IMM01 Series DC-DC Converter 1 Watt * World Wide Medical Approvals * Single and Dual Outputs * 2 A Patient Leakage Current * SIP7 Package * -20 C to +100 C Operation * Full Load at 60 C Ambient * 1500 VAC Isolation, 1 MOPP * MTBF 1 MHrs * 3 Year Warranty Dimensions: IMM01: 0.76 x 0.36 x 0.44" (19.5 x 9.2 x 11.1 mm) Models & Ratings Input Voltage Output Voltage 4.5-9.0 V 9.0-18.0 V Output Current Input current No Load Full Load Maximum capacitive load Model Number 3V3 303 mA 75 mA 337 mA 2200 F IMM0105S3V3 5V 200 mA 75 mA 337 mA 2200 F IMM0105S05 12 V 83 mA 75 mA 327 mA 470 F IMM0105S12 15 V 67 mA 75 mA 327 mA 470 F IMM0105S15 3.3 V 150 mA 55 mA 337 mA 1000 F IMM0105D03 5 V 100 mA 55 mA 337 mA 1000 F IMM0105D05 12 V 42 mA 75 mA 327 mA 220 F IMM0105D12 15 V 33 mA 75 mA 327 mA 220 F IMM0105D15 3V3 303 mA 25 mA 163 mA 2200 F IMM0112S3V3 5V 200 mA 25 mA 163 mA 2200 F IMM0112S05 12 V 83 mA 25 mA 150 mA 470 F IMM0112S12 15 V 67 mA 25 mA 150 mA 470 F IMM0112S15 3.3 V 150 mA 55 mA 163 mA 1000 F IMM0112D03 5 V 100 mA 55 mA 163 mA 1000 F IMM0112D05 12 V 42 mA 30 mA 150 mA 220 F IMM0112D12 15 V 33 mA 30 mA 150 mA 220 F IMM0112D15 Notes 0 Input currents measured at low input voltage. Mechanical Details Pin Connections 0.76 (19.5) 0.02 (0.50) 0.36 (9.20) Printed Face 0.42 (10.6) 0.12 (3.0) 1 2 3 0.02 (0.50) 5 6 7 0.44 (11.1) Bottom View 0.36 (9.20) 0.13 (3.20) 0.08 (2.13) 0.6 (15.24) 1 Single -Vin +Vin Dual -Vin +Vin 3 Remote On/Off Remote On/Off 5 6 7 +Vout -Vout No Pin +Vout Common -Vout Notes 1. All dimensions are in inches (mm) 2. Weight: 0.008 lbs (3.6 g) approx. 3. Pin diameter: 0.020.002 (0.50.05) 4. Pin pitch tolerance: 0.014 (0.35) 5. Case tolerance: 0.02 (0.5) 0.10 (2.54) 0 0.02 (0.50) 0.01 (0.25) Pin 1 2 www.xppower.com IMM01 Series DC-DC Converter Input Characteristic Input Voltage Range Minimum Typical 4.5 9 Inrush Current Input Reflected Ripple Current Maximum Units Notes & Conditions 9 VDC 5 V nominal 12 V nominal 18 VDC 0.05 A2s 16 VDC for 100 ms 5 V models 25 VDC for 100 ms 12 V models Maximum Units Notes & Conditions 30 VDC See Models and Ratings table 2 A 30 mA pk-pk Input Surge Through 12 H inductor and 47 F capacitor Output Characteristic Minimum Output Voltage 3.3 Typical Patient Leakage Current 2 Initial Set Accuracy Minimum Load 3.3 V & 5.0 V at full load % 1 0 12.0 V & 15.0 V at full load % No minimum load required Line Regulation 0.5 % From minimum to maximum input Load Regulation 2.0 % From 10% to full load (2% from 0% to 10% load) Cross Regulation 5 % On dual output models, when one output is at 25% load and other is varied from 10% load to full load Ripple & Noise 1/2 % pk-pk For 12 V & 15 V/ 3.3 V & 5 V models. 20 MHz bandwidth. Measured using 0.1 F ceramic capacitor Short Circuit Protection Continuous fold-back mode, with auto recovery Maximum Capacitive Load See Models and Ratings table Temperature Coefficient Overload Protection Remote On/Off 190 250 0.03 %/ C 310 % Of nominal output current at nominal input voltage Output is on if remote on/off (pin 3) is open Output turns off if 2-4 mA is applied to remote on/off (pin 3) General Characteristic Efficiency Minimum Typical Maximum Units 64 78 81 % Isolation: Input to Output 1500 Switching Frequency 175 Isolation Resistance 109 Isolation Capacitance Power Density Mean Time Between Failure Typical value is for IMM0112S12 VAC At 250 VAC working voltage, 1 MOPP 1000 kHz May enter burst mode frequency of 12-28 kHz at light load 27 pF W/in3 8.31 1 Weight Notes & Conditions MHrs 0.008 (3.6) MIL-HDBK-217F, +25 C GB lb (g) Environmental Characteristic Maximum Units Operating Altitude Minimum 3048 M Transport Altitude 10,000 Operating Temperature -20 Storage Temperature -40 Case Temperature Humidity Typical +100 C +125 C +105 C 95 %RH Cooling Notes & Conditions Derate linearly from 100% load at +60 C to 0% at +100 C Non-condensing Natural convection www.xppower.com 2 IMM01 Series DC-DC Converter EMC: Emissions Phenomenon Standard Test Level Notes & Conditions Conducted EN55011 Class A See Application Note Radiated EN55011 Class A EMC: Immunity Standard Test Level Criteria Medical Device EMC Phenomenon IEC60601-1-2 Ed 4.0:2014 As below ESD Immunity EN61000-4-2 15 kV A Air Discharge A 80% mod, 80 MHz - 2.7 GHz plus discrete communcation proximity field frequencies Radiated Immunity EN61000-4-3 10 V/m Notes & Conditions EFT/Burst EN61000-4-4 2 kV A External input filter required, see applications note Surge EN61000-4-5 2 kV A External input filter required, see applications note Conducted Immunity EN61000-4-6 10 V rms A Magnetic Fields EN61000-4-8 30 A/m A Safety Approvals Safety Agency Safety Standard Notes & Conditions 0 ANSI/AMMI ES60601-1 UL CSA CSA C22.2 No. 60601-1 TUV EN60601-1 CB IEC60601-1 Mechanical Details 0.76 (19.5) 0.02 (0.50) 0.36 (9.20) Printed Face 0.42 (10.6) 0.44 (11.1) 0.12 (3.0) 1 2 3 5 6 Bottom View 0.36 (9.20) 0.08 (2.13) 7 0.02 (0.50) 0.02 (0.50) Pin Connections 0.10 (2.54) 0.6 (15.24) Notes 4. Pin pitch tolerance: 0.014 (0.35) 5. Case tolerance: 0.02 (0.5) 1. All dimensions are in inches (mm) 2. Weight: 0.008 lbs (3.6 g) approx. 3. Pin diameter: 0.020.002 (0.50.05) 0 0.01 (0.25) 0.13 (3.20) Application Note External Filter for Surge and EFT -Vin 3 Remote On/Off Remote On/Off 5 6 7 +Vout -Vout No Pin +Vout Common -Vout + Vin L1 IMM01 Dual -Vin +Vin + Vout + Vout D1 Single -Vin +Vin EMI Filter Conducted Emissions + Vin C1 Pin 1 2 C1 Load C2 IMM01 -Vin -Vout C1 is 220 F, 100 V electrolytic capacitor D1 is 18 V, 3 kW TVS for 5 V input or 28 V, 3 kW TVS for 12 V input L -Vout C1 & C2 are 10 F, 35 V multilayer ceramic chip capacitors, placed as close as possible to the input pins L1 is 12 H inductor www.xppower.com 3 Load 15 March 19