AC Voltage Simple Logger (R) Models L205, L230 & L260 Electrical Number of Channels: One Measurement Range: 0 to 25Vrms (L205) 0 to 300Vrms (L230) 0 to 600Vrms (L260) Input Connection: Recessed safety 4mm banana jacks Resolution: 8 Bit L205: 12.5mV max L230: 250mV max L260: 500mV max Accuracy*: 1% of Reading + Resolution Sample Rate: 4096/hr max, decreases by 50% each time memory is full Storage Technique: Time Extension RecordingTM (TXRTM) Data Storage: 8192 readings Power: 9V Alkaline battery Battery Life: Up to one yr of recording @ 25C Output: RS-232 via DB-9 connector Indicators: Red LED double flashes in LOGGING, single flashes in STANDBY, and is off in OFF mode Controls: One button used to start and stop recording sessions and to turn the data logger on and off MECHANICAL Size: 2 7/8 x 2 5/16 x 15/8" (73 x 59 x 41mm) Weight: (with battery) 5 oz (140g) Mounting: Base plate mounting holes or Velcro(R) pads ENVIRONMENTAL Operating Temperature: -4 to +158F (-20 to +70C) Storage Temperature: -4 to +174F (-20 to +80C) Relative Humidity: 5 to 95% non-condensing *Three input types to choose from: Model L205 0 to 25Vrms input (stray voltage) Model L230 0 to 300Vrms input Model L260 0 to 600Vrms input *True RMS measurements *Inputs via color coded safety Banana Jacks ORDERING INFORMATION *Scale automatically selected by software *Includes test leads and alligator clips *Double Insulation *CE Mark *HVAC troubleshooting *Line voltage monitoring *Surge/Sag monitoring *Stray voltage monitoring *Dropout monitoring CATALOG NO. Simple Logger(R) Model L205, Stray Voltage with Leads (0 to 25.5Vac Input). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2116.05 Simple Logger(R) Model L230, RMS Voltage with Leads (0 to 300Vac Input) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2113.93 Simple Logger(R) Model L260, RMS Voltage with Leads (0 to 600Vac Input). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2113.94 All AC Voltage Simple Loggers(R) are supplied with software (CD-ROM), 6 ft DB-9 RS-232 cable, 9V Alkaline battery, 5 ft lead set, and user manual Accessories Safety Grip Probes, set of 2. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2111.31 110V US Outlet Adaptor with Banana Jacks (for Models L230 or L260) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2118.49 Voltage Leads, (2) with clips, 5 ft. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Cat. #2118.51 L205-30-60_211605-211393-211394 (pdf) 0308 Rev. 05 Technical Assistance (800) 343-1391 www.aemc.com 1 of 4 Simple Logger (R) Software Features Computer Requirements *Very easy to setup and use -- no programming required *Export data to spreadsheet and database software *Statistics appear on screen: Max, Min, Avg, Date and Time *Print both graphs and tabular listings *On-Line display of data logger values *Zoom capability with automatic statistics update *User keyed text annotations *Automatically displays values at the cursor position *Paste graphs into other applications *Programmable scale and units for all DC loggers. Automatic selection for AC loggers *Automatic baud rate selection (version 6.0 or later) *486 or higher *8MB of RAM *Windows(R)95/98/2000/ME/XP or Windows (R)NT *One serial port not in use by any other device *8MB hard drive space for program and 400kB max per file stored Plot graphs and view statistics quickly and easily when you download data from Simple Logger (R) to your computer. Scale and time axis update automatically. Click and drag your mouse to zoom up the desired area of the graph for better viewing. The Min, Max and Average statistics at the bottom of the graph update automatically with zoom window. Click the Annotate button to add a title and descriptive information to your graph. This information can be stored and printed with a graph. L205-30-60_211605-211393-211394 (pdf) 0308 Rev. 05 Technical Assistance (800) 343-1391 www.aemc.com 2 of 4 On-line display of data Automatic update of Y data at cursor position Annotate button Plot area with value and time data shown on the X and Y axis Graph Statistics Click the Annotate button to add a title and descriptive information to your graph. This information can be stored and printed with a graph. L205-30-60_211605-211393-211394 (pdf) 0308 Rev. 05 Technical Assistance (800) 343-1391 www.aemc.com 3 of 4 Contact Us United States & Canada: Chauvin Arnoux(R), Inc. d.b.a. AEMC (R) Instruments 200 Foxborough Blvd. Foxborough, MA 02035 USA (508) 698-2115 * Fax (508) 698-2118 www.aemc.com Customer Support - for placing an order, obtaining price & delivery: customerservice@aemc.com Sales Department - for general sales information: sales@aemc.com Repair and Calibration Service - for information on repair & calibration, obtaining a user manual: repair@aemc.com Technical and Product Application Support - for technical and application support: techinfo@aemc.com Webmaster - for information regarding www.aemc.com: webmaster@aemc.com South America, Central America, Mexico, Caribbean, Australia & New Zealand: Chauvin Arnoux(R), Inc. d.b.a. AEMC (R) Instruments 15 Faraday Drive Dover, NH 03820 USA (978) 526-7667 * Fax (978) 526-7605 export@aemc.com www.aemc.com All other countries: Chauvin Arnoux SCA 190, rue Championnet 75876 Paris Cedex 18, France 33 1 44 85 45 28 * Fax 33 1 46 27 73 89 info@chauvin-arnoux.com www.chauvin-arnoux.com L205-30-60_211605-211393-211394 (pdf) 0308 Rev. 05 Technical Assistance (800) 343-1391 www.aemc.com 4 of 4