Embedded Power for Business-Critical Continuity Rev. 9.8.09_28 LPS40-M Series of 3 LPS40-M Series 40 Watts Medical Total Power: Input Voltage: # of Outputs: 40-55 Watts 85-264 VAC 120-300 VDC Single Electrical Specifications Special Features * * * * * * * * * * * Medical safety approvals Universal input 3" x 5" footprint Less than 1U high Remote sense on main output Low output ripple Adjustable 5V output Overvoltage protection Overload protection 110 KHz switching frequency LPX40 enclosure kit available Input Input range: 85 - 264 VAC; 120 - 300 VDC Frequency: 47 - 440 Hz Inrush current: < 18A peak @ 115 VAC; <36A peak @ 230 VAC, cold start @ 25 C Input current: 1A max. (RMS) @ 115 VAC Efficiency: 70% typical at full load EMI filter: FCC Class B conducted; CISPR 22 Class A conducted; EN55022 Class A conducted; VDE 0878 PT3 Class A conducted < 75A @ 50/60 Hz, 264 VAC input Safety ground leakage current: Output Maximum power: 40 W for convection; 55 W with 30CFM forced air Adjustment range: -5, +10% minimum Hold-up time: 20ms @ 40 W load, 115 VAC nominal line Overload protection: Safety * VDE 0750/EN60601-1 (IEC601) * UL UL2601 E182560 * CSA CSA 22.2 No.601.1 LR109492 * CE Mark (LVD) Short circuit protection on all outputs Case overload protected @ 110 - 145% above peak rating Overvoltage protection: 5V output; 5.7 to 6.7 VDC. Other outputs 10% to 25% above nominal output Remote sense: Compensates for 0.5 V lead drop minimum. Will operate without remote sense connected. Reverse connection protected. Embedded Power for Business-Critical Continuity Rev. 9.8.09_28 LPS40-M Series of 3 Environmental Specifications Operating temperature: 0 to 50 C ambient. Derate each output 2.5% per degree from 50 to 70 C, -20 C startup Storage temperature: -40 C to +85 C Temperature coefficient: 0.4% per C Electromagnetic Designed to meet IEC 801, -2, -3, -4, -5, -6, Level 3 susceptibility: Humidity: Operating; non-condensing 5% to 95% RH Vibration: Three orthogonal axes, sweep at 1 oct/min, 5 min. dwell at four major resonances 0.75G peak 5 Hz to 500 Hz, operational > 550,000 hours at full load and 25 C ambient conditions MTBF demonstrated: Ordering Information Output Voltage Minimum Load LPS42-M 5V 0A LPS43-M 12 V 0A Model Number Maximum Load with Convection cooling Maximum Load with 30CFM Forced Air Peak Load1 Regulation2 8A 11 A 12 A 2% 50mV 3.3 A 4.5 A 5A 2% 120 mV LPS44-M 15 V 0A 2.6 A 3.6 A 4A 2% LPS45-M 24 V 0A 1.6 A 2.3 A 2.5 A 2% 1. Peak current lasting < 30 seconds with a maximum 10% duty cycle. 2. At 25 C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings. 3. Peak-to-peak with 20 MHz bandwidth and 10 F in parallel with a 0.1 F capacitor at rated line voltage and load ranges. Power Derating Curve 55W 30 CFM Forced Air Cooling 40W 30W Output Power (Watts) 20W Natural Convection Cooling 27.5W 20W 10W 0W 0C 10C 20C 30C 40C 50C 60C 70C TA - Ambient Temperature - C Ripple P/P (PARD)3 150 mV 240 mV Embedded Power for Business-Critical Continuity Mechanical Drawing 0.22" (5.6) Americas 4.55" (115.6) 5810 Van Allen Way Carlsbad, CA 92008 USA Telephone: +1 760 930 4600 Facsimile: +1 760 930 0698 5.00" (127) M1 GND SK201 3.00" (76.2) 2.55" (64.8) 3 1 SK 1 Rev. 9.8.09_28 LPS40-M Series of 3 M2 _ 1 2 3 4 5 6 N + Remote Sense Connector Voltage Adj. Pot SK 2 Europe (UK) Waterfront Business Park Merry Hill, Dudley West Midlands, DY5 1LX United Kingdom Telephone: +44 (0) 1384 842 211 Facsimile: +44 (0) 1384 843 355 Asia (HK) L 0.15" (4) DIA (4 places) 14/F, Lu Plaza 2 Wing Yip Street Kwun Tong, Kowloon Hong Kong Telephone: +852 2176 3333 Facsimile: +852 2176 3888 For global contact, visit: www.PowerConversion.com techsupport.embeddedpower @emerson.com 1.14" (29) While every precaution has been taken to ensure accuracy and completeness in this literature, Emerson Network Power assumes no responsibility, and disclaims all liability for damages resulting from use of this information or for any errors or omissions. 0.197" (5) Emerson Network Power. The global leader in enabling business-critical continuity. Pin Assignments Mating Connectors Connector LPS42-M LPS43-M LPS44-M LPS45-M AC Input: SK1-1 SK1-3 SK2-1 SK2- 2 SK2- 3 SK2-4 SK2-5 SK2-6 SK201-1 SK201-2 Neutral Line +5V +5V +5V Common Common Common +Sense -Sense Neutral Neutral Line Line +12V +15V +12V +15V +12V +15V Common Common Common Common Common Common +Sense +Sense -Sense -Sense Neutral Line +24V +24V +24V Common Common Common +Sense -Sense Molex 09-50-8031 (USA) 09-91-0300 (UK) PINS: 08-52-0113 DC Outputs: Molex 09-50-8061 (USA) 09-91-0600 (UK) PINS: 08-52-0113 Remote Sense: Molex 22-01-2025 PINS: 08-52-0123 Emerson Network Power Connector Kit #70-841-006, includes all of the above 1. Specifications subject to change without notice. 2. All dimensions in inches (mm), tolerance is 0.02" ( 0.5mm) 3. Mounting holes M1 and M2 should be grounded for EMI purposes. 4. Mounting hole M1 is safety ground connection. 5. Specifications are for convection rating at factory settings at 115 VAC input, 25 C unless otherwise stated. 6. Warranty: 2 year 7. Weight: 0.5lbs/0.23kg AC Power Connectivity DC Power Embedded Computing Embedded Power Monitoring Outside Plant Power Switching & Controls Precision Cooling Racks & Integrated Cabinets Services Surge Protection EmersonNetworkPower.com Emerson Network Power and the Emerson Network Power logo are trademarks and service marks of Emerson Electric Co. (c)2009 Emerson Electric Co.