Specification
(Typ.)
(Typ.)
(Typ.)
ACINPUTVOLTAGE
ACINPUTCURRENT
INRUSHCURRENT
SETUP,RISETIME
HOLD-UP TIME
LEAKAGECURRENT
DCOUTPUT ADJ.RANGE
OVERLOADPROTECTION
OVERVOLTAGEPROTECTION
POWERGOOD/FAIL SIGNAL (OPTIONAL)
WORKINGTEMP.
WORKINGHUMIDITY
STORAGETEMP.,HUMIDITY
90~264VAC,47~440Hz/127~370VDC.
Maximuminputcurrent2.9A at115VAC,60Hzor
1.7A at230VAC,60Hzwith100%outputload.
Inrushcurrentislessthan22A at115VACorlessthan
45A at230VACundercoldstartconditions.Limiting
providedbyinternalthermistors.
MPS-120: 800ms,20ms/230VACatfullload
2000ms,50ms/115VACatfullload
MPD,T,Q-120:500ms,20ms/230VACatfullload
1200ms,50ms/115VACatfullload
80ms/230VACatfullload
14ms/115VACatfullload
Leakagecurrentislessthan180 A at264VAC
DCoutputvoltage(orCh1ofmultipleoutputmodels)
canbeadjustedbetween-5%~+10%ratedoutput
voltagebypotentialmeter.
Fullyprotectedagainstshortcircuitandoutputoverload.
Thehiccuptypeprotection willbeactivatedat110~150%
(ForMPD,T,Q-120),120~160%(ForMPS-120)ratedload
andrecoversautomaticallyafterfaultconditionis removed.
Providedonoutputchannel1onlyat115%~135%
ratedoutputvoltage.(120%~140%forMPS-120-15/24/48).
Outputwillbeshutdownwhenthisprotectionisactivated.
*MPS-120-3.3doesnothavethisoptionalfunction.
Wholeseriescanoperatefrom-20~70 .Pleaserefer
tothederatingcurves.
20~90%RHnon-condensing.
-40~+85 ,10~90%RH
TTL logichighforpowergoodand TTL lowfor
powerfail.Whentheoutputvoltagereaches90%of
ratedvalue,a+5V TTL signalwillbesentoutwitha
10~500msdelay; Atleast1msbeforetheoutput
voltagegoesbelow90%oftheratedvalue,the TTL
signalwillbeturnedoff.
ComplianceLevel
EN55011,ClassB
EN55022,ClassB
EN61000-3-2
EN61000-3-3
ComplianceLevel
EN61000-4-2,
EN61000-4-2,
EN61000-4-3,
EN61000-4-4,
EN61000-4-5,
EN61000-4-6,
EN61000-4-8,
EN61000-4-11,
ENV50204,
Level3,8KV
Level2,4KV
Level2,3V/m
Level3,10V/m
Level2,1KV/5KHz
Level3,2KV/5KHz
Level4,2KV/Line-Line
4KV/Line-Earth
Level2,3Vrms/m
Level3,10Vrms/m
Level2,3A/m
Level3,10A/m
Compliance
Level2,3V/m,900MHz
Level3,10A/m,900MHz
1~4OutputMedical Type 120WMedical series
Universal ACinput/Fullrange
Lowleakagecurrent<180 A
Protections:Shortcircuit/Overload/Overvoltage
UL60601-1medicalsafetyapproved
Withpowergoodandfailsignaloutput(Optional)
100%fullloadburn-intest
Fixedswitchingfrequencyat45KHz
3yearswarranty
Features
TEMP.COEFFICIENT
VIBRATION
SAFETY STANDARDS
:
WITHSTANDVOLTAGE
ISOLATIONRESISTANCE
EMICOMPLIANCE
EMSCOMPLIANCE
MTBF
DIMENSION(L*W*H)
PACKING
2Gofacceleration,vibratingfrequencyadjustfrom10Hz~500Hzwithina
10-minutecycle.6testingcycles(60minutes)eachalongX, Y,Zaxes.
Medical UL60601-1, TUVEN60601-1 approved
Commercial: AlsodesignrefertoUL60950-1, TUVEN60950-1
4000VACbetweeninputandoutput
1500VACbetweeninputandF.G.
500VACbetweenoutputandF.G.
EMISpecifications
EMSSpecification
ESDair
ESDcontact
RFfieldsusceptibility
EFT(ElectricalFast Transient)/Burst
Lightning/Surge
ConductedRFsusceptibility
Magneticfieldimmunity
Voltagedip,interruption
Digitalphonecarrierimmunity
262,100hoursmin.atfullloadand25 ofambienttemperature,
calculatedperMIL-HDBK-217F.
177.8x107.95x35.5mmor7"x4.25"x1.4"
0.55Kg;24pcs/14.5Kg/0.99CUFT
,
0.04%/ onalloutputsatfullloadbetween0~50 ofambienttemperature.
>100MOhmsforI/P-O/P,I/P-FG,O/P-FGbyusing500VDCtestvoltage.
Conducted&Radiation
Harmonicdistortion
Voltageflicker
IEC60601-1
FileName:120WMEDICAL-SPEC2010-10-18
EN60601-1
IEC60601-1
http://store.iiic.cc/