120W Reliable Green Medical Power Supply RPS-120 series
RPS - 120
Model Encoding
Description
RPS-120 is a 120W highly reliable green PCB type medical power supply with a high power density on
a 4 by 2 footprint. It accepts 80~264VAC input and offers various models with the output voltages between
12V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down
below 0.3W. RPS-120 is able to be used for both Class (with FG) or Class (no FG) system design. The extremely ⅠⅡ
low leakage current is less than A. In addition, it conforms to the international medical regulations (2*MOPP) 150
and EMC EN55011, perfectly fitting all kinds of BF rated patient contact medical system equipment.
4"×2" compact size
Medical safety approved (2 x MOPP) according to
ANSI/AAMI ES60601-1 and IEC/EN60601-1
Suitable for BF application with appropriate system consideration
84W convention, 120W force air
EMI Class B for both Class (with FG) & Class (no FG) configurationⅠⅡ
No load power consumption<0.3W
Extremely low leakage current
12V/0.5A fan supply
Protections: Short circuit / Overload / Over voltage / Over temperature
Operating altitude up to 4000 meters
3 years warranty
Oral irrigator
Hemodialysis machine
Medical monitors
Sleep apnea devices
Pumps machine
Features Applications
Output voltage
Rated wattage
Series name
-12-C
Blank : PCB type
-C : Enclosed type
{
++
EN 60601-1
IEC 60601-1
8
IEC60601-1EN60601-1ANSI/AAMI ES60601-1
μ
File Name:RPS-120-SPEC 2020-10-20
RPS-120 series
SPECIFICATION
MODEL
DC VOLTAGE
CURRENT 10CFM
10CFM
Convection
Convection
RATED
POWER
OUTPUT
LINE REGULATION
VOLTAGE ADJ. RANGE
LOAD REGULATION
SETUP, RISE TIME
HOLD UP TIME
(Typ.)
VOLTAGE RANGE
Note.4
FREQUENCY RANGE
EFFICIENCY
(Typ.)
INPUT
FUNCTION
INRUSH CURRENT
(Typ.)
ISOLATION RESISTANCE
WORKING HUMIDITY
WORKING TEMP.
FAN SUPPLY
OVER TEMPERATURE
STORAGE TEMP., HUMIDITY
TEMP. COEFFICIENT
VIBRATION
MTBF
DIMENSION (L*W*H)OTHERS
NOTE
PACKING
OVERLOAD
OVER VOLTAGE
AC CURRENT
(Typ.)
500ms, 30ms/230VAC 500ms, 30ms/115VAC at full load
50ms/230VAC 10ms/115VAC at full load
80 ~ 264VAC 113 ~ 370VDC
47 ~ 63Hz
89%89% 90%
16.5 ~ 19.5V
90%
26.4 ~ 31.2V
91%
29.7 ~ 35V 52.8 ~ 62.4V
2.1A/115VAC 1.2A/230VAC
COLD START 30A/115VAC 60A/230VAC
115~150% rated output power
13.2 ~ 15.6V
Protection type : Hiccup mode, recovers automatically after fault condition is removed
20 ~ 90% RH non-condensing
-30 ~ +70 (Refer to "Derating Curve")
12V@0.5A for driving a fan ; tolerance -15% ~ +10%
-40 ~ +85 , 10 ~ 95% RH non-condensing
±℃ 0.03%/ (0 ~ 50 )
10 ~ 500Hz, 2G 10min./1cycle, 60min. each along X, Y, Z axes
653.5Khrs min. MIL-HDBK-217F (25 )
PCB:101.6*50.8*29mm ; Enclosed type:103.4*62*40mm
or 4" * 2" *1.141" inch or 4.07" * 2.44" *1.57" inch
PCB:0.15Kg; 72pcs/11.8Kg/0.84CUFT ; Enclosed type:0.24Kg; 60pcs/15.4Kg/1.06CUFT
±1.0%
±0.5%
±1.0%
±0.5%
±1.0%
±0.5% ±0.5%
±1.0% ±1.0%
±0.5%
±2.0% ±2.0% ±1.0% ±1.0% ±1.0%
100mVp-p
11.4~12.6V 14.3~15.8V 22.8~25.2V 25.6 ~ 28.4V 45.6 ~50.4V
120mVp-p 150mVp-p 150mVp-p 150mVp-p
15V 24V 27V 48V
10A
7.0A
8A
5.6A
5A
3.5A
4.5A
3.15A
2.5A
1.75A
120W
84W 84W
120W
84W
120W
85W
121.5W
84W
120W
RPS-120-12RPS-120-15RPS-120-24RPS-120-27RPS-120-48
SAFETY STANDARDS
RIPPLE & NOISE (max.)
Note.2
VOLTAGE TOLERANCE Note.3
1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature.
2. Ripple & noise are measured at 20MHz of bandwidth by using a 12" twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor.
3. Tolerance : includes set up tolerance, line regulation and load regulation.
4. Derating may be needed under low input voltages. Please check the derating curve for more details.
5. Touch current was measured from primary input to DC output.
6. The ambient temperature derating of 3.5 /1000m with fanless models and of 5 /1000m with fan models for operating altitude higher than 2000m(6500ft).℃℃
7. The power supply is considered a component which will be installed into a final equipment. All the EMC tests are been executed by
mounting the unit on a mm* mm metal plate with 1mm of thickness. The final equipment must be re-confirmed that it still meets 360 360
EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies.
(as available on http://www.meanwell.com)
Product Liability Disclaimer For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
ENVIRONMENT
SAFETY &
EMC
(Note 7)
PROTECTION
Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP
IEC60601-1, TUV EN60601-1, , UL ANSI / AAMI ES60601-1 (3.1 version),EAC TP TC 004
CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved; Design refer to EN60335 1-
12V
LEAKAGE CURRENT(max.) Note.5
Earth leakage current < 150 A/264VAC , Touch current < 80 A/264VAC
WITHSTAND VOLTAGE
ISOLATION RESISTANCE
I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC
I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25 / 70% RH
Protection type : Shut down o/p voltage, re-power on to recover
Protection type : Shut down o/p voltage, re-power on to recover
Parameter
Parameter
ESD
Standard
Standard
EN55011 (CISPR11)
EN55011 (CISPR11)
EN61000-3-2
EN61000-3-3
EN61000-4-2
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Test Level / Note
Test Level / Note
Class B
Class B
Class A
-----
EMC IMMUNITY
EMC EMISSION
EN60601-1-2
100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
OPERATING ALTITUDE
Note.6
4000 meters
120W Reliable Green Medical Power Supply
μ
μ
μ μ
RF field susceptibility
EFT bursts
Surge susceptibility
Conducted susceptibility
Magnetic field immunity
Voltage dip, interruption
Conducted emission
Radiated emission
Harmonic current
Voltage flicker
Level 4, 15KV air ; Level 4, 8KV contact
Level 4, 30A/m
Level 3, 10V
Level 3, 2KV
Level 4, 4KV/Line-FG; 2KV/Line-Line
Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
File Name:RPS-120-SPEC 2020-10-20
Block Diagram
Output Derating VS Input Voltage
Derating Curve
fosc : 65KHz
CIRCUIT
DETECTION
O.L.P.O.T.P.
EMI
FILTER
POWER
SWITCHING FILTER
FILTER
&
&
RECTIFIERS
RECTIFIERS
+V
-V
+12V
FAN SUPPLY
I/P
FG
CONTROL
O.V.P.
EMI FILTER
RECTIFIER
&
INPUT VOLTAGE (VAC) 60Hz
80 11590 130120 160140 200180 240220 264
LOAD (%)
90
100
80
70
60
50
40
AMBIENT TEMPERATURE ( )
LOAD (%)
-30 0 2010 30 40 45
20
40
80
60
70
100
6050 70
(HORIZONTAL)
With 10CFM Fan for RPS-120(-C)
Convection
for RPS-120(-C) 24~48V
Convection
for RPS-120(-C) 12~15V
RPS-120 series
120W Reliable Green Medical Power Supply
File Name:RPS-120-SPEC 2020-10-20
Mechanical Specification
FS1
HS1
HS2
T4A/250V
T4A/250V
RPS-120 series
50.8
44.45 3.175
1
2
3
CN1
FS2
3.175 95.25
101.6
4- 3.3ψ
29
3 max.
1
2
3
CN2
4
12
CN101 SVR1 LED
FAN 10CFM
direction
Air flow
5cm
Note
Top View
Side View
120W Reliable Green Medical Power Supply
RPS-120 (PCB Type)
File Name:RPS-120-SPEC 2020-10-20
50 8.
Mechanical Specification
Case No.245A Unit:mm
FAN 10CFM
direction
Air flow
5cm
RPS-120 series
30.8
62
20
25
40
100
103.4
1.7
2-M3 L=2
2-M3 L=4
2-R1.75
12.7 78
24.2 55
1
2
3
CN1 1
2
3
4
CN2
CN101
12
Side View
Top View
Side View
Side View
Side View
120W Reliable Green Medical Power Supply
RPS-120-C (Enclosed Type)
Bottom View
File Name:RPS-120-SPEC 2020-10-20
571.
AC Input Connector (CN1) : JST B3P-VH or equivalent
Pin No.
1
2
3
Assignment
AC/N
No Pin
AC/L
Mating Housing
JST VHR JST SVH-21T-P1.1
or equivalent or equivalent
Terminal
DC Output Connector (CN2) : JST B4P-VH or equivalent
Pin No.
1,2
3,4
Assignment
-V
+V
Mating Housing
JST VHR
or equivalent
Terminal
JST SVH-21T-P1.1
or equivalent
Mating Housing
JST PHR-2
or equivalent
Terminal
JST SPH-002T-P0.5S
or equivalent
FAN Connector(CN101) : or equivalentJST S2B-PH-K-S
Pin No.
1
2
Assignment
+12V(FAN+)
DC COM(FAN-)
1.HS1,HS2 cannot be shorted.
2.HS1 must have safety isolation distance with system case.
RPS-120 series
Note : 1. The FAN supply is designed to serve as the source of the additive external fan for the cooling of the power supply,
enabling the full load delivery and assuring the best life span of the product. Please do not use this FAN supply to
drive other devices.
2.The PCB type(Blank type)model delivers EMI Class B for both conducted emission and radiated emission for the power
supply, when configured into either Class (with FG) or ClassⅠⅡ(no FG) system.
(no FG)3.The Enclosed type(-C type) model is not suitable for the configuration within a Class system but is
suggested to used within a Class (with FG) system.
Installation Manual
Please refer to : http://www.meanwell.com/manual.html
120W Reliable Green Medical Power Supply
File Name:RPS-120-SPEC 2020-10-20