+/- 0.4 % of full scale at 25 °C for thermocouple J - 200...0 °C
+/- 0.2 % of full scale at 25 °C for thermocouple J 0...1000 °C
+/- 0.4 % of full scale at 25 °C for thermocouple E - 200...0 °C
+/- 0.2 % of full scale at 25 °C for thermocouple E 0...800 °C
+/- 0.4 % of full scale at 25 °C for thermocouple T - 200...0 °C
+/- 0.2 % of full scale at 25 °C for thermocouple T 0...400 °C
+/- 0.4 % of full scale at 25 °C for thermocouple N - 200...0 °C
+/- 0.2 % of full scale at 25 °C for thermocouple N 0...1300 °C
Temperature drift +/- 0.01 %FS/°C
Repeat accuracy +/-0.5 %FS
Non-linearity +/- 0.2 %FS
Cross talk <= 1 LSB
[Us] rated supply voltage 24 V DC
Supply voltage limits 20.4...28.8 V
Type of cable Twisted shielded pairs cable 30 m for input circuit
Current consumption 30 mA at 24 V DC via external supply
45 mA at 5 V DC via bus connector
40 mA at 5 V DC via bus connector
Local signalling 1 LED green for PWR
Electrical connection 10 x 1.5 mm² removable screw terminal block with pitch 3.81 mm adjustment for inputs and supply
10 x 1.5 mm² removable screw terminal block with pitch 3.81 mm adjustment for inputs
Insulation 500 V AC between input and internal logic
1500 V AC between input and supply
Marking CE
Surge withstand 1 kV for power supply with common mode protection conforming to EN/IEC 61000-4-5
0.5 kV for power supply with differential mode protection conforming to EN/IEC 61000-4-5
1 kV for input with common mode protection conforming to EN/IEC 61000-4-5
Mounting support Top hat type TH35-15 rail conforming to IEC 60715
Top hat type TH35-7.5 rail conforming to IEC 60715
Plate or panel with fixing kit
Height 90 mm
Depth 70 mm
Width 23.6 mm
Product weight 0.11 kg
Environment
Standards EN/IEC 61131-2
EN/IEC 61010-2-201
Resistance to electrostatic discharge 4 kV on contact conforming to EN/IEC 61000-4-2
8 kV in air conforming to EN/IEC 61000-4-2
Resistance to electromagnetic fields 10 V/m at 80 MHz...1 GHz conforming to EN/IEC 61000-4-3
3 V/m at 1.4 GHz...2 GHz conforming to EN/IEC 61000-4-3
1 V/m at 2 GHz...3 GHz conforming to EN/IEC 61000-4-3
Resistance to magnetic fields 30 A/m conforming to EN/IEC 61000-4-8
Resistance to fast transients 1 kV I/O conforming to EN/IEC 61000-4-4
Resistance to conducted disturbances 10 V at 0.15...80 MHz conforming to EN/IEC 61000-4-6
3 V at spot frequency (2, 3, 4, 6.2, 8.2, 12.6, 16.5, 18.8, 22, 25 MHz) conforming to Marine
specification (LR, ABS, DNV, GL)
Electromagnetic emission Radiated emissions, test level: 40 dBμV/m QP class A (10 m at 30...230 MHz) conforming to EN/IEC
55011
Radiated emissions, test level: 47 dBμV/m QP class A (10 m at 230...1000 MHz) conforming to EN/
IEC 55011
Immunity to microbreaks 10 ms
Ambient air temperature for operation -10...55 °C (horizontal installation)
-10...35 °C (vertical installation)
Ambient air temperature for storage -25...70 °C
Relative humidity 10...95 % without condensation in operation
10...95 % without condensation in storage
IP degree of protection IP20
Pollution degree 2
2