Technical Data
Product highlights
The 721 and 721Ex can be configured by select-
ing either a 16 PSI (1.1 bar) or 36 psi (2.48 bar)
low pressure sensor then add any of 7 high
pressure ranges including 100, 300, 500, 1000,
1500, 3000 or 5000 psi (6.9, 20, 24.5, 69, 103.4,
200, 345 bar).
Intrinsically Safe (721Ex version only) – IECEx
and Atex Ex ia IIB T3 Gb (Zone 1)
Ideal for gas flow calibration (custody transfer)
applications
Two isolated, stainless steel, pressure sensors
with 0.025
% accuracy
Switch test on both pressure inputs
Pressure displayed in the user’s choice of 19
engineering units
Pt100 RTD input for temperature measurement,
(probe optional)
Measures 4 to 20 mA signals
Internal 24 V loop power supply can power a
transmitter under test (721 models only)
Measure up to 30 V dc, check 24 V loop power
supplies (721 models only)
Extend pressure measurement range with
connection to one of 50 external 750 series
pressure modules from 0 to 1 inH20 to 10,000
psi (2.5 mBar to 690 bar) (721 models only)
Large backlit graphic display can display up to
three inputs simultaneously
Store five instrument setups for recall and use
Fluke 721 and 721Ex
Precision Pressure
Calibrators
Pressure calibration and temperature
measurement for custody transfer
applications
The Fluke 721 and 721Ex Precision Pressure Calibrators
with dual isolated pressure sensors are the ideal tool
for gas custody transfer applications allowing you to
take simultaneous static and differential pressure mea-
surements with a single tool.
2 Fluke Corporation Fluke 721 and 721Ex Precision Pressure Calibrators
Low pressure sensor High pressure sensor
Model Description Range sensor 1 Resolution
sensor 1
Accuracy
sensor 1
Range sensor 2 Resolution
sensor 2
Accuracy
sensor 2
Fluke-721-1601
Fluke-721Ex-1601
Dual Sensor Pressure Cali-
brator, 16 PSIG, 100 PSIG
-14 psi to +16 psi
-0.97 bar to 1.1
bar
0.001 psi,
0.0001 bar
0.025
% of
full scale
-12 ps i to +100 psi
-0.83 bar to 6.9 bar
0.01 psi
0.0001 bar
0.025
% of
full scale
Fluke-721-1603
Fluke-721Ex-1603
Dual Sensor Pressure Cali-
brator, 16 PSIG, 300 PSIG
-12 ps i to +300 psi
-0.83 bar to 20 bar
0.01 psi
0.001 bar
Fluke-721-1605
Fluke-721Ex-1605
Dual Sensor Pressure Cali-
brator, 16 PSIG, 500 PSIG
-12 psi to +500 psi
-0.83 bar to 34.5 bar
0.01 psi
0.001 bar
Fluke-721-1610
Fluke-721Ex-1610
Dual Sensor Pressure Cali-
brator, 16 PSIG, 1000 PSIG
0 psi to +1000 psi
0.00 bar to 69 bar
0.1 psi
0.001 bar
Fluke-721-1615
Fluke-721Ex-1615
Dual Sensor Pressure Cali-
brator, 16 PSIG, 1500 PSIG
0 psi to +1500 psi
0.00 bar to 103.4 bar
0.1 psi
0.001 bar
Fluke-721-1630
Fluke-721Ex-1630
Dual Sensor Pressure Cali-
brator, 16 PSIG, 3000 PSIG
0 psi to +3000 psi
0.00 bar to 200 bar
0.1 psi
0.01 bar
Fluke-721-1650
Fluke-721Ex-1650
Dual Sensor Pressure Cali-
brator, 16 PSIG, 5000 PSIG
0 p si to +5000 psi
0.00 bar to 345 bar
0.1 psi
0.01 bar
0.035
% of
full scale
Fluke-721-3601
Fluke-721Ex-3601
Dual Sensor Pressure Cali-
brator, 36 PSIG, 100 PSIG
-14 psi to +36 psi
-0.97 bar to 2.48
bar
-12 to +100 psi
-0.83 bar to 6.9 bar
.01 psi
0.0001 bar
0.025
% of
full scale
Fluke-721-3603
Fluke-721Ex-3603
Dual Sensor Pressure Cali-
brator, 36 PSIG, 300 PSIG
-12 ps i to +300 psi
-0.83 bar to 20 bar
0.01 psi
0.001 bar
Fluke-721-3605
Fluke-721Ex-3605
Dual Sensor Pressure Cali-
brator, 36 PSIG, 500 PSIG
-12 psi to +500 psi
-0.83 bar to 34.5 bar
0.01 psi
0.001 bar
Fluke-721-3610
Fluke-721Ex-3610
Dual Sensor Pressure Cali-
brator, 36 PSIG, 1000 PSIG
0 psi to +1000 psi
0.00 bar to 69 bar
0.1 psi
0.001 bar
Fluke-721-3615
Fluke-721Ex-3615
Dual Sensor Pressure Cali-
brator, 36 PSIG, 1500 PSIG
0 psi to +1500 psi
0.00 bar to 103.4 bar
0.1 psi
0.01 bar
Fluke-721-3630
Fluke-721Ex-3630
Dual Sensor Pressure Cali-
brator, 36 PSIG, 3000 PSIG
0 psi to +3000 psi
0.00 bar to 200 bar
0.1 psi
0.01 bar
Fluke-721-3650
Fluke-721Ex-3650
Dual Sensor Pressure Cali-
brator, 36 PSIG, 5000 PSIG
0 p si to +5000 psi
0.00 bar to 345 bar
0.1 psi
0.01 bar
0.035
% of
full scale
Temperature effect
No effect on accuracy on all functions from 15 °C to 35 °C
Add ± 0.002
% F.S./°C for temps outside of 15 °C to 35 °C
General specifications
Instrument setup recall Last used on power-up
Environmental Operating temperature -10 °C to
+50 °C (Fluke 721)
Operating temperature -10 °C to
+45 °C (Fluke 721Ex)
Storage temperature (without
batteries) -20 °C to +60 °C
Altitude 2,000 meters
Ingress protection IP-40
Power requirements 6.0 v dc
Battery life >35 hours, typical usage
Batteries required (4) Standard AA cells
Dimentions (HxWxD) 7.9” X 4.3” X 2.3” (20 cm X
11 cm X 5.8 cm)
Weight 03539 kg (1 lb 3 oz)
Connectors/ports pressure Two, 1/8 in NPT, 750P Series
Pressure Module (721 models
only); RTD Probe
Specifications 1-year
Electrical and temperature measurement 1-year
Function Range Resolution Accuracy
mA measure 0 to 24 mA 0.001 mA ±0.015% of rdg
±0.002mA
VDC measure** 0 to 30 VDC 0.001V ±0.015% of rdg
±0.002V
*Temperature mea-
sure (RTD/Ohms)
-40 °C to 150 °C
(-40 °F to 302 °F)
0.01 °C, 0.01 °F ±0.015% of rdg
±0.02 ohms;
±0.1°C (±0.2°F)
±0.25 degC (±0.45
degF) combined
uncertainty when
using 720 RTD
probe accessory
Loop power supply** 24V N/A 24 mA @ 24V
*Temperature measurement requires optional Fluke-720RTD Pt-100 RTD probe
available as an accessory **721 models only
3 Fluke Corporation Fluke 721 and 721Ex Precision Pressure Calibrators
User selectable pressure engineering units
psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C,
mH2O@4°C, mH2O@20°C, inH2O@4°C, inH2O@20°C,
inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C,
ftH2O@20°C, ftH2O@60°F
Fluke Corporation
PO Box 9090, Everett, WA 98206 U.S.A.
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800)-36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2013, 2015 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 3/2015 6000077B_EN
Pub_ID: 12085-eng
Modification of this document is not permitted
without written permission from Fluke Corporation.
Fluke. Keeping your world
up and running.®
Ordering information
Fluke-721-1601
Fluke-721Ex-1601
Dual Sensor Pressure Calibrator
16 PSIG, 100 PSIG
Fluke-721-1603
Fluke-721Ex-1603
Dual Sensor Pressure Calibrator
16 PSIG, 300 PSIG
Fluke-721-1605
Fluke-721Ex-1605
Dual Sensor Pressure Calibrator
16 PSIG, 500 PSIG
Fluke-721-1610
Fluke-721Ex-1610
Dual Sensor Pressure Calibrator
16 PSIG, 1000 PSIG
Fluke-721-1615
Fluke-721Ex-1615
Dual Sensor Pressure Calibrator
16 PSIG, 1500 PSIG
Fluke-721-1630
Fluke-721Ex-1630
Dual Sensor Pressure Calibrator
16 PSIG, 3000 PSIG
Fluke-721-1650
Fluke-721Ex-1650
Dual Sensor Pressure Calibrator
16 PSIG, 5000 PSIG
Fluke-721-3601
Fluke-721Ex-3601
Dual Sensor Pressure Calibrator
36 PSIG, 100 PSIG
Fluke-721-3603
Fluke-721Ex-3603
Dual Sensor Pressure Calibrator
36 PSIG, 300 PSIG
Fluke-721-3605
Fluke-721Ex-3605
Dual Sensor Pressure Calibrator
36 PSIG, 500 PSIG
Fluke-721-3610
Fluke-721Ex-3610
Dual Sensor Pressure Calibrator
36 PSIG, 1000 PSIG
Fluke-721-3615
Fluke-721Ex-3615
Dual Sensor Pressure Calibrator
36 PSIG, 1500 PSIG
Fluke-721-3630
Fluke-721Ex-3630
Dual Sensor Pressure Calibrator
36 PSIG, 3000 PSIG
Fluke-721-3650
Fluke-721Ex-3650
Dual Sensor Pressure Calibrator
36 PSIG, 5000 PSIG
Fluke-720RTD RTD Probe for Fluke-721, 721Ex
and 719Pro
Fluke-720URTDA Universal RTD adapter for
Fluke-721, 721Ex and 719Pro
Included:
Soft case
Four (4) AA batteries
Manual
Traceable calibration certificate
Test leads
721Ex Approval Details
ATEX Ex ia IIB T3 Gb (Ta= -10… +45 °C)
KEMA 10 ATEX 0168X
IECEx Ex ia IIB T3 Gb (Ta= -10…+45 °C)
II 2 G IECEx CSA 10.0013X