SMU INSTRUMENTS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
2450
SourceMeter
®
SMU Instrument
Capabilities of analyzers,
curve tracers, and I-V systems
at a fraction of their cost.
Five-inch, high resolution
capacitive touchscreen GUI
0.012% basic measure accuracy
with 6½-digit resolution
Enhanced sensitivity with
new 20mV and 10nA source/
measure ranges
Source and sink (4-quadrant)
operation
Four “Quickset” modes for fast
setup and measurements
Built-in, context-sensitive front
panel help
Front panel input banana
jacks; rear panel input triaxial
connections
2450 SCPI and TSP® scripting
programming modes
Model 2400 SCPI-compatible
programming mode
Front panel USB memory
port for data/programming/
configuration I/O
The Model 2450 is Keithley’s next-generation SourceMeter source measure
unit (SMU) Instrument that truly brings Ohm’s law (current, voltage, and
resistance) testing right to your fingertips. Its innovative graphical user
interface (GUI) and advanced, capacitive touchscreen technology allow
intuitive usage and minimize the learning curve to enable engineers and
scientists to learn faster, work smarter, and invent easier. The 2450 is the
SMU for everyone: a versatile instrument, particularly well-suited for char-
acterizing modern scaled semiconductors, nano-scale devices and materi-
als, organic semiconductors, printed electronics, and other small-geometry
and low-power devices. All this combined with Keithley SMU precision and
accuracy allow users to Touch, Test, InventTM with the new favorite go-to
instrument in the lab for years to come.
Learn Faster, Work Smarter, Invent Easier
Unlike conventional instruments with dedicated pushbutton technology and small, obscure, limit-
ed-character displays, the 2450 features a five-inch, full-color, high resolution touchscreen that facil-
itates ease of use, learning, and optimizes overall speed and productivity. A simple icon-based menu
structure reduces configuration steps by as much as 50 percent and eliminates the cumbersome
multi-layer menu structures typically used on soft-key instruments. Built-in, context-sensitive help
enables intuitive operation and minimizes the need to review a separate manual. These capabilities
combined with its application versatility make the 2450 the SMU instrument inherently easy to
use for basic and advanced measurement applications, regardless of your experience level with
SMU instruments.
2450 main home screen. View of 2450 menu.
Fourth-Generation,
All-in-One SMU Instrument
The 2450 is the fourth-generation member of
Keithley’s award-winning SourceMeter family
of SMU instruments and leverages the proven
capabilities of the Model 2400 SourceMeter SMU
Instrument. It offers a highly flexible, four-quad-
rant voltage and current source/load coupled with
precision voltage and current meters. This all-in-
one instrument can be used as a:
Precision power supply with V and I readback
True current source
Digital multimeter (DCV, DCI, ohms, and
power with 6½-digit resolution).
Precision electronic load
Trigger controller
+20V-20V-200V +200V
+100mA
+1A
-100mA
-1A
Quad. II Quad. I
Quad. III Quad. IV
2450 power envelope.
Model 2450 SourceMeter
®
SMU Instrument
SMU INSTRUMENTS
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Confidence A Tektronix Company
Online
HELP key
USB 2.0
memory I/O
Front/rear
input selector
Rotary
navigation/
control knob
5˝ color graphical touchscreen display
2450
SourceMeter
®
SMU Instrument
Ordering Information
2450 200V, 1A, 20W
SourceMeter Instrument
2450-NFP 200V, 1A, 20W
SourceMeter Instrument,
with No Front Panel
2450-RACK
200V, 1A, 20W
SourceMeter Instrument,
without Handle
2450-NFP-RACK
200V, 1A, 20W
SourceMeter Instrument,
with No Front Panel
and No Handle
Accessories Supplied
8608 High Performance
Test Leads
USB-B-1 USB Cable, Type A to
Type B, 1m (3.3 ft)
CS-1616-3 Safety Interlock
Mating Connector
CA-180-3A TSP-Link/Ethernet Cable
Documentation CD
2450 QuickStart Guide
Test Script Builder Software
(supplied on CD)
KickStart Startup Software
(supplied on CD)
LabVIEW and IVI Drivers
available at www.keithley.com
Model 2400 Model 2450
V-Ranges: 200mV – 200V V-Ranges: 20mV – 200V
I-Ranges: 1μA – 1A I-Ranges: 10nA – 1A
0.012% Basic Accuracy 0.012% Basic Accuracy
Wideband Noise: 4mVrms Typ.Wideband Noise: 2mVrms Typ.
Sweep Types:
Linear, Log, Custom,
Source-Memory
Sweep Types:
Linear, Log, Dual Linear, Dual Log, Custom,
Source-Memory (SCPI 2400 Mode)
5000 Point Reading Buffer >250,000 Point Reading Buffer
>2000 Readings/Sec. >3000 Readings/Sec.
SCPI Programming SCPI (2400 + 2450) + TSP Programming
GPIB GPIB, USB, Ethernet (LXI)
Front/Rear Banana Jacks Front: Banana Jacks, Rear: Triax
Comparison of Model 2400 vs Model 2450.
Ease of Use Beyond the Touchscreen
In addition to its five-inch, color touchscreen, the 2450 front panel has many features that
supplement its speed, user-friendliness, and learnability, including a USB 2.0 memory I/O port, a
HELP key, a rotary navigation/control knob, a front/rear input selector button, and banana jacks for
basic bench applications. The USB 2.0 memory port supports easy data storing, saving instrument
configurations, loading test scripts, and system upgrades. Plus, all front panel buttons are backlit to
enhance visibility in low-light environments.
Model 2450 front panel with high resolution, capacitive touchscreen.
Four “Quickset” modes simplify user setup. With one touch, the instrument can be quickly configured
for various operating modes without the need to configure the instrument indirectly for this operation.
Comprehensive Built-in Connectivity
Rear panel access to rear-input triax connectors,
remote control interfaces (GPIB, USB 2.0, and
LXI/Ethernet), D-sub 9-pin digital I/O port (for
internal/external trigger signals and handler
control), instrument interlock control, and
TSP-Link® jacks enables easy configuration
of multiple instrument test solutions and
eliminates the need to invest in additional
adapter accessories.
Quickset modes enable fast setup and time
to measurements.
Model 2450 SourceMeter
®
SMU Instrument
SMU INSTRUMENTS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
2450 SourceMeter
®
SMU Instrument
Ethernet
Triax inputs
Digital I/O TSP-Link GPIB
Interlock
USB
Rear panel connections are optimized for signal integrity.
Convert Raw Data to Information
The 2450 provides a full plotting and sheet view to display sweeps,
measurement data, and charting right on the screen. It also supports
exporting to a spreadsheet for further analysis, dramatically improving
productivity for research, bench-top testing, device qualification,
and debugging.
Full data display, charting, and export to a spreadsheet lets you
convert raw data to useful information.
TYPICAL APPLICATIONS
Ideal for current/voltage characterization and
functional test of a wide range of todays modern
electronics and devices, including:
Nanomaterials and Devices
Graphene
Carbon nanotubes
Nanowires
Low power nanostructures
Semiconductor Structures
Wafers
Thin films
Organic Materials and Devices
E-inks
Printable electronics
Energy Efficiency and Lighting
LEDs/AMOLEDs
Photovoltaics/Solar Cells
Batteries
Discrete and Passive Components
Two-leaded: Resistors, diodes,
zener diodes, LEDs, disk drive
heads, sensors
Three-leaded: Small signal
bipolar junction transistors
(BJTs), field effect transistors
(FETs), and more
Material Characterization
Resistivity
Hall Effect
TriggerFlow™ Building Blocks for
Instrument Control and Execution
The 2450 incorporates Keithley’s new TriggerFlow triggering
system that allows user control of instrument execution. Similar to
developing a flow chart, TriggerFlow diagrams are created using
four fundamental building blocks:
Wait – Waits for an event to occur before the flow continues
Branch – Branches when a condition has been satisfied
Action – Initiates an action in the instrument, for example,
measure, source, delay, set digital I/O, etc.
Notify – Notifies other equipment that an event has occurred
Side text
Model 2450 SourceMeter
®
SMU Instrument
SMU INSTRUMENTS
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Confidence A Tektronix Company
2450 SourceMeter
®
SMU Instrument
TriggerFlow building blocks let users create very simple to very
complex triggering models.
A TriggerFlow model using a combination of these building blocks can
be created from the front panel or by sending remote commands. With
the TriggerFlow system, users can build triggering models from very sim-
ple to complex with up to 255 block levels. The 2450 also includes basic
triggering functions, including immediate, timer, and manual triggering.
Unmatched System Integration and
Programming Flexibility
When the 2450 is integrated as part of a multi-channel I-V test
system, the Test Script Processor (TSP®) embedded scripting capability
allows test scripts to be run by the instrument, enabling the user to
create powerful measurement applications with significantly reduced
development times. TSP technology also offers channel expansion
without a mainframe. Keithleys TSP-Link® channel expansion bus,
which uses a 100 Base T Ethernet cable, connects multiple 2450
instruments and other TSP instruments such as Keithley’s Series 2600B
SourceMeter SMU instruments and Series 3700A Switch/Multimeter
systems in a master-slave configuration that behaves as one integrated
system. The TSP-Link expansion bus supports up to 32 units per GPIB
or IP address, making it easy to scale a system to fit an application’s
particular requirements.
The 2450 also includes a SCPI programming mode that optimizes the
instrument’s new features, as well as a SCPI 2400 mode that provides
backwards compatibility with the existing Model 2400 SourceMeter
instrument. Not only does this preserve your 2400 investment, but it
also eliminates re-work normally associated with upgrading to a new
instrument with new capabilities.
Parallel Test Capability
With the TSP technology in the 2450, multiple devices can be
tested in parallel to meet the needs of device research, advanced
semiconductor lab applications, and even high throughput production test.
This parallel testing capability enables each instrument in the system to
run its own complete test sequence, creating a fully multi-threaded test
environment. The number of tests that can be run in parallel on a 2450
can be as high as the number of instruments in the system.
Free Instrument Control Start-up
Software and Web Interface
KickStart, Keithley’s new instrument control non-programming start-up
software, lets users start taking measurements in minutes. In most cases,
users merely need to make quick measurements, graph the data, and
store the data to disk to perform analysis in software environments
such as Excel.
KickStart offers the following functionality:
Instrument configuration control to perform I-V characterization
Native X-Y graphing, panning, and zooming
Spreadsheet/tabular viewing of data
Saving and exporting data for further analysis
Saving of test setups
Screenshot capturing of graph
Annotation of tests
Command line dialog for sending and receiving data
HTML help
GPIB, USB 2.0, Ethernet compliant
With KickStart start-up software, users are ready to take
measurements in minutes.
Simplified Programming
with Ready-to-Use Instrument Drivers
For users who want to create their own customized application software,
native National Instruments LabVIEW® drivers, IVI-C, and IVI-COM
drivers are available at www.keithley.com.
Model 2450 SourceMeter
®
SMU Instrument
Side text
SMU INSTRUMENTS
A Greater Measure of Confidence
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Tektronix Company
2450 SourceMeter
®
SMU Instrument
Voltage Specifications1,2
Source Measure3
Range Resolution
Accuracy (23° ± 5°C)
1 Year
±(% setting + volts)
Noise
(RMS)
(<10Hz) Resolution
Input
Resistance
Accuracy (23° ± 5°C)
1 Year
±(% rdg. + volts)
20.00000 mV 500 nV 0.100% + 200 μV 1 μV 10 nV >10 GW0.100% + 150 μV
200.0000 mV 5 μV 0.015% + 200 μV 1 μV 100 nV >10 GW0.012% + 200 μV
2.000000 V 50 μV 0.020% + 300 μV 10 μV 1 μV >10 GW0.012% + 300 μV
20.00000 V 500 μV 0.015% + 2.4 mV 100 μV 10 μV >10 GW0.015% + 1 mV
200.0000 V 5 mV 0.015% + 24 mV 1 mV 100 μV >10 GW0.015% + 10 mV
Current Specifications1,2
Source Measure3
Range Resolution
Accuracy (23° ±C)4
1 Year
±(% setting + amps)
Noise
(RMS)
(<10Hz) Resolution
Voltage
Burden
Accuracy (23° ±C)
1 Year
±(% rdg. + amps)
10.00000 nA5500 fA 0.100% + 100 pA 500 fA 10 fA <100 μV 0.100% + 50 pA
100.0000 nA55 pA 0.060% + 150 pA 500 fA 100 fA <100 μV 0.060% + 100 pA
1.000000 μA 50 pA 0.025% + 400 pA 5 pA 1 pA <100 μV 0.025% + 300 pA
10.00000 μA 500 pA 0.025% + 1.5 nA 40 pA 10 pA <100 μV 0.025% + 700 pA
100.0000 μA 5 nA 0.020% + 15 nA 400 pA 100 pA <100 μV 0.020% + 6 nA
1.000000 mA 50 nA 0.020% + 150 nA 5 nA 1 nA <100 μV 0.020% + 60 nA
10.00000 mA 500 nA 0.020% + 1.5 μA 40 nA 10 nA <100 μV 0.020% + 600 nA
100.0000 mA 5 μA 0.025% + 15 μA 100 nA 100 nA <100 μV 0.025% + 6 μA
1.000000 A 50 μA 0.067% + 900 μA 3 μA 1 μA <100 μV 0.030% + 500 μA
TEMPERATURE COEFFICIENT (0°–18°C and 28°–50°C): ±(0.15 × accuracy specification)/°C.
1. Speed = 1 PLC.
2. All specifications are guaranteed with output ON.
3. Accuracies apply to 2- and 4-wire mode when properly zeroed.
4. For sink mode, 1μA to 100mA range accuracy is ±(0.15% + offset × 4). For 1A range, accuracy is ±(1.5% + offset × 8).
5. Rear panel triax connections only.
Resistance Measurement Accuracy (Local or Remote Sense)2,3
Range
Default
Resolution6
Default
Test Current
Normal Accuracy
(23°C ±5°C)
1 Year, ±(% rdg. + ohms)
Enhanced Accuracy7
(23°C ±5°C)
1 Year, ±(% rdg. + ohms)
<2.000000 W81 μWUser defined Source IACC + Meas. VACC Meas. IACC + Meas. VACC
20.00000 W10 μW100 mA 0.098% + 0.003 W0.073% + 0.001 W
200.0000 W100 μW10 mA 0.077% + 0.03 W0.053% + 0.01 W
2.000000 kW1 mW1 mA 0.066% + 0.3 W0.045% + 0.1 W
20.00000 kW10 mW100 μA 0.063% + 3 W0.043% + 1 W
200.0000 kW100 mW10 μA 0.065% + 30 W0.046% + 10 W
2.000000 MW1 W1 μA 0.110% + 300 W0.049% + 100 W
20.00000 MW10 W1 μA 0.110% + 1 kW0.052% + 500 W
200.0000 MW100 W100 nA 0.655% + 10 kW0.349% + 5k W
>200.0000 MW8User defined Source IACC + Meas. VACC Meas. IACC + Meas. VACC
TEMPERATURE COEFFICIENT (0°–18°C and 28°–50°C): ±(0.15 × accuracy specification)/°C.
SOURCE CURRENT, MEASURE RESISTANCE MODE:
Total uncertainty = Isource accuracy + Vmeasure accuracy (4-wire remote sense).
SOURCE VOLTAGE, MEASURE RESISTANCE MODE:
Total uncertainty = Vsource accuracy + Imeasure accuracy (4-wire remote sense).
GUARD OUTPUT IMPEDANCE: 0.5W (DC) in ohms mode.
6. 6.5 digit measure resolution
7. Source readback enabled. Offset compensation ON.
8. Source Current, Measure Resistance or Source Voltage, Measure Resistance only.
ACCESSORIES AVAILABLE
TEST LEADS AND PROBES
1754 2-wire Universal 10-Piece Test Lead Kit
5804 Kelvin (4-Wire) Universal 10-Piece Test Lead Kit
5805 Kelvin (4-Wire) Spring-Loaded Probes
5806 Kelvin Clip Lead Set
5808 Low Cost Single-pin Kelvin Probe Set
5809 Low Cost Kelvin Clip Lead Set
8605 High Performance Modular Test Leads
8606 High Performance Modular Probe Kit
8608 High Performance Clip Lead Set
CABLES, CONNECTORS, ADAPTERS
237-ALG-2 3-slot Male Triax Connector to 3 Alligator Clips
237-BAN-3A Triax to Banana Plug
2450-TRX-BAN Triax to Banana Adapter. Converts the 4 Triax
adapters on the rear panel to 5 banana jacks
7078-TRX-* 3-slot, Low Noise Triax Cable
7078-TRX-GND 3-slot Male Triax To BNC Adapter
(guard removed)
8607 2-wire, 1000V Banana Cables, 1m (3.3 ft)
CA-18-1 Shielded Dual Banana Cable, 1.2m (4 ft)
CAP-31 Protective Shield/Cap for 3-lug Triax Connectors
CS-1546 Triax 3-lug Special Shorting Plug. Shorts center
pin to outer shield
CS-1616-3 Safety Interlock Mating Connector
COMMUNICATION INTERFACES & CABLES
KPCI-488LPA IEEE-488 Interface for PCI Bus
KUSB-488B IEEE-488 USB-to-GPIB Interface Adapter
7007-1 Shielded GPIB Cable, 1m (3.3 ft)
7007-2 Shielded GPIB Cable, 1m (6.6 ft)
CA-180-3A CAT5 Crossover Cable for TSP-Link/Ethernet
USB-B-1 USB Cable, Type A to Type B, 1m (3.3 ft)
TRIGGERING AND CONTROL
2450-TLINK DB-9 to Trigger Link Connector Adapter.
8501-1 Trigger Link Cable, DIN-to-DIN, 1m (3.3 ft)
8501-2 Trigger Link Cable, DIN-to-DIN, 2m (6.6 ft)
RACK MOUNT KITS
4299-8 Single Fixed Rack Mount Kit
4299-9 Dual Fixed Rack Mount Kit
4299-10 Dual Fixed Rack Mount Kit. Mount one 2450
and one Series 26xxB
4299-11 Dual Fixed Rack Mount Kit. Mount one 2450
and one Series 2400, Series 2000, etc.
2450-BenchKit Ears and Handle for 2450-NFP-RACK and
2450-RACK models
TEST FIXTURES
8101-PIV DC Test Fixture
SERVICES AVAILABLE
2450-3Y-EW 1 Year Factory Warranty extended to 3 years
from date of shipment
2450-5Y-EW 1 Year Factory Warranty extended to 5 years
from date of shipment
C/2450-3Y-17025 KeithleyCare® 3 Year ISO 17025 Calibration Plan
C/2450-3Y-DATA KeithleyCare 3 Year Calibration w/Data Plan
C/2450-3Y-STD KeithleyCare 3 Year Std. Calibration Plan
C/2450-5Y-17025 KeithleyCare 5 Year ISO 17025 Calibration Plan
C/2450-5Y-DATA KeithleyCare 5 Year Calibration w/Data Plan
C/2450-5Y-STD KeithleyCare 5 Year Std. Calibration Plan
Model 2450 specifications
Model 2450 specifications
SMU INSTRUMENTS
www.keithley.com
1.888.KEITHLEY (U.S. only)
A Greater Measure of Confidence A Tektronix Company
OPERATING CHARACTERISTICS
MAX. OUTPUT POWER: 20W, four-quadrant source or sink operation.
SOURCE LIMITS: Vsource: ±21V ( 1A range), ±210V ( 100mA range)
Isource: ±1.05A ( 20V range), ±105mA ( 200V range)
OVERRANGE: 105% of range, source and measure.
REGULATION: Voltage: Line: 0.01% of range. Load: 0.01% of range + 100μV.
Current: Line: 0.01% of range. Load: 0.01% of range + 100pA.
SOURCE LIMITS: Voltage Source Current Limit: Bipolar current limit set with single value. Min. 10% of range.
Current Source Voltage Limit: Bipolar voltage limit set with single value. Min. 10% of range.
V-LIMIT / I-LIMIT ACCURACY: Add 0.3% of setting and ±0.02% of reading to base specification.
OVERSHOOT: Voltage Source: <0.1% typical (full scale step, resistive load, 20V range, 10mA I-Limit.
Current Source: <0.1% typical (1mA step, RLoad = 10kW, 20V range)
RANGE CHANGE OVERSHOOT: Overshoot into a fully resistive 100kW load, 10Hz to 20MHz BW, adjacent ranges: 250mV typical
OUTPUT SETTLING TIME: Time required to reach 0.1% of final value, 20V range, 100mA I-Limit: <20s typical.
MAXIMUM SLEW RATE: 0.2V/μs, 200V range, 100mA limit into a 2kW load (typical)
OVER VOLTAGE PROTECTION: User selectable values, 5% tolerance. Factory default = none.
VOLTAGE SOURCE NOISE: 10Hz–1MHz (RMS): 2mV typical into a resistive load.
COMMON MODE VOLTAGE: 250V DC.
COMMON MODE ISOLATION: >1GW, <1000pF.
NOISE REJECTION (TYPICAL):
LOAD IMPEDANCE: 20nF typical (standard). Stable into 50μF typical (High-C mode).
High-C mode valid for 100μA ranges, 200mV ranges.
MAX. VOLTAGE DROP BETWEEN FORCE and SENSE TERMINALS: 5V.
MAX. SENSE LEAD RESISTANCE: 1MW for rated accuracy.
SENSE INPUT IMPEDANCE: >10 GW.
GUARD OFFSET VOLTAGE: <300μV, typical
System Measurement Speeds 9
READING RATES (READINGS/SECOND) TYPICAL FOR 60Hz (50Hz):
SCRIPT (TSP) Programmed
Measure Source-Measure Sweep
NPLC/Trigger Origin To Mem. To GPIB To USB To LAN To Mem. To GPIB To USB To LAN
0.01 / Internal 3130 (2800) 2830 (2570) 2825 (2600) 2790 (2530) 1710 (1620) 1620 (1540) 1630 (1540) 1620 (1540)
0.01 / External 2170 (2050) 2150 (2030) 2170 (2040) 2160 (1990) 1670 (1590) 1580 (1500) 1590 (1510) 1580 (1510)
0.1 / Internal 540 (460) 530 (450) 530 (450) 530 (450) 470 (410) 460 (400) 470 (400) 470 (400)
0.1 / External 500 (430) 490 (420) 500 (430) 500 (420) 470 (400) 460 (390) 460 (400) 460 (400)
1.00 / Internal 59 (49) 58 (49) 59 (49) 59 (49) 58 (48) 58 (48) 58 (48) 58 (48)
1.00 / External 58 (48) 57 (48) 58 (48) 58 (48) 57 (48) 57 (47) 57 (48) 57 (48)
SCPI Programmed 10
Measure Source-Measure Sweep
NPLC/Trigger Origin To Mem. To GPIB To USB To LAN To Mem. To GPIB To USB To LAN
0.01 / Internal 3130 (2800) 3060 (2760) 3000 (2790) 3010 (2710) 1710 (1630) 1610 (1600) 1440 (1380) 1690 (1590)
0.01 / External 2350 (2200) 2320 (2170) 2340 (2190) 2320 (2130) 1680 (1590) 1560 (1570) 1410 (1360) 1660 (1560)
0.1 / Internal 540 (460) 540 (450) 540 (460) 540 (450) 470 (410) 470 (410) 450 (390) 470 (410)
0.1 / External 510 (440) 510 (430) 510 (440) 510 (430) 470 (400) 470 (400) 450 (390) 470 (400)
1.00 / Internal 59 (49) 59 (49) 59 (49) 59 (49) 58 (48) 58 (48) 57 (48) 58 (48)
1.00 / External 58 (49) 58 (49) 58 (49) 58 (49) 58 (48) 58 (48) 57 (47) 58 (48)
9. Reading rates applicable for voltage or current measurements, autozero off, autorange off, filter off, binary reading format, and source readback off.
10. SCPI programming mode. Speeds do not apply to SCPI 2400 mode.
2450 SourceMeter
®
SMU Instrument
NPLC NMRR CMRR
0.01 60 dB
0.1 60 dB
160 dB 100 dB*
* Except lowest two current ranges ~90dB.
Model 2450 specifications
SMU INSTRUMENTS
2450 SourceMeter
®
SMU Instrument
GENERAL CHARACTERISTICS (default mode unless specified)
FACTORY DEFAULT STANDARD POWER-UP: SCPI MODE.
SOURCE OUTPUT MODES: Fixed DC Level, Memory/Configuration
List (mixed function), Stair (linear and log).
SOURCE MEMORY LIST: 100 points max. (SCPI 2400 Mode only).
MEMORY BUFFER: >250,000 readings. Includes selected measured value(s)
and time stamp.
REAL-TIME CLOCK: Lithium battery backup (3 yr. + battery life).
REMOTE INTERFACES:
GPIB: IEEE-488.1 compliant. Supports IEEE-488.2 common commands
and status model topology.
USB Device (rear panel, type B): 2.0 Full Speed USBTMC.
USB Host (front panel, type A): USB 2.0, support for flash drives, FAT32.
Ethernet: RJ-45 (10/100BT)
PROGRAMMABILITY: SCPI or TSP command sets.
TSP MODE: Embedded Test Script Processor (TSP) accessible from any host interface.
IP CONFIGURATION: Static or DHCP
EXPANSION INTERFACE: The TSP-Link expansion interface allows TSP enabled
instruments to trigger and communicate with each other.
LXI COMPLIANCE: 1.4 LXI Core 2011.
DISPLAY: 5 inch capacitive touch, color TFT WVGA (800x480) with LED backlight.
INPUT SIGNAL CONNECTIONS: Front: Banana. Rear: Triaxial (3-Lug)
INTERLOCK: Active High Input
DIGITAL I/O INTERFACE:
Lines: 6 Input/Output user defined for digital I/O or triggering
Connector: 9-pin female D
Input Signal Levels: 0.7 V (maximum logic low), 3.7 V (minimum logic high)
Input Voltage Limits: -0.25 V (Abs. minimum), +5.25 V (Abs. maximum)
Maximum Source Current: +2.0 mA @ >2.7 V (per pin)
Maximum Sink Current: -50 mA @ 0.7 V (per pin, solid-state fuse protected)
5 V Power Supply Pin: Limited to 500 mA @ >4V (solid-state fuse protected)
Handler: User definable Start of Test, End of Test, 4 category bits
COOLING: Forced air, variable speed.
OVER TEMPERATURE PROTECTION: Internally sensed temperature overload puts unit in
standby mode.
POWER SUPPLY: 100V to 240V RMS, 5060Hz (automatically detected at power up).
VA RATING: 190 volt-amps max.
ALTITUDE: Maximum 2000 meters above sea level.
EMC: Conforms to European Union EMC Directive.
SAFETY: NRTL listed to UL61010-1 and UL61010-2-30. Conforms with European Union Low
Voltage Directive.
VIBRATION: MIL-PRF-28800F Class 3 Random.
WARM-UP: 1 hour to rated accuracies.
DIMENSIONS: (With handle and bumpers): 106mm high × 255mm wide × 425mm deep
(4.18 in × 10.05 in × 16.75 in). (Without handle and bumpers): 88mm high × 213mm wide
× 403mm deep (3.46 in × 8.39 in × 15.87 in).
WEIGHT: With bumpers & handle: 4.04 kg (8.9 lbs.). Without bumpers & handle 3.58 kg (7.9 lbs.).
ENVIRONMENT: Operating: 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C.
Storage: –25°C to 65°C.
ACCESSORIES SUPPLIED: Test Leads, USB Cable, Ethernet/TSP Cable, Interlock Adapter, Power
Cord, Quick Start Guide, CD User’s Manual.
Model 2450 specifications
SMU INSTRUMENTS
2450 SourceMeter
®
SMU Instrument
Model 2450 SourceMeter
®
SMU Instrument
Specifications are subject to change without notice. All Keithley trademarks and trade names are the property of Keithley Instruments, Inc.
All other trademarks and trade names are the property of their respective companies.
KEITHLEY INSTRUMENTS, INC. 28775 AURORA RD. CLEVELAND, OH 44139-1891 440-248-0400 Fax: 440-248-6168 1-888-KEITHLEY www.keithley.com
A Greater Measure of Confidence
BENELUX
+31-40-267-5506
www.keithley.nl
BRAZIL
55-11-4058-0229
www.keithley.com
CHINA
86-10-8447-5556
www.keithley.com.cn
FRANCE
+33-01-69-86-83-60
www.keithley.fr
GERMANY
+49-89-84-93-07-40
www.keithley.de
INDIA
080-30792600
www.keithley.in
ITALY
+39-049-762-3950
www.keithley.it
JAPAN
81-120-441-046
www.keithley.jp
KOREA
82-2-6917-5000
www.keithley.co.kr
MALAYSIA
60-4-643-9679
www.keithley.com
MEXICO
52-55-5424-7907
www.keithley.com
RUSSIA
+7-495-664-7564
www.keithley.ru
SINGAPORE
01-800-8255-2835
www.keithley.com.sg
TAIWAN
886-3-572-9077
www.keithley.com.tw
UNITED KINGDOM
+44-1344-39-2450
www.keithley.co.ukw
For further information on how to purchase or to locate a sales partner please visit www.keithley.com/company/buy
© Copyright 2013 Keithley Instruments, Inc. Printed in the U.S.A No. 3219 07.26.13