Agilent 34330A Current Shunt
© Copyright Agilent Technologies, Inc.
2003, 2011
Part Number: 34330-90001
Edition 2, July 2011
Printed in Malaysia
Agilent Technologies, Inc.
5301 Stevens Creek Blvd.
Santa Clara, CA 95052 USA
Safety Notices
Do not attempt to repair this sealed
device. Do not use this device for any
purpose other than indicated in these
instructions. Doing so may expose you
to bodily injury or cause equipment
damage.
Do not install substitute parts or
perform any unauthorized modification
to the product. Return the product to an
Agilent Technologies Sales and Service
Office for service and repair to ensure
that safety features are maintained.
WARNING
A WARNING notice denotes a hazard.
It calls attention to an operating proce-
dure, practice, or the like that, if not
correctly performed or adhered to,
could result in personal injury or death.
Do not proceed beyond a WARNING
notice until the indicated conditions are
fully understood and met.
CAUTION
A CAUTION notice denotes a hazard.
It calls attention to an operating proce-
dure, practice, or the like that, if not
correctly performed or adhered to,
could result in damage to the product or
loss of important data. Do not proceed
beyond a CAUTION notice until the
indicated conditions are fully understood
and met.
Symbols
Instructions for Use
The Agilent 34330A Current Shunt is
a precision 0.001Ω resistor housed in a
plastic case surrounded by epoxy.
Output is 1 mV per A of current
passing through the shunt.
WARNING
Install the shunt in the multimeter
voltage input terminals before use. Do
not remove the shunt with a current
source still connected, as there may be
danger of shorting the banana plugs to
ground. Observe the current and volt-
age limits listed under “Specifications.”
The following simplified diagram shows
how the shunt is used:
Insert the shunt’s banana-plug outputs
into the main voltage inputs on the
multimeter (not the high ampere
inputs). Then connect the current
source to be measured to the shunt’s
input terminals (binding posts). The
multimeter will read the IR drop caused
by the input current through the shunt.
The precision resistance of the shunt is
0.001Ω, so the voltage measured by the
multimeter is equal to the current
multiplied by 0.001Ω. Thus, the voltage
reading sensitivity is 1 mV/A. For a
15 A current, the voltage reading is
15 mV. Select an appropriate voltage
range on the multimeter. See your
multimeter User’s Guide for further
information.
Cleaning Instructions
Wipe with a soft cloth, moistened
slightly with water, if needed.
*34330-90001*
34330-90001
Specifications
Operating Environment: Indoor use
only
Operating Temperature: 5 to 40 °C
Operating Humidity: 0 to 80% RH
Storage Temperature: -30 to 70 °C
Input Current:
15 A (continuous)
30 A (15 minutes, max)
Safety Limit, Max Voltage to Ground:
60 Vdc, max
30 Vac (rms), max
Voltage Sensitivity: 1 mV/A
Accuracy (DC - 1 kHz): ±0.3%
Accuracy (1 kHz - 5 kHz): ±5%
Assistance
Product maintenance agreements and
other customer assistance agreements
are available for Agilent Technologies
products. For assistance, contact your
nearest Agilent Technologies Sales and
Service Office. Further information is
available on the Agilent web site at
www.agilent.com/find/assist.
Certification
Agilent Technologies certifies that this
product met its published specifications
at the time of shipment from the factory.
The shunt is not subject to periodic
calibration.
Warranty
The material contained in this
document is provided “as is,” and is
subject to being changed, without
notice, in future editions. Further,
to the maximum extent permitted by
applicable law, Agilent disclaims all
warranties, either express or implied
with regard to this manual and any
information contained herein, includ-
ing but not limited to the implied
warranties of merchantability and fit-
ness for a particular purpose. Agilent
shall not be liable for errors or for
incidental or consequential
damages in connection with the
furnishing, use, or performance of this
document or any information contained
herein. Should Agilent and the user
have a separate written agreement with
warranty terms covering the material
in this document that conflict with
these terms, the warranty terms in the
separate agreement will control.
instructions for import-
Refer to the operating
ant safety information.
Current Shunt.backup.fm Page 0 Wednesday, June 22, 2011 3:51 PM