Applications
Medical and dental equipment
Test and measurement
Chemical analysis equipment
Drug infusion equipment
Peripherals
Description
Single-Output Module Selection
Module No. of Slots Nominal Voltage Range Imax
Module 1 1 5V 3 to 5.6V 30A
Module 2 1 12V 5 to 13V 20A
Module 3 1 18V 8 to 20V 15A
Module 4 1 24V 12 to 28V 12A
Module 70 2 5V 1.45 to 5.6V 80A
Dual-Output Module Selection
Module No. of Slots Nominal Voltage Range Imax
Module 5 1 24V 10 to 28V 3A
24V 10 to 28V 3A
Module 6 1 5V 3 to 5.6V 10A
24V 10 to 28V 3A
ESM Series, AC-DC Medical Power Supply Data Sheet
Page 1 of 5
REV. MAY 24, 2004
Features
EN60601-1, UL2601-1, IEC601-1 approved
12mm creepage
Low leakage current <300 µA
Stackable for up to 2000W output power
1 to 12 isolated outputs with full user
configurability
Isolated bias supply voltage 5V @ 50mA
Class B conducted emissions (ESM6)
400, 600, and 1000 watts of output power
Series and parallel capability
EN61000-3-2 compliant
Individual control signals on each module
2-year warranty
The ESM4 and ESM6 series of modular AC/DC power supplies provide quick-turn-around, cost-effective power
solutions for medical applications from 400 to 1000 watts. Low leakage currents, under 300 µA (consult factory for
leakage currents under 150 µA), combined with conformance to IEC601-1 isolation and sp acing requirements make
these products ideal power sources for applications such as blood analyzers, DNA sequencers, and MRI and CAT
scanners.
The ESM4 Series is available in 400 and 600 watt configurations, both providing up to eight outputs from 2.56" x 5"
x 10.63" (65 x 127 x 270mm) chassis. The ESM6 Series is available in 600 and 1000 watt versions, both providing
up to twelve outputs from 2.56" x 7.36" x 10.63" (65 x 187 x 270mm) packages. Five single-output and two dual-
output modules can be configured in series or parallel to provide outputs from 1.45 to 58VDC.
Specification All specifications are typical at nominal input, full load at 25°C unless otherwise stated.
Output Specifications
Maximum power Input module B (for 4 slot only) 400W
Input module C (for 4 & 6 slot) 600W
Input module D (for 6 slot only) 1000W
(1)
Output adjustment (Note 2) Multi-turn potentiometer
Line regulation ±0.1%
Load regulation 50% load change ±0.2%
Cross regulation ±0.2% typ.
Transient response (Note 3) <10%, <0.5ms
Temperature coefficient ±0.02%/°C
Ripple and noise (Note 4) 1.0% or
100mV pk-pk
Overvoltage protection Standard on all outputs
Overcurrent protection (Note 5) Individual current limit
Thermal protection Standard
Mains failure signal Option 06 or 07 5ms warning
Output isolation (Note 6) Each single and dual
output fully floating
Margin See application note for individual
module margin capabilities
Minimum load (Note 7) Zero
Turn-on delay 500ms
Remote sense Single output modules only 0.5V drop
Input Specifications
Input voltage range Universal input 88 to 264VAC
125 to 370VDC
Input frequency range (Note 8) 47Hz to 63Hz
Inrush current 230VAC @ 25°C 50A max.
Harmonic distortion (Power factor) EN61000-3-2
NOTES
1 1000W peak power for 10s at low line. 800W average power for input
voltage less than 180VAC.
2 Outputs are user adjustable or factory set to your requested voltage.
3 25% to 75% load change.
4 Whichever is greater. 20MHz bandwidth. (See application note for
specification below 0°C).
5 Straight line on all outputs. On Module 70 current limit adjustable from
50% to 110%. Optional foldback on Module 70 or contact factory for
details see application note.
EMC Characteristics
Emissions: EN55022, FCC
Conducted ESM4B & ESM4C Level A
ESM6C & ESM6D Level B
Immunity:
Electrostatic discharge EN61000-4-2 Level 4
Radiated
RFI EN61000-4-3 Level 3
Fast transients - burst EN61000-4-4 Level 3
Input line surges EN61000-4-5 Class 3
Conducted RFI EN61000-4-6 Level 3
Voltage dips EN61000-4-11 Compliant
General Specifications
Hold-up time (Note 9) 20ms typ after loss of AC power
Efficiency 82% typ.
Isolation voltage Input/output 4000VAC
Input/chassis 1500VAC
Switching frequency 200kHz
Approvals and (Note 11) IEC60950, UL1950
standards CSA22.2 No. 950
Leakage current <300µA, 250VAC, 60Hz
Weight ESM4B, ESM4C 2.5kg
ESM6C, ESM6D 3.5kg
Size See mechanical specifications
MTBF See application note 400,000 hours
Environmental Specifications
Operating temperature See application note -20°C to +50°C
(See derating curve) Derate 2.5% per °C
up to +70°C
Storage temperature -40°C to +85°C
Relative humidity Non-condensing 5% to 95% RH
Shock 3000 bumps, 10G (16ms) half sine
Vibration 10-200Hz, 1.5G
6 100V isolation between each output and 500V to chassis.
7 All outputs except Module 70, which has 5.0% minimum load for full
specification.
8 Contact factory for 400Hz operation.
9 For nominal output voltages and full load.
10 The specifications contained in this data sheet are believed to be correct at
time of publication. Specifications are subject to change without notice.
11 This product is not intended for use as a stand alone unit and must be
installed by authorized personnel in order to maintain approvals.
Principles of Operation
ESM Series, AC-DC Medical Power Supply Data Sheet
Page 2 of 5
REV. MAY 24, 2004
7.87 (200.00)
1.63 (41.5)
TOP VIEW
SLOT A
SLOT B
SLOT C
SLOT D
0.39 (10.00)
1.38
(35.00)
1.38
(35.00)
1.36 (34.70)
2.36
(60.00)
1.27 (32.30)
5.00
(127.00)
7.87 (200.00)
10.63 (270.00)
1.38
(35.00)
1.38
(35.00)
1.28 (32.50)
2.56
(65.00)
AIRFLOW
SIDE VIEW
J1
J2
J4 J3
Third angle projection
All Dimensions Inches (mm)
Mounting Holes [M4] on Base and Side
Max screw penetration is 4mm
E
S
M4
E
S
M6
7.87 (200.00)
TOP VIEW
SLOT A
SLOT B
SLOT C
SLOT D
SLOT E
SLOT F
1.38
(35.00)
1.38
(35.00)
1.36 (34.70)
4.72
(120.00)
1.27 (32.30)
7.36
(187.00)
J2
J1
0.39 (10.00) 1.63 (41.5)
J4 J3
Mechanical Drawing Connectors:
J1 Line Input Connector
J2 Options
See application note for Pinout
J3 Output Signals
See application note for Pinout
J4 Output Connector
Accessories:
Parallel Link
Series Link
“U” Link
Mating Connector for options + Module 1-6 signals
Mating Connector for Module 70 signals
Typical Emissions to EN55022 Level B
ESM Series, AC-DC Medical Power Supply Data Sheet
Page 3 of 5
REV. MAY 24, 2004
ESM Series, AC-DC Medical Power Supply Data Sheet
Page 4 of 5
REV. MAY 24, 2004
Use 00 if no options
are desired.
Use 0 for unfilled slots
E S M 4 C 1 2 2 5 0 7
Single-Output Module Selection
Module No. of Slots Nominal Voltage Range Imax
Module 1 1 5V 3 to 5.6V 30A
Module 2 1 12V 5 to 13V 20A
Module 3 1 18V 8 to 20V 15A
Module 4 1 24V 12 to 28V 12A
Module 70 2 5V 1.45 to 5.6V 80A
Dual-Output Module Selection
Module No. of Slots Nominal Voltage Range Imax
Module 5 1 24V 10 to 28V 3A
24V 10 to 28V 3A
Module 6 1 5V 3 to 5.6V 10A
24V 10 to 28V 3A ESP/ESM Standard Options
06 Mains Power Fail + Global Enable + Bias Supply Voltage
07 Mains Power Fail + Global Inhibit + Bias Supply Voltage
Output Signals
Output control signals are available on all output modules.
(see application note)
Modules 1 to 6 Module 70 Additional Features
• Power good signal Adjustable Current Limit
Output inhibit signal Foldback or Straight Line Current
Limiting
Remote adjust (margin) Bias Voltage
Selectable Output Inhibit or
Enable
Dual output modules:
Output signals available on first [top] output only.
How to Order ESM4
(Available in 400/600 Watt Versions)
Note: Calculate power
requirements by summing
output powers calculated at
application output voltages.
Note: Calculate power
requirements by summing
output powers calculated at
application output voltages.
For ESM6D:
Limit total power from slots
A-C and D-E to 550W each.
How to Order ESM6
(Available in 600/1000 Watt Versions)
Series
Power B = 400W
C = 600W
C = 600W
D = 1000W
Slot A
Slot B Slot C
Slot D
Number of slots Option code
= standard configuration
C = special configuration
Use 00 if no options
are desired.
E S M 6 D 1 2 2 4 0 7
Series
Power
Slot A
Slot B Slot C
Slot D
Slot E
Slot F
Number of slots
Option code
= standard configuration
C = special configuration
}
}
Specification of power supply detailed above:
• 4-slot series
• Maximum output power: 600W
• 5V @ 30A; 12V @ 20A; 24V @ 3A; 24V @ 3A
• Mains Power Fail signal + Logic Inhibit + Bias Supply Voltage
Specification of power supply detailed above:
• 6-slot series
• Maximum output power: 1000W
• 5V @ 30A; 12V @ 20A; 12V @ 20A; 24V @ 12A; 24V @ 12A; 24V @ 3A; 24V @ 3A
• Mains Power Fail signal + Logic Inhibit + Bias Supply Voltage
Production Configuration:
Units are shipped with nominal output voltages unless special con-
figuration is specified. Power-One can configure to your exact
requirements through use of appropriate series and parallel busbars,
and voltage adjustment to specific set points.
4 5
Use 0 for unfilled slots
ESM Series, AC-DC Medical Power Supply Data Sheet
Page 5 of 5
REV. MAY 24, 2004
ESM Flexibility
Notes:
Maximum current = (I1+ I2) x .9
Use two parallel links
Notes:
Maximum voltage to chassis is 500V
Use series link
Reverse bias diodes may be required for certain
applications, eg. large capacitive loads
Notes:
Where the sensing point is remote from the
output of the power supply, to avoid spurious
noise pick-up it may be necessary to:
1 Use twisted pair sense wires.
2 Use R C as shown (R1= 100W) (R2= 10W)
(C = 22µF).
Notes:
See application note for full details.
NUCLEAR AND MEDICAL APPLICATIONS - Power-One products are not authorized for use as critical components in life support systems,
equipment used in hazardous environments, or nuclear control systems without the express written consent of the respective divisional president of
Power-One, Inc.
TECHNICAL REVISIONS - The appearance of products, including safety agency certifications pictured on labels, may change depending on the date
manufactured. Specifications are subject to change without notice.