Technical Data
Fluke 355/353
True-rms 2000 A Clamp Meters
High current and
a large jaw, make
these clamps
well suited for
industrial and
utility applications
Confidently take reliable readings
with the true-rms, Fluke 355/353
Clamp Meters; the tools of choice
for high-current measurements
up to 2000 A. The extra-wide
jaw easily clamps around large
conductors typically found in high-
current applications. The rugged
design and CAT IV 600 V, CAT III
1000 V ratings, add an extra ele-
ment of protection when taking
high-powered measurements.
Accurate peak measurements
can be taken using the in-rush
current mode, ideal for motors
and inductive loads. The 355 also
measures voltage and resistance,
making this the most versatile tool
for utilities, electrical contractors,
and industrial service technicians.
Reliably handle a wide range of high-current
applications with 2000 A ac + dc true-rms,
1400 A ac, and 2000 A dc
The large 58 mm (2.3 in) jaw capacity is suitable for
large, or multiple conductors
CAT IV 600 V, CAT III 1000 V rating for added user
protection
In-rush current measurement captures ‘power-on’
surge current with accuracy and repeatability
High voltage measurement of 1000 V ac + dc
true-rms, 600 V ac, and 1000 V dc allows users to
perform multiple tests with only one tool (355 only)
Resistance to 400 kW coupled with a continuity
beeper, provide the convenience of a multimeter
(355 only)
Accurately measure frequency up to 1 kHz for
optimum troubleshooting
Quickly analyze readings using the MIN, MAX, and
AVG functions
A large backlit display allows for easy visibility in
low-lit areas
Use the display hold feature to capture readings
even when the display cannot be viewed
Use the low-pass filter to smooth out noisy loads
and stabilize readings
N10140
2 Fluke Corporation Fluke 355/353 True-rms 2000 A Clamp Meters
Specifications
Electrical specifications
Current measurement dc and ac 10 Hz to 100 Hz
Range Resolution Accuracy, A Trigger Level for
Inrush
Trigger Level
for Hz
Filter OFF
Trigger Level
for Hz
Filter ON
40 A 10 mA 1.5 % rdg + 15 digits 0.50 A 2.50 A 0.50 A
400 A 100 mA 1.5 % rdg + 5 digits 5.0 A 2.5 A 2.5 A
2000 A;
1400 ac rms
1 A 1.5 % rdg + 5 digits 5 A 8 A 8 A
Crest Factor (50/60 Hz)
Range Crest Factor*
40 A 2 @ 33 A, 2.4 @ 27 A
400 A 2 @ 330 A, 2.4 @ 270 A
2000 A; 1400 ac rms 2 @ 1000 A, 2.4 @ 833 A
*Add 2 % to error spec for CF > 2
Current measurement ac 100.1 Hz to 1 kHz
Range Resolution Accuracy > 10 A Trigger Level for
Inrush
Trigger Level
for Hz
Filter OFF
Trigger Level
for Hz
Filter ON
40 A 10 mA 3.5 % rdg + 15 digits 0.50 A 2.50 A 0.50 A
400 A 100 mA 3.5 % rdg + 5 digits 5.0 A 2.5 A 2.5 A
2000 A;
1400 ac rms
1 A 3.5 % rdg + 5 digits 5 A 8 A 8 A
Voltage measurement (355 only) dc and ac 10 Hz to 100 Hz
600 V and 1000 V ranges have 10 % over range to 660 V and 1100 V respectively.
Range Resolution Accuracy Trigger Level for Hz
Filter OFF
Trigger Level for Hz
Filter ON
4 V 1 mV 1 % rdg + 10 digits 0.050 V 0.050 V
40 V 10 mV 1 % rdg + 5 digits 0.25 V 0.25 V
400 V 100 mV 1 % rdg + 5 digits 6 V 6 V
600 V ac rms 1 V 1 % rdg + 5 digits 6 V 6 V
1000 V dc 1 V 1 % rdg + 5 digits N/A N/A
Voltage measurement (355 only) ac 100.1 Hz to 1 kHz
600 V and 1000 V ranges have 10 % over range to 660 V and 1100 V respectively.
Range Resolution Accuracy Trigger Level for Hz
Filter OFF
Trigger Level for Hz
Filter ON
4 V 1 mV 3 % rdg + 10 digits 0.050 V 0.050 V
40 V 10 mV 3 % rdg + 5 digits 0.25 V 0.25 V
400 V 100 mV 3 % rdg + 5 digits 6 V 6 V
600 V ac rms 1 V 3 % rdg + 5 digits 6 V 6 V
Ohms measurement (355 only)
Range Resolution Accuracy
400 W0.1 W1.5 % + 5 digits
4 kW1 W1.5 % + 5 digits
40 kW10 W1.5 % + 5 digits
400 kW100 W1.5 % + 5 digits
3 Fluke Corporation Fluke 355/353 True-rms 2000 A Clamp Meters
Fluke Corporation
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.
PO Box 1186, 5602 BD
Eindhoven, The Netherlands
For more information call:
In the U.S.A. (800) 443-5853 or
Fax (425) 446-5116
In Europe/M-East/Africa +31 (0) 40 2675 200 or
Fax +31 (0) 40 2675 222
In Canada (800) 36-FLUKE or
Fax (905) 890-6866
From other countries +1 (425) 446-5500 or
Fax +1 (425) 446-5116
Web access: http://www.fluke.com
©2007 Fluke Corporation.
Specifications subject to change without notice.
Printed in U.S.A. 9/2007 3094045 D-EN-N Rev A
Fluke. Keeping your world
up and running.®
Ordering information:
Fluke-353 AC/DC True-rms Clamp Meter, 2000 A, Amps only
Includes: C43 Soft Carrying Case, 6 AA batteries, and users manual
Fluke-355 AC/DC True-rms Clamp Meter, 2000 A
Includes: C43 Soft Carrying Case, 6 AA batteries, TL224 1.5 m
Silicone Rubber Test Leads, TP2 Test Probes, AC285 Alligator Clips,
and users manual
Continuity beeper (355 only)
On at 30 W
Off at 100 W
Frequency measurement
Measurement range 5.0 Hz to 1 kHz
Resolution 0.1 Hz (15 Hz to 399.9 Hz); 1 Hz (400 Hz to 1 kHz)
Accuracy – 5.0 Hz to 100 Hz 0.2 % + 2 counts
Accuracy – 100.1 Hz to 1 kHz 0.5 % + 5 counts
Trigger level Refer to current and voltage tables
General specifications
Batteries Six 1.5 V AA NEDA 15 A or IEC LR6
Battery life (with typical usage, backlight off) 100 hours
Test leads Rated to 1000 V
Weight .814 kg (1.8 lb)
Jaw size 58 mm (2.28 in)
Dimensions (LxWxD) 300 mm x 98 mm x 52 mm (12 in x 3.75 in x 2 in)
Safety rating IEC 61010-2-032, 600 V CAT IV, 1000 V CAT III
Environmental specifications
Operating temperature 32 °F to + 122 °F (0 °C to +50 °C)
Storage temperature -4 °F to 140 °F (-20 °C to +60 °C)
Operating humidity 0 to 95 % (non-condensing)
Operating altitude 2000 m
Storage altitude 10,000 m
IP rating 42 (indoor use only)
Drop test requirements 1 m
EMI, RFI, EMC FCC part 15, IEC/EN 61326-1:1997 class B, IEC/EN
61326:1997 3V/m, performance criteria B, EN61325
Temperature coefficients Current: 0.1 % of reading per °C outside 22 °C to 24 °C
Voltage: 0.1 % of reading per °C outside 22 °C to 24 °C
Fluke-353Fluke-355