Specification
50/80/120 Watts
•4000 VAC Isloation
•Low Leakage Current
•Single, Dual and Triple Outputs
•Worldwide Medical Approvals
•CCC Approved (PCM80)
•Both Line and Neutral Fused
•Non-standard Connectors Available
PCM Series
Input
Input Voltage • 90-264 VAC
Input Frequency • 47-63 Hz
Input Current • PCM50: 1.35 A max
PCM80: 1.30 A max
PCM120: 1.60 A max
Inrush Current • 30 A at 115 VAC or
60 A at 230 VAC at 25 ˚C cold start
Power Factor • PCM50: 0.55 typical
PCM80/120: 0.94 min
Earth Leakage Current • 180 µA max at 132 VAC/60Hz
300 µA max at 264 VAC/50Hz
Output
Output Voltage • See tables
Minimum Load • No minimum load required for single
output models. See tables for multi
output models
Start Up Delay • 2 s max
Hold Up Time • 10 ms min at low line
Line Regulation • ±1% max at full load
Load Regulation • ±5% max
Transient Response • 5% max. deviation, 500 µs recovery time
for a 50% load change
Ripple & Noise • 2% pk-pk, 20MHz BW
Overvoltage Protection • 112-132%
Overcurrent Protection • 110-200% with auto recovery
on all outputs
Temperature • ±0.05% /˚C
Coefficient
xppower.com
General
Efficiency • Single output models: 80% typical
Multi output models: 75% typical
230 VAC 100% load
Isolation • 4000 VAC Input to Output
1500 VAC Input to Ground
500 VAC Output to Ground
Switching Frequency • PCM50: 64 kHz typical
PCM80: 55 kHz typical
PCM120: 120 kHz typical (PWM),
70 kHz (PFC)
MTBF • 100 kHrs at full load and 25 ˚C ambient
per MIL-HDBK-217F
Environmental
Operating Temperature • PCM50: 0 ˚C to +70 ˚C, derate from 100%
load at 50 ˚C to 50% load at +70 ˚C
PCM80: 0 ˚C to +70 ˚C, derate from 100%
load at 40 ˚C to 50% load at +70 ˚C
PCM120: 0 ˚C to +60 ˚C, derate from 100%
load at +40 ˚C to 50% load at +60 ˚C
Storage Temperature • PCM50/80: -40 ˚C to +80 ˚C
PCM120: -20 ˚C to +85 ˚C
EMC & Safety
Emissions • EN55022, level B conducted & radiated
Harmonic Currents • EN61000-3-2, class A
Voltage Flicker • EN61000-3-3
ESD Immunity • IEC61000-4-2, level 3
Perf Criteria A
Radiated Immunity • IEC61000-4-3 10 V/m, level 3
Perf Criteria A
EFT/Burst • IEC61000-4-4, level 3, Perf Criteria A
Surge • IEC61000-4-5, level 3, Perf Criteria A
Conducted Immunity • IEC61000-4-6 10 V/m, level 3
Perf Criteria A
Power Frequency • IEC61000-4-8 10 A/m, Perf Criteria A
Magnetic Fields
Dips & Interruptions • IEC61000-4-11, 30% 500 ms,
60% 100 ms, 100 % 5000 ms
Perf Criteria A, B, B
Safety Approvals • PCM80/120: UL60601-1, CSA C22.2
No. 601.1 per cUL
PCM50: EN60601-1,
PCM80: CCC
AC-DC