cui.com
date 09/05/2017 │ page 2 of 7CUI Inc │ SERIES: VMS-550 │ DESCRIPTION: AC-DC POWER SUPPLY
INPUT
parameter conditions/description min typ max units
voltage 90 264 Vac
frequency 47 63 Hz
current at 115 Vac, full load
at 230 Vac, full load
6
3
A
A
inrush current at 230 Vac, cold start 45 A
leakage current at 115 Vac
at 230 Vac
0.2
0.4
mA
mA
touch current 0.1 mA
power factor at full load 0.95
no load power consumption 0.7 W
input fuse 8 A/250 V time delay fuse (included)
OUTPUT
parameter conditions/description min typ max units
initial set point accuracy ±1 %
line regulation ±0.5 %
load regulation from 100% to 10% load ±1 %
start-up delay time 2 s
rise time at 115/230 Vac 55 ms
hold-up time at 115/230 Vac, full load 16 ms
adjustability built in trim pot ±3 %
switching frequency 50 300 kHz
transient response
25% step load change, at 0.1 A/µS slew rate, 50%
duty cycle, 50/60 Hz, max excursion 4%, recovery
time 5 ms
temperature coefcient at 0~50°C ±0.05 %/°C
fan output112 Vdc / 500 mA
Notes: 1. Fan supply output voltage tolerance including set point accuracy, line and load regulation is ±10% and ripple and noise is less than 10%.
PROTECTIONS
parameter conditions/description min typ max units
over voltage protection hiccup, auto recovery 110 140 %
over current protection hiccup, auto recovery 110 %
short circuit protection hiccup, auto recovery
over temperature protection goes into hiccup mode when the temperature of
the PCB exceeds 110±10°C, auto recovery
SAFETY & COMPLIANCE
parameter conditions/description min typ max units
isolation voltage
input to output (2 x MOPP)
input to ground (1 x MOPP)
output to ground
4,200
1,500
1,500
Vac
Vac
Vac
safety approvals
IEC 60601-1: 2005 +CORR1:2006
+CORR2:2007 +AM1:2012, EN 60601-1:2006
+A11:2011 +A1:2013, ANSI/AAMI ES 60601-1
(2005+CI:09+A2:10), AMD1:2012 CAN/CSA-
C22.2 No 60601-1 (2008) 60601-1:14 ISO 14971,
2nd edition complies with LVD directive
safety class class I
Notes: 2. The power supply is considered a component which will be installed into nal equipment. The nal equipment still must be tested to meet the necessary EMC directives.