 4010 K Tol. :1%, Dened
at 25°C/85°C
Absolute Maximum Ratings
Operating Temperature
-20°C ~ 100°C
Storage Temperature
-40°C ~ 120°C
Thermistor for Temperature
Compensation
Thermistor Resistance (R-T Table)
Lead Length (mm) Part No.
3.0 ZTP-115
6.7 ZTP-115L1
13.2 ZTP-115L
Ordering Information
Sensitivity Field of View
Filter Transmission Data
Parameter Limits Units Condition
Min Typ Max
Field of
View
50 55 60 Degree 50% of
Maximum
Output
0 20 40 60 80 100
-1
0
1
2
3
Output[mV]
Object Temperature[
o
C] @Tamb.=25[
o
C
]
-60 -50 -40 -30 -20 -10 0 10 20 30 40 50 60
0.0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1.0
Normalized Output
Degree
2 4 6 8 10 12 14 16 18 20
0
20
40
60
80
100
T rans miss ion(%)
W avelength (um)
Unit: mm
Pin arrangement:
1. GND Thermistor
2. Thermopile
3. GND Thermopile
4 . Thermistor
Outline of Sensor Packaging and Pin Arrangement (unit: mm)
LEAD LENGTH :
Variable (3.0, 6.7, 13.2mm±0.5mm)
AAS-920-549B - 04/2019
www.amphenol-sensors.com
© 2019 Amphenol Corporation. All Rights Reserved. Specications are subject to change without notice.
Other company names and product names used in this document are the registered trademarks or
trademarks of their respective owners.
Amphenol
Advanced Sensors